Skip to content

NA 2274-08-000Recalls & reported problems

Depuy (Ireland) · HEXAGON SCREWDRIVER SHAFT CARDAN · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Hexagon screwdriver shaft is a surgical instrument used to tighten screws during hip and other orthopedic procedures. Most of the 55 reports describe the tool breaking during surgery, with the tip snapping off so pieces had to be retrieved, and some mention it would not fit or work with other instruments. No deaths and no injuries are reported. There are no recalls naming this device, and reporting has been fairly steady in recent years.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
55
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 55 adverse event reports that list this exact device (DI 10603295109051), received between Jul 20, 2017 and Feb 9, 2026 — classified as 55 malfunctions.

The most frequently reported specific device problem is “Break” (31 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break31
  • Device-Device Incompatibility14
  • Material Deformation5
  • Material Twisted/Bent4
  • Entrapment of Device2
  • Failure to Cut1
  • Naturally Worn1
  • No Apparent Adverse Event1

Patient problems

  • No Clinical Signs, Symptoms or Conditions27
  • No Consequences Or Impact To Patient22
  • Insufficient Information3
  • No Code Available3
  • Not Applicable3
  • No Information1

Recent adverse event reports

The 25 most recent of 55 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 9, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THE DEVICE EXPERIENCED AN UNDETERMINED FAILURE DURING SURGERY. DURING MANUFACTURER'S INVESTIGATION OF THE RETURNED DEVICE IT WAS IDENTIFIED THAT THE WELD BETWEEN THE CARDAN AND HEX SWIVEL, BROKE, RESULTING IN THE HEX SWIVEL DETACHING FROM THE DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24291769 ↗
  • MalfunctionJun 10, 2025Break; Break

    IT WAS REPORTED THAT THE CARDAN DRIVER BROKE FOLLOWED BY KNUCKLE DRIVER BREAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22176931 ↗
  • MalfunctionJan 15, 2025Break; Break

    IT WAS REPORTED THAT THE DOCTOR WAS IMPLANTING A SCREW WITH THE CARDAN SCREWDRIVER ON POWER, AND WHILE IMPLANTING THE END OF THE SCREWDRIVER BROKE. THE DOCTOR HAD TO DRILL THE TIP OF THE SCREWDRIVER EXTENDING THE CASE 10 MINUTES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21158981 ↗
  • MalfunctionSep 16, 2024Break; Break

    IT WAS REPORTED THAT PATIENT HAD TOTAL HIP. ONE INSTRUMENT BROKE DURING THE CASE. THERE WAS BACKUPS AVAILABLE SO THERE WAS NO DELAYS IN SURGERY.

    Patient problems: Insufficient Information

    MAUDE report 20224294 ↗
  • MalfunctionAug 14, 2024Break; Break

    ADDITIONAL INFORMATION RECEIVED STATES THAT THERE WAS NO ADVERSE CONSEQUENCE AND ALL BROKEN PIECES WERE RETRIEVED FROM THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19981334 ↗
  • MalfunctionMar 5, 2024Break; Break

    IT WAS REPORTED THAT THE TIP OF PINNACLE SCREWDRIVER IS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18835533 ↗
  • MalfunctionJul 14, 2023Break; Break

    IT WAS REPORTED THAT SCREWDRIVER BROKE . SURGICAL DELAY UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17327523 ↗
  • MalfunctionMay 22, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE SCREWDRIVER RATCHET MECHANISM BROKE WHILE INSERTING THE SCREW. THE WELD BROKE ON THE CUP IMPACTOR HANDLE ABOVE THE GRIP, AND THE QUICK CONNECT END WOULDN¿T FIT ALL THE WAY INTO SCREWDRIVER HANDLE. NO DELAYS OR PATIENT HARM OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16975753 ↗
  • MalfunctionMay 1, 2023Break; Break

    IT WAS REPORTED THAT DURING CASE PHYSICIAN WAS TIGHTENING SCREW. IN THE PROCESS THE DRILL BIT BROKE AND WAS ATTACHED TO THE HEAD OF THE SCREW. IT TOOK THE PHYSICIAN TWENTY MINUTES TO REMOVE THE BROKEN HEX DRILL BIT. X RAYS WERE PERFORMED.

    Patient problems: Insufficient Information

    MAUDE report 16846069 ↗
  • MalfunctionJan 23, 2023Break; Break

    IT WAS REPORTED THAT THE SCREW DRIVER BROKE (TIP). NO DELAY IN THE CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16225023 ↗
  • MalfunctionJan 19, 2023Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    IT WAS REPORTED THAT THE SCREW BROKE AND STUCK ON DRIVER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16204896 ↗
  • MalfunctionJan 9, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT SCREWDRIVER ENDS STICKING IN SCREWS. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16122448 ↗
  • MalfunctionJan 9, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT SCREWDRIVER ENDS STICKING IN SCREWS. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16122397 ↗
  • MalfunctionJan 9, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT SCREWDRIVER ENDS STICKING IN SCREWS. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16122338 ↗
  • MalfunctionJan 9, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT SCREWDRIVER ENDS STICKING IN SCREWS. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16122321 ↗
  • MalfunctionJan 9, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT SCREWDRIVER ENDS STICKING IN SCREWS. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16122297 ↗
  • MalfunctionJan 9, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT SCREWDRIVER ENDS STICKING IN SCREWS. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16122288 ↗
  • MalfunctionJan 9, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT SCREWDRIVER ENDS STICKING IN SCREWS. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16122264 ↗
  • MalfunctionJan 9, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT SCREWDRIVER ENDS STICKING IN SCREWS. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16122240 ↗
  • MalfunctionDec 27, 2022Break; Break

    SCREWDRIVER BROKE WHEN PUTTING IN SCREW. NO SURGICAL DELAY. NO ADVERSE AFFECTS ON THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16056163 ↗
  • MalfunctionSep 12, 2022Break; Break

    IT WAS REPORTED THAT THE SWIVEL HEAD SHATTERED WHEN INSERTING SCREW.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15405572 ↗
  • MalfunctionJun 15, 2022Break; Entrapment of Device; Break; Entrapment of Device

    IT WAS REPORTED THAT THE TIP BROKE OFF THE DRIVER DURING SCREW INSERTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14695047 ↗
  • MalfunctionJun 3, 2022Break; Break

    IT WAS REPORTED THAT THE TIP OF THE SCREW DRIVER BROKE OFF. BOTH PIECE FOUND AND THROWN AWAY. NO DELAY IN SURGERY, NO OTHER INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14593407 ↗
  • MalfunctionDec 8, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE HEX HEAD IS DULL AND WILL NOT ARTICULATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12959602 ↗
  • MalfunctionOct 1, 2021Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE SCREWDRIVER TIP WOULD NOT SEAT INTO SCREW HEAD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12561241 ↗

Other SCREWDRIVER devices

All SCREWDRIVER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page