MORELAND 2422-10-000Recalls & reported problems

Depuy Orthopaedics, Inc. · MORELAND REV INST THIN OSTEOTOME HANDLE · Class 1

All models of MORELAND (22)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Moreland flexible osteotome handle is a surgical instrument handle used in joint replacement procedures. Most of the 53 reports describe the handle or its locking screw breaking or snapping during use, sometimes cracking or coming apart where the blade is inserted, with a few mentioning worn or bent parts and mismatched equipment. No deaths or injuries are reported, and many reports note no effect on the patient. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
53
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 53 adverse event reports that list this exact device (DI 10603295123071), received between Sep 8, 2017 and Mar 6, 2026 — classified as 53 malfunctions.

The most frequently reported specific device problem is “Break” (40 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break40
  • Device-Device Incompatibility9
  • Naturally Worn4
  • Material Deformation3
  • Material Twisted/Bent3
  • Appropriate Device Problem Term/Code Not Available2
  • Scratched Material2
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions25
  • No Consequences Or Impact To Patient22
  • Insufficient Information3
  • No Information2
  • Not Applicable1

Recent adverse event reports

The 25 most recent of 53 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 6, 2026Break

    IT WAS REPORTED THAT SCREW THAT LOCKS DOWN BLADES IN THE MORELAND FLEXIBLE OSTEOTOME HANDLE SHEARED IN HALF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24532298 ↗
  • MalfunctionJan 7, 2026Break

    IT WAS REPORTED THAT LOCKING SCREW FROM FLEXIBLE OSTEOTOME HANDLE BROKE. THE SET SCREW THAT HOLDS THE BLADE BROKE IN 2 PIECES. NO FRAGMENTS GENERATED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23994043 ↗
  • MalfunctionSep 4, 2025Break; Break

    IT WAS REPORTED THAT OSTEOTOME HANDLE BROKE. SCREW BROKE OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22963813 ↗
  • MalfunctionMar 17, 2025Break; Break

    IT WAS REPORTED THAT TIBIA EXTRACTOR BROKE WHILE REMOVING THE TIBIA. THE FEMORAL EXTRACTOR WON¿T GRIP SIDES OF FEMUR ANYMORE. THE CLAWS WERE WORN DOWN. OSTETOME HANDLE BROKE APART WHERE BLADE IS INSERTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21620741 ↗
  • MalfunctionMar 17, 2025Break; Break

    IT WAS REPORTED THAT TIBIA EXTRACTOR BROKE WHILE REMOVING THE TIBIA DURING A TOTAL KNEE REPLACEMENT. FEMORAL EXTRACTOR WON¿T GRIP SIDES OF FEMUR ANYMORE AS THE CLAWS ARE WORN DOWN. OSTEOTOME HANDLE BROKE APART WHERE BLADE IS INSERTED. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21620210 ↗
  • MalfunctionFeb 20, 2025Break; Break

    ADDITIONAL INFORMATION WAS RECEIVED: THERE WERE NO FRAGMENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21422010 ↗
  • MalfunctionJan 31, 2025Crack; Crack

    IT WAS REPORTED THAT HOSPITAL SHOWED SALES REP THINS HANDLE HAS A CRACK IN HANDLE

    Patient problems: Insufficient Information

    MAUDE report 21283323 ↗
  • MalfunctionNov 14, 2024Break; Break

    IT WAS REPORTED DURING A TOTAL HIP PROCEDURE ON (B)(6) 2024 THE SCREW IN THE INSTRUMENT HAS BROKEN OFF DURING USE. NO PATIENT CONSEQUENCE. NO KNOWN SURGICAL DELAY

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20685942 ↗
  • MalfunctionOct 15, 2024Break; Material Deformation; Naturally Worn; Scratched Material; Break; Material Deformation; Naturally Worn; Scratched Material

    IT WAS REPORTED THAT ON (B)(6) 2024, THERE WAS A BROKEN AND WORN PINNACLE 32MM +4/10 DEGREE LINER TRIAL. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20451963 ↗
  • MalfunctionSep 30, 2024Break; Break

    IT WAS REPORTED THAT (B)(6) 2024, THE CORAIL EXTRACTION OSTEOTOME HANDLE SCREW BROKE IN HANDLE DURING USE. THE FRAGMENTS WERE GENERATED AND EASILY REMOVED. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20333286 ↗
  • MalfunctionSep 4, 2024Break; Break

    IT WAS REPORTED THAT THE SCREW THREAD SHEARED OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20139535 ↗
  • MalfunctionJul 23, 2024Break; Break

