Skip to content

DELTA XTEND 2307-20-202Recalls & reported problems

Depuy (Ireland) · DELTA Xtend MODULAR ECCENTRIC EPIPHYSIS TRIAL SZ 2 LEFT · Class 1

All models of DELTA XTEND (194)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10603295115755), received between Mar 27, 2023 and Apr 18, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break1
  • Device-Device Incompatibility1
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 18, 2023Device-Device Incompatibility; Material Integrity Problem; Device-Device Incompatibility; Material Integrity Problem

    IT WAS REPORTED THAT THE HOSPITAL HAVE SOME DELTA XTEND INSTRUMENTS THAT NEED TO BE EXCHANGED DUE TO GENERAL WEAR AND TEAR, DETAILS ARE AS FOLLOWS: 3.5 MM CANNULATED HEX SCREWDRIVER: GENERAL WEAR AND TEAR OVER TIME, THE HEX IS NOW MISSHAPEN AND DUE TO IT BECOMING MORE ROUNDED, DOES NOT GRIP AND ENGAGE WITH THE SCREW SUFFICIENTLY TO TIGHTEN IT. MODULAR HUMERAL STEM TRIAL Ø 10 MM: THE PROXIMAL PART OF THE TRIAL STEM HAS BEEN IMPACTED OVER TIME CAUSING THE CURVATURE OF THE HOLE TO BE SLIGHTLY MISSHAPEN. THIS, HOWEVER, MAKES IT EXTREMELY DIFFICULT TO INSERT THE EPIPHYSIS ONTO THE TRIAL STEM. WITH A LOT OF FORCE, THE EPIPHYSIS CANNOT BE FULLY INSERTED, MEANING THAT IT DOES NOT SIT FLUSH AGAINST THE STEM. TRIALLING IS THEREFORE INACCURATE. SIZE 2 LEFT EPI TRIAL: INSERTION POST DAMAGED TRYING TO INSERT INTO ABOVE BROKEN TRIAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16760813 ↗
  • MalfunctionMar 27, 2023Break; Break

    IT WAS REPORTED THAT THIS DEVICE IS MISSING A SCREW.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16621237 ↗

Other TEMPLATE devices

All TEMPLATE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page