Skip to content

PINNACLE 2218-36-154Recalls & reported problems

Depuy (Ireland) · PINNACLE TRIAL LINER 54OD x 36ID + 4 10 DEG · Class 1

All models of PINNACLE (198)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Pinnacle trial liner is a plastic sizing trial used during hip replacement surgery. The 31 reports describe liners with scratches, deep cuts and cracks, and screws that strip or won't hold, with some liners breaking during a case; a few were found damaged after sterilization or back at the office. None of the reports describe deaths or injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
31
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 31 adverse event reports that list this exact device (DI 10603295101277), received between Nov 8, 2017 and Mar 4, 2026 — classified as 31 malfunctions.

The most frequently reported specific device problem is “Scratched Material” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Scratched Material11
  • Break9
  • Naturally Worn8
  • Device-Device Incompatibility4
  • Material Twisted/Bent4
  • Material Integrity Problem3
  • Component Missing2
  • Corroded2

Patient problems

  • No Consequences Or Impact To Patient15
  • No Clinical Signs, Symptoms or Conditions11
  • Insufficient Information3
  • No Information2
  • Not Applicable1

Recent adverse event reports

The 25 most recent of 31 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 4, 2026Break; Device-Device Incompatibility

    IT WAS REPORTED THAT THE LINER WOULD NOT SCREW IN AND FINALLY PUSHED THROUGH AND BROKE DURING THE CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24507679 ↗
  • MalfunctionJul 23, 2024Scratched Material; Scratched Material

    IT WAS REPORTED THAT THERE WERE SCRATCHES ON TRIALS. SURGICAL DELAY WAS UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 19806013 ↗
  • MalfunctionJul 16, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE SCREW CAME OUT OF TRIAL.

    Patient problems: Insufficient Information

    MAUDE report 19750266 ↗
  • MalfunctionFeb 19, 2024Detachment of Device or Device Component; Naturally Worn; Detachment of Device or Device Component; Naturally Worn

    IT WAS REPORTED THAT THE SCREW STRIPPED AND CAME OUT OF LINER TRIAL.

    Patient problems: Insufficient Information

    MAUDE report 18730320 ↗
  • MalfunctionJun 12, 2023Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT THE INSTRUMENTS WERE DAMAGED DURING STERILIZATION. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17105952 ↗
  • MalfunctionApr 3, 2023Scratched Material; Scratched Material

    IT WAS REPORTED THAT INSTRUMENTS DAMAGED DURING STERILIZATION. ALL OF THE POLYS ARE SCRATCHED. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16663916 ↗
  • MalfunctionDec 9, 2022Crack; Crack

    IT WAS REPORTED THAT 2 POLY TRIALS WERE INSPECTED BEFORE A CASE AND WERE CRACKED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15950509 ↗
  • MalfunctionJul 6, 2022Naturally Worn; Scratched Material; Naturally Worn; Scratched Material

    IT WAS REPORTED THAT THE TRIAL OLD, BADLY WORN WITH DEEP SCRATCHES, CUTS. FOUND WHEN SET RETURNED TO THE OFFICE FOR INSPECTION. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14935216 ↗
  • MalfunctionMar 23, 2022Break; Crack; Break; Crack

    TRIAL LINER CRACKED AND BROKE ON RIM. PIECES WERE RETRIEVED. NO SURGICAL DELAY. NO ADVERSE AFFECTS ON PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13861321 ↗
  • MalfunctionNov 18, 2021Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED AS "TRIALS IN POOR CONDITION."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12836666 ↗
  • MalfunctionMay 21, 2021Break; Break

    IT WAS REPORTED AS "BROKEN IN STERILIZATION". NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11864130 ↗
  • MalfunctionMar 16, 2021Corroded; Material Twisted/Bent; Corroded; Material Twisted/Bent

    IT WAS REPORTED THAT THE INSTRUMENTS WERE REPORTED DUE TO CORROSION AND WEAR. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11497421 ↗
  • MalfunctionMar 16, 2021Corroded; Material Twisted/Bent; Corroded; Material Twisted/Bent

    IT WAS REPORTED THAT THE INSTRUMENTS WERE REPORTED DUE TO CORROSION AND WEAR. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11497410 ↗
  • MalfunctionMar 16, 2021Break; Break

