Skip to content

NA 2274-02-000Recalls & reported problems

Depuy (Ireland) · RATCHET SCREWDRIVER HANDLE · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy ratchet screwdriver handle is a surgical screwdriver used to tighten screws during orthopedic surgery. Many of its 300 reports describe the handle breaking or cracking, and the ratchet mechanism jamming or locking. The reports mention no deaths and 1 injury. Reports fell from 39 in the 12 months before to 17 in the latest 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
303
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 303 adverse event reports that list this exact device (DI 10603295109020), received between Jun 28, 2017 and Sep 29, 2026 — classified as 1 injury, 302 malfunctions.

The most frequently reported specific device problem is “Break” (112 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 3 reportsMar 2024: 1 reportsApr 2024: 3 reportsMay 2024: 7 reportsJun 2024: 3 reportsJul 2024: 0 reportsAug 2024: 3 reportsSep 2024: 1 reportsOct 2024: 4 reportsNov 2024: 0 reportsDec 2024: 5 reportsJan 2025: 1 reportsFeb 2025: 2 reportsMar 2025: 3 reportsApr 2025: 8 reportsMay 2025: 1 reportsJun 2025: 5 reportsJul 2025: 6 reportsAug 2025: 3 reportsSep 2025: 2 reportsOct 2025: 3 reportsNov 2025: 3 reportsDec 2025: 3 reportsJan 2026: 2 reportsFeb 2026: 1 reportsMar 2026: 2 reportsApr 2026: 2 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 5 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break112
  • Mechanical Jam76
  • Device-Device Incompatibility36
  • Unintended Movement23
  • Physical Resistance/Sticking19
  • Appropriate Device Problem Term/Code Not Available15
  • Detachment of Device or Device Component11
  • Material Deformation11

Patient problems

  • No Clinical Signs, Symptoms or Conditions149
  • No Consequences Or Impact To Patient111
  • Insufficient Information20
  • No Information17
  • Not Applicable9
  • No Code Available6
  • Foreign Body In Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 303 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Break; Mechanical Jam

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50936313 ↗
  • MalfunctionSep 14, 2026No Apparent Adverse Event

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50764867 ↗
  • MalfunctionSep 10, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50722046 ↗
  • MalfunctionSep 10, 2026Break

    It was reported that the grey Screw driver handle lock mechanism broke.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50714063 ↗
  • MalfunctionSep 2, 2026Mechanical Jam

    It was reported that the screwdriver handle no longer ratchets. It was reported that there was no delay in the procedure due to the event. There were no adverse consequences to the patient. No additional information could be provided.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50572008 ↗
  • MalfunctionMay 29, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE IT WAS OBSERVED THAT THE PINNACLE 25MM DRILL BIT WAS BLUNT, AND THE METAL RING FROM PINNACLE SCREWDRIVER HANDLE FROM PINNACLE SCREW TRAY WAS LOOSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25337081 ↗
  • MalfunctionMay 26, 2026Material Integrity Problem

    IT WAS REPORTED THAT THE T HANDLE HAS A CRACK IN IT AND THE SCREW DRIVER HANDLES COLLAR WAS DAMAGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25277084 ↗
  • MalfunctionApr 10, 2026Break

    IT WAS REPORTED THAT THE SCREWDRIVER HANDLE WAS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24845160 ↗
  • MalfunctionApr 7, 2026Break

    IT WAS REPORTED THAT DURING CUP IMPLANTATION TWO DRILL BITS, SCREWDRIVER AND A SCREWDRIVER HANDLE WERE BROKEN. NO PIECES WERE LOST. RATCHET HANDLE WORKED POORLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24808906 ↗
  • MalfunctionMar 4, 2026Break

    IT WAS REPORTED THAT RUBBER ON SCREWDRIVER IS TORN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24514232 ↗
  • MalfunctionMar 3, 2026Break; Device-Device Incompatibility; Material Deformation

    ADDITIONAL INFORMATION WAS RECEIVED PROVIDING ANSWERS FROM THE FOLLOWING QUESTIONS: 1. DID IT BREAK INTO TWO OR MORE PIECES? IF YES, WERE ALL THE FRAGMENTS RETRIEVED FROM THE PATIENT? IF NO, PLEASE SPECIFY WHAT IS THE ALLEGED DEFICIENCY, IS IT BENT, CRACKED, STRIPPED, SCRATCHED, WORN OR CROSS THREADED? - THE DRIVER DID NOT BREAK, IT IS STILL COMPLETE IN ONE PIECE. IT IS NOT BENT BUT POSSIBLE STRIPPED OR CROSS THREADED. IT IS THE TORQUE LOCKING MECHANISM THAT IS MALFUNCTIONING, WHILE ATTEMPTING TO TIGHTEN/LOOSEN A SCREW. ALL OF THE OTHER INFORMATION WAS ALREADY GIVEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24499784 ↗
  • MalfunctionFeb 20, 2026Break

    IT WAS REPORTED THAT THE SPRING BROKE OFF FROM THE SCREWDRIVER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24402783 ↗
  • MalfunctionJan 24, 2026Mechanical Jam

    IT WAS REPORTED THAT THE RATCHET FUNCTION OF THE SCREWDRIVER HANDLE WAS NO LONGER WORKING WHILE IMPLANTING A CANCELLOUS BONE SCREW. IT WAS REPORTED THAT SURGICAL TIME WAS NOT EXTENDED AND NO ADVERSE CONSEQUENCES FOR THE PATIENT OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24173644 ↗
  • MalfunctionJan 9, 2026Mechanical Jam

