Skip to content

DELTA XTEND 2307-38-303Recalls & reported problems

Depuy (Ireland) · DELTA Xtend HIGH MOBILITY HUMERAL CUP TRIAL DIA 38 +3MM · Class 1

All models of DELTA XTEND (194)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10603295115786), received between Oct 29, 2022 and Jan 8, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Naturally Worn2

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 8, 2025Break; Break

    IT WAS REPORTED THAT THE LINER TRIAL HAS HOLES IN IT. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21103374 ↗
  • MalfunctionSep 22, 2023Break; Naturally Worn; Break; Naturally Worn

    ADDITIONAL INFORMATION RECEIVED: 1. WAS THE REPORTED EVENT ASSOCIATED WITH A REMOVAL/REVISION SURGERY? IF YES, PLEASE PROVIDE ALL RELEVANT DETAILS. NO 2. CAN YOU PLEASE CLARIFY WHAT YOU MEAN BY "DAMAGED"? WAS IT WORN, CRACKED, BROKEN INTO PIECES, BENT, STRIPPED, THREADED, OR ANY DEVICE INTERACTION? WORN AND CHIPPED AROUND THE EDGES OF THE LINER TRIAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17799819 ↗
  • MalfunctionOct 29, 2022Break; Naturally Worn; Break; Naturally Worn

    IT WAS REPORTED THAT THE POLY TRIAL HMO 38 + 3 PLASTIC IS PEELING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15695132 ↗

Other TEMPLATE devices

All TEMPLATE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page