DELTA XTEND 2307-10-001Recalls & reported problems
Depuy (Ireland) · DELTA Xtend MODULAR HUMERAL STEM TRIAL DIA 10 MM · Class 1
All models of DELTA XTEND (194)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10603295115557), received between Apr 19, 2023 and Apr 19, 2023 — classified as 1 malfunction.
The most frequently reported specific device problem is “Device-Device Incompatibility” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device-Device Incompatibility1
- Material Deformation1
- Naturally Worn1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 19, 2023Device-Device Incompatibility; Material Deformation; Naturally Worn; Device-Device Incompatibility; Material Deformation; Naturally Worn
IT WAS REPORTED THAT THE HOSPITAL HAVE SOME DELTA XTEND INSTRUMENTS THAT NEED TO BE EXCHANGED DUE TO GENERAL WEAR AND TEAR, DETAILS ARE AS FOLLOWS: 3.5 MM CANNULATED HEX SCREWDRIVER. GENERAL WEAR AND TEAR OVER TIME, THE HEX IS NOW MISSHAPEN AND DUE TO IT BECOMING MORE ROUNDED, DOES NOT GRIP AND ENGAGE WITH THE SCREW SUFFICIENTLY TO TIGHTEN IT. MODULAR HUMERAL STEM TRIAL Ø 10 MM. THE PROXIMAL PART OF THE TRIAL STEM HAS BEEN IMPACTED OVER TIME CAUSING THE CURVATURE OF THE HOLE TO BE SLIGHTLY MISSHAPEN. THIS, HOWEVER, MAKES IT EXTREMELY DIFFICULT TO INSERT THE EPIPHYSIS ONTO THE TRIAL STEM. WITH A LOT OF FORCE, THE EPIPHYSIS CANNOT BE FULLY INSERTED, MEANING THAT IT DOES NOT SIT FLUSH AGAINST THE STEM. TRAILING IS THEREFORE INACCURATE. SIZE 2 LEFT EPI TRIAL. INSERTION POST DAMAGED TRYING TO INSERT INTO ABOVE BROKEN TRIAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16768802 ↗
Other TEMPLATE devices
- ATTUNE 2544-01-014Depuy (Ireland) · TEMPLATE802 reports
- ATTUNE 2545-00-546Depuy (Ireland) · TEMPLATE1 recall301 reports
- ARTICUL/EZE 2531-51-000Depuy (Ireland) · TEMPLATE323 reports
- DELTA XTEND 2307-95-000Depuy (Ireland) · TEMPLATE315 reports
- ATTUNE 2545-00-547Depuy (Ireland) · TEMPLATE1 recall231 reports
- ATTUNE 2544-00-520Depuy (Ireland) · TEMPLATE279 reports
- ATTUNE 2545-00-545Depuy (Ireland) · TEMPLATE1 recall221 reports
- ARTICUL/EZE 2530-91-000Depuy (Ireland) · TEMPLATE271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page