ARTICUL/EZE 2531-53-000Recalls & reported problems

Depuy (Ireland) · ARTICUL/EZE TRIAL HEAD 36mm +8.5 · Class 1

All models of ARTICUL/EZE (26)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy ARTICUL/EZE trial head is a surgical instrument used during hip replacement. Many of the 136 reports describe instruments scratched, cracked or worn out, and some mention trial heads that would not stay on the neck. There are no deaths and 1 injury. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
136
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 136 adverse event reports that list this exact device (DI 10603295129202), received between Dec 13, 2017 and Feb 3, 2026 — classified as 1 injury, 135 malfunctions.

The most frequently reported specific device problem is “Scratched Material” (82 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 0 reportsJan 2024: 3 reportsFeb 2024: 10 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 2 reportsJun 2024: 4 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 4 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 5 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202610
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Scratched Material82
  • Naturally Worn49
  • Break25
  • Material Deformation18
  • Material Twisted/Bent9
  • Crack8
  • Device-Device Incompatibility8
  • Peeled/Delaminated7

Patient problems

  • No Clinical Signs, Symptoms or Conditions92
  • No Consequences Or Impact To Patient38
  • Insufficient Information5
  • No Code Available3
  • Bone Fracture(s)1
  • Not Applicable1

Recent adverse event reports

The 25 most recent of 136 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 3, 2026Crack; Material Deformation

    IT WAS REPORTED THAT SEVERAL INSTRUMENTS HAVE BEEN DEEMED NO LONGER USABLE DUE TO BEING CRACKED, CROSS THREADED WITH WEAR AND TEAR OVER TIME. THERE WAS NO PATIENT HARM AND NO SURGICAL DELAY REPORTED. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24247437 ↗
  • MalfunctionFeb 3, 2026Crack; Material Deformation

    IT WAS REPORTED THAT SEVERAL INSTRUMENTS HAVE BEEN DEEMED NO LONGER USABLE DUE TO BEING CRACKED, CROSS THREADED WITH WEAR AND TEAR OVER TIME. THERE WAS NO PATIENT HARM AND NO SURGICAL DELAY REPORTED. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24247436 ↗
  • MalfunctionFeb 3, 2026Crack; Material Deformation

    IT WAS REPORTED THAT SEVERAL INSTRUMENTS HAVE BEEN DEEMED NO LONGER USABLE DUE TO BEING CRACKED, CROSS THREADED WITH WEAR AND TEAR OVER TIME. THERE WAS NO PATIENT HARM AND NO SURGICAL DELAY REPORTED. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24247177 ↗
  • MalfunctionFeb 3, 2026Crack; Material Deformation

    IT WAS REPORTED THAT SEVERAL INSTRUMENTS HAVE BEEN DEEMED NO LONGER USABLE DUE TO BEING CRACKED, CROSS THREADED WITH WEAR AND TEAR OVER TIME. THERE WAS NO PATIENT HARM AND NO SURGICAL DELAY REPORTED. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24247173 ↗
  • MalfunctionFeb 3, 2026Crack; Material Deformation

    IT WAS REPORTED THAT SEVERAL INSTRUMENTS HAVE BEEN DEEMED NO LONGER USABLE DUE TO BEING CRACKED, CROSS THREADED WITH WEAR AND TEAR OVER TIME. THERE WAS NO PATIENT HARM AND NO SURGICAL DELAY REPORTED. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24247115 ↗
  • MalfunctionDec 3, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT DURING THE HIP REPLACEMENT PROCEDURE, THE 32+9 TRIAL HEAD WAS FOUND TO BE BADLY SCRATCHED AND WORN. THERE WERE NO BROKEN PIECES, NO DELAYS, AND NO COMPLICATIONS FOR THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23708366 ↗
  • MalfunctionMar 3, 2025Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT THE HEAD TRIALS WERE STRIPPED AND WON'T STAY ON NECK. SURGICAL DELAY WAS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21509437 ↗
  • MalfunctionMar 3, 2025Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT THE HEAD TRIALS WERE STRIPPED AND WON'T STAY ON NECK. SURGICAL DELAY WAS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21509401 ↗
  • MalfunctionNov 6, 2024Scratched Material; Scratched Material

    IT WAS REPORTED THAT THE HEADS ARE SCRATCHED AND NEED REPLACED. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20624434 ↗
  • MalfunctionNov 6, 2024Scratched Material; Scratched Material

    IT WAS REPORTED THAT THE HEADS ARE SCRATCHED AND NEED REPLACED. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20623770 ↗
  • MalfunctionOct 31, 2024Break; Material Deformation; Naturally Worn; Scratched Material; Break; Material Deformation; Naturally Worn; Scratched Material

    IT WAS REPORTED THAT THE HEAD TRIALS WERE WORN, SCRATCHED AND CHIPPED. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20578977 ↗
  • MalfunctionOct 28, 2024Material Deformation; Naturally Worn; Scratched Material; Material Deformation; Naturally Worn; Scratched Material

    SURGEON AND OR STAFF NOTED THAT VARIOUS FEMORAL HEAD TRIALS IN HIP SETS ARE SCUFFED, GOUGED AND VISIBLY WORN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20551144 ↗
  • MalfunctionOct 28, 2024Material Deformation; Naturally Worn; Scratched Material; Material Deformation; Naturally Worn; Scratched Material

    SURGEON AND OR STAFF NOTED THAT VARIOUS FEMORAL HEAD TRIALS IN HIP SETS ARE SCUFFED, GOUGED AND VISIBLY WORN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20550603 ↗
  • MalfunctionOct 28, 2024Material Deformation; Naturally Worn; Scratched Material; Material Deformation; Naturally Worn; Scratched Material

