QUICKSET 2440-00-549Recalls & reported problems

Depuy (Ireland) · QUICKSET GRATER HEAD 49mm · Class 1

All models of QUICKSET (66)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy QuickSet grater head is a reamer attachment used in surgery. Most of its 1,113 reports describe the grater failing to cut, with staff finding it too dull or chipped during reprocessing. The reports include no deaths and no injuries, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,113
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,113 adverse event reports that list this exact device (DI 10603295123965), received between Aug 16, 2017 and Apr 6, 2026 — classified as 1,113 malfunctions.

The most frequently reported specific device problem is “Failure to Cut” (833 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 18 reportsDec 2023: 18 reportsJan 2024: 10 reportsFeb 2024: 52 reportsMar 2024: 15 reportsApr 2024: 25 reportsMay 2024: 34 reportsJun 2024: 16 reportsJul 2024: 15 reportsAug 2024: 13 reportsSep 2024: 19 reportsOct 2024: 1 reportsNov 2024: 5 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 7 reportsJun 2025: 4 reportsJul 2025: 8 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 5 reportsFeb 2026: 6 reportsMar 2026: 0 reportsApr 2026: 10 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202652
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Cut833
  • Dull, Blunt147
  • Naturally Worn129
  • No Apparent Adverse Event48
  • Corroded43
  • Material Twisted/Bent34
  • Break31
  • Adverse Event Without Identified Device or Use Problem16

Patient problems

  • No Clinical Signs, Symptoms or Conditions868
  • No Consequences Or Impact To Patient182
  • Insufficient Information57
  • No Information20
  • No Code Available16
  • Not Applicable6

Recent adverse event reports

The 25 most recent of 1,113 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT DURING MEDICAL DEVICE REPROCESSING THE STAFF NOTICED THAT THE GRATERS ARE TOO DULL AND CHIPPED TO BE USED ON A PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24793681 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT DURING MEDICAL DEVICE REPROCESSING THE STAFF NOTICED THAT THESE GRATERS ARE TOO DULL AND CHIPPED TO BE USED ON A PATIENT. THE SUMMIT BROACH HANDLE CLASPS NO LONGER TIGHTEN SUFFICIENTLY ON THE BROACHES. THE EXCEL T HANDLES ARE CRACKED ON THE HANDLE. THE CUP INSURERS ARE CROSS THREADED. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24791313 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT DURING MEDICAL DEVICE REPROCESSING THE STAFF NOTICED THAT THESE GRATERS ARE TOO DULL AND CHIPPED TO BE USED ON A PATIENT. THE SUMMIT BROACH HANDLE CLASPS NO LONGER TIGHTEN SUFFICIENTLY ON THE BROACHES. THE EXCEL T HANDLES ARE CRACKED ON THE HANDLE. THE CUP INSURERS ARE CROSS THREADED. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24791312 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT DURING MEDICAL DEVICE REPROCESSING THE STAFF NOTICED THAT THESE GRATERS ARE TOO DULL AND CHIPPED TO BE USED ON A PATIENT. THE SUMMIT BROACH HANDLE CLASPS NO LONGER TIGHTEN SUFFICIENTLY ON THE BROACHES. THE EXCEL T HANDLES ARE CRACKED ON THE HANDLE. THE CUP INSURERS ARE CROSS THREADED. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24791311 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT DURING MEDICAL DEVICE REPROCESSING THE STAFF NOTICED THAT THESE GRATERS ARE TOO DULL AND CHIPPED TO BE USED ON A PATIENT. THE SUMMIT BROACH HANDLE CLASPS NO LONGER TIGHTEN SUFFICIENTLY ON THE BROACHES. THE EXCEL T HANDLES ARE CRACKED ON THE HANDLE. THE CUP INSURERS ARE CROSS THREADED. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24791306 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT DURING MEDICAL DEVICE REPROCESSING THE STAFF NOTICED THAT THESE GRATERS ARE TOO DULL AND CHIPPED TO BE USED ON A PATIENT. THE SUMMIT BROACH HANDLE CLASPS NO LONGER TIGHTEN SUFFICIENTLY ON THE BROACHES. THE EXCEL T HANDLES ARE CRACKED ON THE HANDLE. THE CUP INSURERS ARE CROSS THREADED. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24791302 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT DURING MEDICAL DEVICE REPROCESSING, THE STAFF NOTICED THAT THESE GRATERS ARE TOO DULL AND CHIPPED TO BE USED ON A PATIENT. THE SUMMIT BROACH HANDLE CLASPS NO LONGER TIGHTEN SUFFICIENTLY ON THE BROACHES. THE EXCEL T HANDLES ARE CRACKED ON THE HANDLE. THE CUP INSURERS ARE CROSS THREADED. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24791286 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT DURING MEDICAL DEVICE REPROCESSING THE STAFF NOTICED THAT THESE GRATERS ARE TOO DULL AND CHIPPED TO BE USED ON A PATIENT. THE SUMMIT BROACH HANDLE CLASPS NO LONGER TIGHTEN SUFFICIENTLY ON THE BROACHES. THE EXCEL T HANDLES ARE CRACKED ON THE HANDLE. THE CUP INSURERS ARE CROSS THREADED. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24791271 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT DURING MEDICAL DEVICE REPROCESSING, THE STAFF NOTICED THAT THESE GRATERS ARE TOO DULL AND CHIPPED TO BE USED ON A PATIENT. THE SUMMIT BROACH HANDLE CLASPS NO LONGER TIGHTEN SUFFICIENTLY ON THE BROACHES. THE EXCEL T-HANDLES ARE CRACKED ON THE HANDLE. THE CUP INSURERS ARE CROSS THREADED. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24791243 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT DURING MEDICAL DEVICE REPROCESSING, THE STAFF NOTICED THAT THESE GRATERS ARE TOO DULL AND CHIPPED TO BE USED ON A PATIENT. THE SUMMIT BROACH HANDLE CLASPS NO LONGER TIGHTEN SUFFICIENTLY ON THE BROACHES. THE EXCEL T-HANDLES ARE CRACKED ON THE HANDLE. THE CUP INSURERS ARE CROSS THREADED. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24791221 ↗
  • MalfunctionFeb 3, 2026Failure to Cut; No Apparent Adverse Event