    IT WAS REPORTED THAT THE RATCHET HANDLE WOULD NOT RATCHET FORWARD AND ONLY MOVED BACKWARD, REVERSING THE SCREW. A BACKUP SET WAS OPENED AND THERE WAS NO DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24024432 ↗
  • MalfunctionDec 23, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT DURING STERILE PROCESSING IT WAS OBSERVED THAT THE SCREWDRIVER HANDLE HAD SOMETHING STUCK IN THE CANNULATED HOLE AND RP PS ARTIC SURFACE HAD A BROKEN PEG. NOTHING FURTHER TO REPORT. THE EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23882292 ↗
  • MalfunctionDec 3, 2025Break; Break

    IT WAS REPORTED THAT ONE RATCHET HANDLE FROM THE PINNACLE CORE SET WAS DAMAGED AND WAS BROKEN. THE FAILURE WAS NOT ATTRIBUTABLE TO WEAR AND TEAR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23708730 ↗
  • MalfunctionDec 2, 2025Device-Device Incompatibility; Mechanical Jam; Device-Device Incompatibility; Mechanical Jam

    PINNACLE RATCHET HANDLE WAS NOT WORKING AS IT SHOULD DURING THE CASE. UNABLE TO WIND FROM 'REVERSE' TO 'FORWARD' TO PLACE ACETABULAR SCREW AND RATCHET WAS NOT WORKING. ENDED UP HAVING TO PLACE SCREW MANUALLY WITH THE SCREWDRIVER. I TESTED THIS ONCE CASE WAS FINISHED, AND THE RATCHET HANDLE DOES NOT FUNCTION AS IT SHOULD. SURGEON HAS REQUESTED TO HAVE THIS REPLACED OUT WITH A NEW ONE IN THE SET (LIR TO BE COMPLETED BY MYSELF ONCE PC LODGED). CUSTOMER HAS REQUESTED TO HOLD THIS ITEM IN THE TRAY UNTIL A REPLACEMENT CAN BE SOURCED. WAS SURGERY DELAYED DUE TO THE REPORTED EVENT? NO, ACTION TAKEN WHEN EVENT OCCURRED? SURGEON MENTIONED TO ME THAT RATCHET HANDLE WAS NOT WORKING AS IT SHOULD AND REQUESTED THAT I SORT A REPLACEMENT., WAS PROCEDURE SUCCESSFULLY COMPLETED? YES, WERE FRAGMENTS GENERATED? NO, IF YES, WERE THEY REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION? UNKNOWN, PATIENT STATUS/ OUTCOME / CONSEQUENCES NO, WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED: NO, IS THE PATIENT PART OF A CLINICAL STUDY NO, DEVICE PROPERTY OF NONE, DEVICE IN POSSESSION OF NONE BY CHECKING THIS BOX I CERTIFY THAT ALL INFORMATION THAT ARE KNOW

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23691623 ↗
  • MalfunctionNov 18, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE TORQUE LIMITING FUNCTION ON A SCREWDRIVER WAS NOT WORKING DURING SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23583160 ↗
  • MalfunctionNov 10, 2025Break; Mechanical Jam; Break; Mechanical Jam

    IT WAS REPORTED THAT SCREWDRIVER HANDLE WAS NOT RATCHET, SOMETHING IN THE INNER DRIVER WAS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23513932 ↗
  • MalfunctionNov 6, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DIRECTION KNOB CAME APART.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23487569 ↗
  • MalfunctionOct 27, 2025Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT THE SCREWDRIVER WILL NOT RATCHET ANYMORE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23393464 ↗
  • MalfunctionOct 21, 2025Break; Break

    IT WAS REPORTED THAT THERE WAS A BROKEN LOCKING GUARD AT BASE OF SCREW HANDLE. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23349158 ↗
  • MalfunctionOct 16, 2025Break; Break

    IT WAS REPORTED THAT PATIENT UNDERWENT TOTAL HIP REPLACEMENT SURGERY. THE SURGEON HIT THE RATCHETING COLLAR ON THE SCREWDRIVER TO GET THE HANDLE OF THE SCREW AND BROKE THE COLLAR. THERE WAS NO DELAY IN CASE OR CONSEQUENCE TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23309904 ↗
  • MalfunctionSep 12, 2025Break; Break

    IT WAS REPORTED THE HUDSON/CHUCK IS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23042521 ↗
  • MalfunctionSep 4, 2025Break; Break

    IT WAS REPORTED THAT DURING SURGERY THAT THE SCREWDRIVER HANDLE WITH RATCHET WAS DAMAGED, BUT STILL IT WAS POSSIBLE TO SOLVE THE ISSUE BY SCREWING DIRECTLY WITH THE PIECE IN THE HAND AND USING A LARGE HANDLE THAT WAS INSIDE ANOTHER INSTRUMENT, THUS ACHIEVING TO FINISH THE SURGERY SUCCESSFULLY, WITHOUT AFFECTING THE PATIENT AND WITHOUT ALTERATIONS IN THE SURGICAL TIMES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22963761 ↗

Other SCREWDRIVER devices

All SCREWDRIVER devices →

Sources and updates