Quick summary

This DePuy Delta Xtend screwdriver is a surgical instrument used during shoulder replacement surgery to tighten implant parts. Its 154 reports mostly describe worn parts, bent or broken tips, and handles sliding off, and most say no patient consequences were reported. The reports include no deaths and no injuries. No recalls name this device. Reports dropped from 5 in the year before to 1 in the latest year.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
154
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 154 adverse event reports that list this exact device (DI 10603295116585), received between Aug 25, 2017 and Nov 24, 2025 — classified as 154 malfunctions.

The most frequently reported specific device problem is “Naturally Worn” (63 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Naturally Worn63
  • Material Twisted/Bent49
  • Material Deformation27
  • Break23
  • Device-Device Incompatibility14
  • Appropriate Device Problem Term/Code Not Available7
  • Adverse Event Without Identified Device or Use Problem1
  • Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions75
  • No Consequences Or Impact To Patient69
  • No Information5
  • Not Applicable4
  • No Code Available3
  • Insufficient Information2

Recent adverse event reports

The 25 most recent of 154 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 24, 2025Break; Break

    IT WAS REPORTED THAT THE YELLOW HANDLE OF THE 3.5 CANNULATED HEX DRIVER SLID OFF THE SHAFT. NO DELAYS IN THE SURGERY OCCURRED, AND THE SURGEON WAS ABLE TO COMPLETE THE CASE WITH NO COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23630425 ↗
  • MalfunctionOct 28, 2025Break; Break

    IT WAS REPORTED THAT TIP BROKE OFF OF THE ITEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23402569 ↗
  • MalfunctionApr 29, 2025Break; Break

    IT WAS REPORTED THAT DURING INSERTION OF THE GLENOSPHERE THE INSTRUMENT´S TIP BROKE OFF INTRAOPERATIVELY. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21926710 ↗
  • MalfunctionApr 24, 2025Break; Break

    IT WAS REPORTED THAT THE SCREWDRIVER SEEMED TO BE COMING APART. THERE WAS NO SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21900459 ↗
  • MalfunctionFeb 5, 2025Break; Break

    IT WAS REPORTED THAT PATIENT UNDERWENT SHOULDER REPAIR ON (B)(6) 2025. THE SCREWDRIVER TIP BROKE OFF WHILE INSERTING THE GLENOSPHERE. ALL FRAGMENTS WERE REMOVED. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21316047 ↗
  • MalfunctionDec 2, 2024Break; Break

    IT WAS REPORTED THAT INTRAOPERATIVELY, THE INSTRUMENT FRACTURED INTO TWO OR MORE PIECES. NO ADVERSE PATIENT CONSEQUENCES, NO SURGICAL DELAY REPORTED. SALES REPRESENTATIVE CONFIRMED THAT THERE WAS MATERIAL DEFECT, BUT THE SCREWDRIVER WAS TILTED BY THE SALES REPRESENTATIVE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20822456 ↗
  • MalfunctionJan 4, 2024Break; Break

    IT WAS REPORTED THAT THE SCREWDRIVER TIP BROKE OFF WHEN TIGHTENING GLENOSOHERE TO BASEPLATE. ALL PARTS RECOVERED NO DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18448833 ↗
  • MalfunctionNov 28, 2023Break; Naturally Worn; Break; Naturally Worn

    THE CANNULATED SCREWDRIVER WAS RECEIVED AS BLIND UNIT HOWEVER, IT COULDN'T BE ASSOCIATED WITH A COMPLAINT OR ANY OTHER EXISTING RECORD. AS A RESULT, THIS COMPLAINT WAS CREATED TO ANALYZE THE RETURNED DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18219008 ↗
  • MalfunctionJun 5, 2023Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE SCREWDRIVER HANDLE WAS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17060317 ↗
  • MalfunctionApr 20, 2023Break; Break

    IT WAS REPORTED THAT THE YELLOW HANDLE BROKE UPON USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16777179 ↗
  • MalfunctionApr 18, 2023Device-Device Incompatibility; Material Deformation; Naturally Worn; Device-Device Incompatibility; Material Deformation; Naturally Worn