    IT WAS REPORTED THAT ALL GRATERS ARE DULL AND NEED REPLACED OR SHARPENED. WILL SEND BACK AS SOON AS I RECEIVE SHARP REPLACEMENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24244234 ↗
  • MalfunctionFeb 3, 2026Failure to Cut; No Apparent Adverse Event

    IT WAS REPORTED THAT ALL GRATERS ARE DULL AND NEED REPLACED OR SHARPENED. WILL SEND BACK AS SOON AS I RECEIVE SHARP REPLACEMENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24244227 ↗
  • MalfunctionFeb 2, 2026Failure to Cut; No Apparent Adverse Event

    IT WAS REPORTED THE QUICKSET ACE GRATER HEADS WERE NO LONGER SHARP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24236064 ↗
  • MalfunctionFeb 2, 2026Failure to Cut; No Apparent Adverse Event

    IT WAS REPORTED THE QUICKSET ACE GRATER HEADS WERE NO LONGER SHARP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24236058 ↗
  • MalfunctionFeb 2, 2026Failure to Cut; No Apparent Adverse Event

    IT WAS REPORTED THE QUICKSET ACE GRATER HEADS WERE NO LONGER SHARP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24236006 ↗
  • MalfunctionFeb 2, 2026Failure to Cut; No Apparent Adverse Event

    IT WAS REPORTED THAT SURGEON HAS COMPLAINED THAT THE ACETABULAR REAMERS WERE NOT SHARP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24233918 ↗
  • MalfunctionJan 28, 2026Failure to Cut

    IT WAS REPORTED THAT THE ACETABULAR GRATERS IN THE HOSPITALS SETS ARE IN NEED OF SHARPENING AND THEY ARE NO LONGER FUNCTIONING AS EXPECTED. NO KNOWN PATIENT HARM THE FOLLOWING COMPLAINTS ARE LINKED TO CAPTURE ALL IMPACTED PRODUCTS: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24202473 ↗
  • MalfunctionJan 28, 2026Failure to Cut

    IT WAS REPORTED THAT THE ACETABULAR GRATERS IN THE HOSPITALS SETS ARE IN NEED OF SHARPENING AND THEY ARE NO LONGER FUNCTIONING AS EXPECTED. NO KNOWN PATIENT HARM. THE FOLLOWING COMPLAINTS ARE LINKED TO CAPTURE ALL IMPACTED PRODUCTS: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24198894 ↗
  • MalfunctionJan 28, 2026Failure to Cut

    IT WAS REPORTED THAT THE ACETABULAR GRATERS IN THE HOSPITALS SETS ARE IN NEED OF SHARPENING AND THEY ARE NO LONGER FUNCTIONING AS EXPECTED. NO KNOWN PATIENT HARM. THE FOLLOWING COMPLAINTS ARE LINKED TO CAPTURE ALL IMPACTED PRODUCTS: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24198886 ↗
  • MalfunctionJan 28, 2026Failure to Cut

    IT WAS REPORTED THAT THE ACETABULAR GRATERS IN THE HOSPITALS SETS ARE IN NEED OF SHARPENING AND THEY ARE NO LONGER FUNCTIONING AS EXPECTED. NO KNOWN PATIENT HARM THE FOLLOWING COMPLAINTS ARE LINKED TO CAPTURE ALL IMPACTED PRODUCTS: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24198876 ↗
  • MalfunctionJan 28, 2026Failure to Cut

    IT WAS REPORTED THAT THE ACETABULAR GRATERS IN THE HOSPITALS SETS ARE IN NEED OF SHARPENING AND THEY ARE NO LONGER FUNCTIONING AS EXPECTED. NO KNOWN PATIENT HARM THE FOLLOWING COMPLAINTS ARE LINKED TO CAPTURE ALL IMPACTED PRODUCTS: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24198873 ↗
  • MalfunctionNov 25, 2025Break; Break

    IT WAS REPORTED THAT THERE WAS A BROKEN GRATER WHICH WAS NOTICED BY STERILE PROCESSING DEPARTMENT. NO PATIENT WAS AFFECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23644839 ↗
  • MalfunctionJul 15, 2025Corroded; Material Deformation; Corroded; Material Deformation

    IT WAS REPORTED THAT OLD PINNACLE GRATERS NEED TO BE REPLACED. THEY ARE REPORTEDLY DULL WITH RUST AND WARPING DURING STERILE PROCESSING THESE ARE OVER 10 YEARS OLD. THE ISSUES DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22511237 ↗
  • MalfunctionJul 15, 2025Corroded; Material Deformation; Corroded; Material Deformation

    IT WAS REPORTED THAT OLD PINNACLE GRATERS NEED TO BE REPLACED. THEY ARE REPORTEDLY DULL WITH RUST AND WARPING DURING STERILE PROCESSING THESE ARE OVER 10 YEARS OLD. THE ISSUES DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22511223 ↗
  • MalfunctionJul 15, 2025Corroded; Material Deformation; Corroded; Material Deformation

    IT WAS REPORTED THAT OLD PINNACLE GRATERS NEED TO BE REPLACED. THEY ARE REPORTEDLY DULL WITH RUST AND WARPING DURING STERILE PROCESSING THESE ARE OVER 10 YEARS OLD. THE ISSUES DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22511211 ↗