    IT WAS REPORTED THAT SCREW BROKE, ALL PIECES WERE FOUND AND THERE WAS NO DELAYS IN CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19805067 ↗
  • MalfunctionApr 9, 2024Break; Break

    IT WAS REPORTED THAT THE OSTEOTOME HANDLE BROKE. THIS OCCURRED DURING SURGERY AND NO PATIENT WAS AFFECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19070312 ↗
  • MalfunctionMar 15, 2024Device-Device Incompatibility; Naturally Worn; Device-Device Incompatibility; Naturally Worn

    IT WAS REPORTED THAT THE THREAD HAS WORN AND WOULDN¿T TIGHTEN THE OSTEOTOME INTO PLACE. DID THE EVENT HAPPEN DURING A PROCEDURE? YES. WERE YOU IN THE PROCEDURE AT THE TIME OF THE EVENT? NO. EVENT OUTCOME/HOW WAS IT MANAGED? NA. WAS THERE ANY CONSEQUENCE TO THE PATIENT DUE TO COMPLAINT? NO. WAS THE SURGERY PROLONGED DUE TO THE EVENT? IF YES, CONFIRM HOW MANY MINUTES DELAY: YES. NO. HAS THE REPORTER FACILITY INDICATED THERE MAY BE LEGAL ACTION? NO. IS THE PRODUCT AVAILABLE FOR RETURN? NO. PLEASE GIVE A DETAILED EXPLANATION OF THE EVENT. THE THREAD HAS WORN AND WOULDN¿T TIGHTEN THE OSTEOTOME INTO PLACE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18912994 ↗
  • MalfunctionSep 22, 2023Break; Break

    IT WAS REPORTED THAT DEVICE BROKE DURING USE. SURGICAL TIME WAS NOT PROLONGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17799384 ↗
  • MalfunctionSep 7, 2023Break; Break

    IT WAS REPORTED THAT THE OSTEOTOME HANDLE WAS BROKEN UPON RETURN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17700259 ↗
  • MalfunctionMar 28, 2023Break; Structural Problem; Break; Structural Problem

    IT WAS REPORTED THAT THE LOCKING SCREW TO HOLD THE OSTEOTOME IN PLACE AS BECOME INEFFECTIVE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16626905 ↗
  • MalfunctionNov 29, 2022Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE HANDLE DOESN¿T HOLD THE OSTEOTOME BLADES SECURELY, LOCKING MECHANISM IS LOCKED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15878501 ↗
  • MalfunctionMar 24, 2022Break; Break

    THE HANDLE HOLDING A FLEXIBLE OSTEOTOME BROKE DURING SURGICAL PROCEDURE. THE SURGEON WAS ABLE TO COMPLETE THE CASE USING AN ALTERNATE OSTEOTOME AND HANDLE. NO SURGICAL DELAY. NO ADVERSE AFFECTS ON PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13892835 ↗
  • MalfunctionMar 15, 2022Break; Break

    IT WAS REPORTED THAT WHEN USING THE FLEXIBLE OSTEOTOMES THE BOLT THAT SECURES THE BLADE TO HANDLE BROKE AND HANDLE CAME APART. SURGICAL DELAY OF 30 MINUTES.

    Patient problems: Insufficient Information

    MAUDE report 13770188 ↗
  • MalfunctionFeb 4, 2022Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON THE FLEXIBLE OSTEOTOME HANDLE - SCREW HAD SEIZED IN PLACE. DID NOT HOLD OSTEOTOME WELL.

    Patient problems: Insufficient Information

    MAUDE report 13452989 ↗
  • MalfunctionSep 15, 2021Break; Break

    COMPASS CASE (B)(4) REPORT: IT WAS REPORTED THAT WHILE TRYING TO REMOVE THE TIBIA WITH THE FLEXIBLE OSTEOTOME, THE HANDLE BROKE AFTER MULTIPLE HITS FROM THE MALLET. THE SCREW THAT HOLDS THE OSTEOTOME ON BROKE. THIS DID NOT SLOW DOWN THE SURGERY AND NO PARTS WERE LEFT IN THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12476123 ↗
  • MalfunctionSep 14, 2021Break; Break

    IT WAS REPORTED THAT THE SET SCREW HAS BROKEN OFF. INSTRUMENT IS DEFECTIVE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12467506 ↗
  • MalfunctionJul 22, 2021Break; Break

    IT WAS REPORTED THAT THE INSTRUMENT IS TAGGED AS DAMAGED WITH A BROKEN OSTEOTOME HANDLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12214703 ↗
  • MalfunctionApr 16, 2021Break; Break

    IT WAS REPORTED THAT THERE WAS A BROKEN OSTEOTOME HANDLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11678560 ↗