    IT WAS REPORTED THAT THE HOSPITAL HAVE SOME DELTA XTEND INSTRUMENTS THAT NEED TO BE EXCHANGED DUE TO GENERAL WEAR AND TEAR, DETAILS ARE AS FOLLOWS: 3.5 MM CANNULATED HEX SCREWDRIVER GENERAL WEAR AND TEAR OVER TIME, THE HEX IS NOW MISSHAPEN AND DUE TO IT BECOMING MORE ROUNDED, DOES NOT GRIP AND ENGAGE WITH THE SCREW SUFFICIENTLY TO TIGHTEN IT. MODULAR HUMERAL STEM TRIAL Ø 10 MM. THE PROXIMAL PART OF THE TRIAL STEM HAS BEEN IMPACTED OVER TIME CAUSING THE CURVATURE OF THE HOLE TO BE SLIGHTLY MISSHAPEN. THIS, HOWEVER, MAKES IT EXTREMELY DIFFICULT TO INSERT THE EPIPHYSIS ONTO THE TRIAL STEM. WITH A LOT OF FORCE, THE EPIPHYSIS CANNOT BE FULLY INSERTED, MEANING THAT IT DOES NOT SIT FLUSH AGAINST THE STEM. TRIALLING IS THEREFORE INACCURATE. SIZE 2 LEFT EPI TRIAL. INSERTION POST DAMAGED TRYING TO INSERT INTO ABOVE BROKEN TRIAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16760828 ↗
  • MalfunctionApr 12, 2023Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE HEX END HAS BECOME SLIGHTLY ROUNDED SO IT IS NOT AS EFFECTIVE AS IT SHOULD BE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16724342 ↗
  • MalfunctionMar 20, 2023Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE INHANCE REAMER SHAFT BROKE AT THE PLASTIC JUNCTION DURING NORMAL USE. THE YELLOW HEX SCREWDRIVER IS STRIPPED FROM NORMAL USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16577085 ↗
  • MalfunctionMar 16, 2023Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE TONG ON FORKED RETRACTOR BENT. THERE WAS NO BROKEN PIECES. NO ADVERSE AFFECTS ON THE PATIENT. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16555548 ↗
  • MalfunctionFeb 20, 2023Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE DELTA EXTEND SCREWDRIVER BECOME STRIPPED DURING SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16403694 ↗
  • MalfunctionJan 23, 2023Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE TIP OF THE DRIVER IS WORN OUT AND WOULD NOT WORK PROPERLY. CASE WAS NOT DELAYED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16226494 ↗
  • MalfunctionJan 18, 2023Break; Break

    ADDITIONAL INFORMATION RECEIVED. INDICATES, THAT THERE WERE NO ADVERSE CONSEQUENCES THAT WOULD AFFECT THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16196279 ↗
  • MalfunctionDec 9, 2022Material Deformation; Material Deformation

    IT WAS REPORTED THAT A 3.5 CANNULATED SCREWDRIVER TIP WAS BENT. NO ADDITIONAL SURGERY TIME WAS REQUIRED. NO ADDITIONAL INFORMATION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15949161 ↗
  • MalfunctionDec 7, 2022Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE TIP OF THE DRIVER IS WORN OUT AND NOT CATCHING SCREWS. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15931553 ↗
  • MalfunctionNov 22, 2022Break; Break

    IT WAS REPORTED THAT THE SCREWDRIVER TIP BROKE WHILE TIGHTENING GLENOSPHERE INTO COLLET. THERE WAS A SURGICAL DELAY OF 10 MINS.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 15842609 ↗
  • MalfunctionOct 31, 2022Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE DRIVER TIP IS WORN DUE TO LOTS OF USE. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15700250 ↗
  • MalfunctionOct 26, 2022Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE DR. COMPLAINED TIP OF SCREW DRIVER WAS TWISTED AND NEEDING REPLACEMENT. PLEASE SEND A REPLACEMENT TO OUR KNOXVILLE OFFICE. SURGICAL DELAY WAS REPORTED UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15674265 ↗
  • MalfunctionOct 18, 2022Naturally Worn; Naturally Worn

    DELTA XTEND YELLOW SCREW DRIVER HAS BECOME WORN AND IS NO LONGER FUNCTIONING AS USUAL. HAS SLOWLY WORSENED OVER TIME. NO ADVERSE EFFECTS TO THE PATIENT. NO DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15625547 ↗
  • MalfunctionOct 6, 2022Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE CANNULATED HEX SCREWDRIVER HAS THE EDGES ROUNDED OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15551154 ↗
  • MalfunctionSep 29, 2022Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE DRIVER IS WORN FROM EXCESSIVE WEAR. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15508539 ↗