ELITE 2522-22-001Recalls & reported problems
Depuy International Ltd · ELITE IMPL TRIAL HEAD 22.225MM -3 · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10603295128489), received between Aug 31, 2020 and Aug 31, 2020 — classified as 1 malfunction.
The most frequently reported specific device problem is “Contamination /Decontamination Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Contamination /Decontamination Problem1
Patient problems
- No Code Available1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 31, 2020Contamination /Decontamination Problem; Contamination /Decontamination Problem
THE PRIMARY SURGERY WAS PERFORMED ON (B)(6) 2020 VIA THA. BEFORE THE SURGERY, THE HOSPITAL STAFF STERILIZED THE DEVICES BY CLEANING AND DRYING WITH A MACHINE. IT WAS REPORTED THAT AFTER STERILIZATION, THE STAFF SET OUT THE DEVICES FOR THE SURGERY, AND HE/SHE FOUND THAT WATER REMAINED IN THE 9 TRIAL HEADS WHEN HE/SHE PULLED OUT THE DEVICES FROM THE CASE. HE/SHE HAD CHECKED THAT THERE WAS NO WATER IN THE TRIAL HEADS WHEN SET THEM IN THE DRYING MACHINE. THE STAFF JUDGED POOR STERILIZATION BECAUSE WATER WAS REMAINED IN THE DEVICES DESPITE STERILIZED BY AUTOCLAVE, AND HE/SHE STERILIZED THE DEVICES AGAIN. PREVENTING A RECURRENCE OF WATER¿S REMAINING, THE TRIAL HEADS WERE STERILIZED INTO SMALL GROUPS. THE REMAINING OF WATER WAS FOUND AT THE LAST MINUTE OF THE SURGERY, BUT BARELY MADE USING IN THE SURGERY. THE SURGERY WAS COMPLETED WITHOUT ANY SURGICAL DELAY. THERE WAS NO HARM TO THE PATIENT.
Patient problems: No Code Available; No Consequences Or Impact To Patient
MAUDE report 10472728 ↗
Other TEMPLATE devices
- ATTUNE 2544-01-014Depuy (Ireland) · TEMPLATE802 reports
- ATTUNE 2545-00-546Depuy (Ireland) · TEMPLATE1 recall301 reports
- ARTICUL/EZE 2531-51-000Depuy (Ireland) · TEMPLATE323 reports
- DELTA XTEND 2307-95-000Depuy (Ireland) · TEMPLATE315 reports
- ATTUNE 2545-00-547Depuy (Ireland) · TEMPLATE1 recall231 reports
- ATTUNE 2544-00-520Depuy (Ireland) · TEMPLATE279 reports
- ATTUNE 2545-00-545Depuy (Ireland) · TEMPLATE1 recall221 reports
- ARTICUL/EZE 2530-91-000Depuy (Ireland) · TEMPLATE271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page