S-ROM 2365-31-500Recalls & reported problems
Depuy (Ireland) · S-ROM TRIAL HEAD 36mm +0 · Class 1
All models of S-ROM (185)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10603295120032), received between Aug 21, 2017 and Apr 20, 2026 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Scratched Material” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Scratched Material3
- Device-Device Incompatibility2
- Appropriate Device Problem Term/Code Not Available1
- Illegible Information1
- Naturally Worn1
- No Apparent Adverse Event1
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient3
- No Information1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 20, 2026No Apparent Adverse Event
IT WAS REPORTED THAT PATIENT UNDERWENT TOTAL HIP REPLACEMENT. PRODUCT HAD REACHED END OF LIFE DUE TO WEAR AND TEAR. THE SURGEONS WERE COMPLAINING ABOUT INSTRUMENT BEING DULL, ILLEGIBLE, AND GENERALLY WORN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24926354 ↗MalfunctionFeb 16, 2023Device-Device Incompatibility; Illegible Information; Device-Device Incompatibility; Illegible Information
IT WAS REPORTED THE TRIALS WOULD NOT STAY ON THE NECK TRIAL. NO SURGICAL DELAY WAS THERE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16385846 ↗MalfunctionJan 27, 2023Naturally Worn; Naturally Worn
IT WAS REPORTED THAT THE TRIAL HEADS SHOW SIGNIFICANT WEAR DUE TO MULTIPLE USES AND NEED TO BE REPLACED. THE SALES REP BECAME AWARE OF THIS ON (B)(6) 2023, WHILE SETTING UP FOR THE PROCEDURE. FOUND TEMPORARY REPLACEMENTS FOR THIS SURGERY. LOT NUMBER COULD NOT BE READ ON THE 32MM +0. NO SURGICAL DELAY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16253497 ↗MalfunctionDec 8, 2022Device-Device Incompatibility; Unintended Movement; Device-Device Incompatibility; Unintended Movement
IT WAS REPORTED THAT ON THE 36+12 SROM NECK TRIAL, THE SCREW-IN PIECE THAT CONNECTS TO THE DISTAL STEM STRIPPED. ALSO, THE 36+0 HEAD TRIAL WAS VERY LOOSE ON THE TRIAL NECK TRUNNION. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15940618 ↗MalfunctionJan 2, 2020Scratched Material; Scratched Material
MalfunctionAug 2, 2019Scratched Material; Scratched Material
IT WAS REPORTED THAT DEEP CUTS AND GOUGES FOUND WHEN SET RETURNED TO THE OFFICE FOR INSPECTION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8855380 ↗MalfunctionAug 8, 2018Scratched Material; Scratched Material
GENERAL WEAR, SCRATCHES ON HEAD MAKING IT DIFFICULT TO READ.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7761444 ↗MalfunctionAug 21, 2017Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT HEAD TRIALS WERE FOUND TO HAVE DEEP GOUGES UPON POST SURGICAL OBSERVATION. IT WAS ALSO STATED THAT THE SURFACES ARE DAMAGED.
Patient problems: No Information
MAUDE report 6808253 ↗
Other TEMPLATE devices
- ATTUNE 2544-01-014Depuy (Ireland) · TEMPLATE802 reports
- ATTUNE 2545-00-546Depuy (Ireland) · TEMPLATE1 recall301 reports
- ARTICUL/EZE 2531-51-000Depuy (Ireland) · TEMPLATE323 reports
- DELTA XTEND 2307-95-000Depuy (Ireland) · TEMPLATE315 reports
- ATTUNE 2545-00-547Depuy (Ireland) · TEMPLATE1 recall231 reports
- ATTUNE 2544-00-520Depuy (Ireland) · TEMPLATE279 reports
- ATTUNE 2545-00-545Depuy (Ireland) · TEMPLATE1 recall221 reports
- ARTICUL/EZE 2530-91-000Depuy (Ireland) · TEMPLATE271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page