    IT WAS REPORTED THAT THE POLY TRIAL USED IN THIS CASE IS BROKEN AND NEEDS REPLACEMENT. THE METAL PART BROKE OFF FROM THE PLASTIC. NO FURTHER INFORMATION IS AVAILABLE. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11496292 ↗
  • MalfunctionMar 30, 2020Break; Component Missing; Break; Component Missing

    IT WAS REPORTED THAT THE 54 X 36/10¿ PINNACLE LINER TRIAL SCREW RING BROKE OFF. THE RING WAS THROWN OUT BY MISTAKE BUT THE LINER AND SCREW ARE BEING RETURNED. NO PIECES WERE MISSING. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9900187 ↗
  • MalfunctionFeb 11, 2020Device-Device Incompatibility; Scratched Material; Device-Device Incompatibility; Scratched Material

    IT WAS REPORTED THAT PLASTIC AROUND LOCKING RING OF THE PINNACLE 36MM +4 10 54 OD GOUGES. THE RING WONT STAY ON PROPERLY. THIS WAS NOTICED BY CSR AFTER MULTIPLE CASES OF THAT DAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9694921 ↗
  • MalfunctionJan 16, 2020Break; Component Missing; Break; Component Missing

    ADDITIONAL INFORMATION INDICATED THAT THE INDEX PROCEDURE DIDN¿T INVOLVE DEPUY PRODUCTS. THE REVISION CASE WAS REMOVAL OF A PROSTALAC AND REINSERTION OF IMPLANTS.

    Patient problems: Not Applicable

    MAUDE report 9597033 ↗
  • MalfunctionJan 3, 2020Contamination /Decontamination Problem; Device-Device Incompatibility; Contamination /Decontamination Problem; Device-Device Incompatibility

    IT WAS REPORTED THAT THE SURGEON COULD NOT GET LINER TRIAL TO SEAT IN CUP DUE TO DEBRIS IN CENTER HOLE OF CUP. HE ELECTED TO REMOVE THE SCREW IN ORDER TO TRIAL. IN DOING THIS, HE DAMAGED THE TRIAL. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9548908 ↗
  • MalfunctionOct 3, 2019Loose or Intermittent Connection; Scratched Material; Unintended Movement; Loose or Intermittent Connection; Scratched Material; Unintended Movement

    IT WAS REPORTED THAT THE LINER TRIAL NOTED BY CENTRAL STERILE DURING INSPECTION AS HAVING LINER SCREW IN MECHANISM "WOBBLY". THE LOCKING RING WILL NOT FULLY SEAT DOWN SECURELY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9152527 ↗
  • MalfunctionSep 24, 2019Break; Material Disintegration; Material Twisted/Bent; Scratched Material; Break; Material Disintegration; Material Twisted/Bent; Scratched Material

    ADDITIONAL INFORMATION RECEIVED THAT THE RING BENT AND CAME OFF BUT DID NOT BREAK IN PIECES.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9112319 ↗
  • MalfunctionSep 3, 2019Device-Device Incompatibility; Material Twisted/Bent; Device-Device Incompatibility; Material Twisted/Bent

    IT WAS REPORTED THAT THE CENTRAL STERILE CLEANING NOTIFIED OF A DAMAGED PINNACLE LINER TRIAL PART NUMBER 221836154. THE LOCKING WASHER IS BENT AND WILL NOT STAY IN POSITION. NEEDS REPLACEMENT ASAP.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8957716 ↗
  • MalfunctionAug 19, 2019Break; Break

    IT WAS REPORTED THAT THE LINER TRIAL BROKE INTRAOPERATIVELY. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8908279 ↗
  • MalfunctionJun 26, 2019Material Discolored; Naturally Worn; Scratched Material; Material Discolored; Naturally Worn; Scratched Material

    ADDITIONAL INFORMATION RECEIVED INDICATING THAT THE LOT NUMBERS COULD NOT BE IDENTIFIED BECAUSE THESE WERE SO WORN AND SCRATCHED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8736522 ↗
  • MalfunctionNov 21, 2018Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE PINNACLE TRIAL LINERS ARE WORN.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8094222 ↗
  • MalfunctionOct 30, 2018Naturally Worn; Scratched Material; Naturally Worn; Scratched Material

    COMPLAINT NOT SPECIFIC TO ONE CASE. SURGEONS COMPLAIN THAT PINNACLE POLY TRIALS ARE WORN AND USED, SCRATCHES ON THEM AND SCREWS WORN OUT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8021597 ↗