    IT WAS REPORTED THAT THE SURGEON AND OPERATING ROOM STAFF NOTED THAT VARIOUS FEMORAL HEAD TRIALS IN THE HIP SETS ARE SCUFFED, GOUGED AND VISIBLY WORN. THE TRIALS WERE USED ANYWAY AS THEY WERE STILL VIABLE AT THAT TIME. THERE WERE NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20546109 ↗
  • MalfunctionJul 26, 2024Naturally Worn; Scratched Material; Naturally Worn; Scratched Material

    STAFF HAVE INDICATED SEVERAL TRIAL FEMORAL HEADS ARE SCRATCHED AND WORN DUE TO WEAR AND TEAR/USAGE OVER TIME. THEY ARE NO LONGER USABLE AND REQUIRE REPLACEMENT. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19843433 ↗
  • MalfunctionJul 2, 2024Break; Break

    IT WAS REPORTED THAT INSTRUMENTS BROKE DURING THE CASE.

    Patient problems: Insufficient Information

    MAUDE report 19656138 ↗
  • MalfunctionJun 28, 2024No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT SURGEONS WANT NEW EMPHASYS HEADS. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19633356 ↗
  • MalfunctionJun 28, 2024No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT SURGEONS WANT NEW EMPHASYS HEADS. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19633330 ↗
  • MalfunctionJun 28, 2024No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT SURGEONS WANT NEW EMPHASYS HEADS. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19633327 ↗
  • MalfunctionJun 28, 2024No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT SURGEONS WANT NEW EMPHASYS HEADS. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19633319 ↗
  • MalfunctionMay 16, 2024Break; Scratched Material; Break; Scratched Material

    IT WAS REPORTED THAT THE SURGEON COMPLAINED THE TRIAL HEADS WERE SCRATCHED AND PLASTIC WAS COMING OFF.

    Patient problems: Insufficient Information

    MAUDE report 19330010 ↗
  • MalfunctionMay 16, 2024Break; Scratched Material; Break; Scratched Material

    IT WAS REPORTED THAT THE SURGEON COMPLAINED THE TRIAL HEADS WERE SCRATCHED AND PLASTIC WAS COMING OFF.

    Patient problems: Insufficient Information

    MAUDE report 19329975 ↗
  • MalfunctionApr 18, 2024Naturally Worn; Scratched Material; Naturally Worn; Scratched Material

    IT WAS REPORTED THAT THE HEADS WERE WORN AND SCRATCHED.

    Patient problems: Insufficient Information

    MAUDE report 19132666 ↗
  • MalfunctionFeb 26, 2024Scratched Material; Scratched Material

    IT WAS REPORTED THAT THE TRIAL IS NICKED/SCRATCHED. PATIENT STATUS/ OUTCOME / CONSEQUENCES: WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED? UNKNOWN. IS THE PATIENT PART OF A CLINICAL STUDY? UNKNOWN. (B)(4). DEVICE PROPERTY OF: NONE. DEVICE IN POSSESSION OF: NONE. (B)(4). DEVICE PROPERTY OF: NONE. DEVICE IN POSSESSION OF: NONE. (B)(4). DEVICE PROPERTY OF: NONE. DEVICE IN POSSESSION OF: NONE. (B)(4). DEVICE PROPERTY OF: NONE. DEVICE IN POSSESSION OF: NONE. (B)(4). DEVICE PROPERTY OF: NONE. DEVICE IN POSSESSION OF: NONE. BY CHECKING THIS BOX I CERTIFY THAT ALL INFORMATION THAT ARE KNOWN/AVAILABLE HAS BEEN DISCLOSED. IF ANY NEW INFORMATION WILL BE MADE AVAILABLE, THE ADDITIONAL INFORMATION WILL BE SUBMITTED THROUGH CST. TRUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18777828 ↗
  • MalfunctionFeb 16, 2024Scratched Material; Scratched Material

    ROUTINE INSPECTION HEAD TRAILS WITH SCRAPES AND GAUGES OUT OF THEM IN NEED OF REPLACEMENT. IT WAS REPORTED THAT THE HEAD TRIALS ARE SCRATCHED/GOUGED. PATIENT STATUS/ OUTCOME / CONSEQUENCES: NO. WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED? UNKNOWN. IS THE PATIENT PART OF A CLINICAL STUDY? UNKNOWN. (B)(4). DEVICE PROPERTY OF: NONE. DEVICE IN POSSESSION OF: NONE. (B)(4). DEVICE PROPERTY OF: NONE. DEVICE IN POSSESSION OF: NONE. (B)(4). DEVICE PROPERTY OF: NONE. DEVICE IN POSSESSION OF: NONE. (B)(4). DEVICE PROPERTY OF: NONE. DEVICE IN POSSESSION OF: NONE. (B)(4). DEVICE PROPERTY OF: NONE. DEVICE IN POSSESSION OF: NONE. (B)(4). DEVICE PROPERTY OF: NONE. DEVICE IN POSSESSION OF: NONE, (B)(4). DEVICE PROPERTY OF: NONE. DEVICE IN POSSESSION OF: NONE. (B)(4). DEVICE PROPERTY OF: NONE. DEVICE IN POSSESSION OF: NONE. BY CHECKING THIS BOX I CERTIFY THAT ALL INFORMATION THAT ARE KNOWN/AVAILABLE HAS BEEN DISCLOSED. IF ANY NEW INFORMATION WILL BE MADE AVAILABLE, THE ADDITIONAL INFORMATION WILL BE SUBMITTED THROUGH CST. TRUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18727186 ↗