ATTUNE 2545-00-524Recalls & reported problems

Depuy (Ireland) · ATTUNE ARTICULATION SURFACE FIXED BEARING CRUCIATE RETAINING SIZE 4 · Class 1

All models of ATTUNE (1475)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This ATTUNE knee articulating surface is a trial instrument used during knee replacement surgery. The reports mostly describe the small spring coil on the instrument's post coming loose or falling off during surgery, along with parts of the instrument breaking, going missing, or bending. There are 85 reports, with no deaths and no injuries. One recall, now closed, started in June 2015 because that spring coil could become damaged and separate during trialing, potentially entering the surgical site.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
85
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 85 adverse event reports that list this exact device (DI 10603295132646), received between Jul 24, 2017 and Aug 26, 2026 — classified as 85 malfunctions.

The most frequently reported specific device problem is “Break” (45 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 4 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 2 reportsJun 2024: 1 reportsJul 2024: 2 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 2 reportsOct 2025: 1 reportsNov 2025: 4 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 12, 2015Class IITerminated

    Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal is separated from the post, it has the potential to enter the surgical site and be left in the patient if the surgeon is unaware of the disassociation.

    Z-2138-2015 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break45
  • Detachment of Device or Device Component28
  • Component Missing9
  • Material Deformation8
  • Use of Device Problem4
  • Material Twisted/Bent3
  • Loose or Intermittent Connection2
  • Crack1

Patient problems

  • No Clinical Signs, Symptoms or Conditions56
  • No Consequences Or Impact To Patient22
  • Insufficient Information5
  • No Information3
  • Not Applicable2
  • No Code Available1

Recent adverse event reports

The 25 most recent of 85 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 26, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50453832 ↗
  • MalfunctionJun 5, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT THERE WERE A COUPLE OF LOST SPRINGS FROM ARTICULATING SURFACES IN AN ATTUNE KIT DURING SURGERIES. IT WAS ALSO NOTED THAT THE IM REAMER HAD A NOT SHARP TIP. NO IMPACT ON PATIENT HAS BEEN REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25414332 ↗
  • MalfunctionMay 18, 2026Break

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE IT WAS OBSERVED THAT THE ATTUNE FB CR ARTIC SURF SZ4 TRIAL DEVICE BROKE. IT WAS REPORTED THAT THERE WAS NO DELAY IN THE PROCEDURE DUE TO THE EVENT. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25209254 ↗
  • MalfunctionDec 18, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE BITS WERE BROKEN. IT WAS FURTHER REPORTED THAT THE SPRING WENT OF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23846239 ↗
  • MalfunctionNov 26, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE TIBIAL DEVICE BROKE DURING KNEE SURGERY. THE SPRING CAME OFF THE ARTICULATING SURFACE AND NEEDS REPLACING. NOTHING WAS LEFT INSIDE THE PATIENT, THE INSTRUMENT DID NOT BREAK IN TWO (THE SPRING CAME OFF), IT DID NOT DELAY THE SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23658731 ↗
  • MalfunctionNov 7, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THE INSTRUMENTS WERE MISSING SPRINGS, AND EASILY DISPLACED FROM THE ARCTIC SURFS. BROKEN INSTRUMENTS, HAPPENED DURING SURGERY BUT NO PATIENT AFFECTED, OR SURGERY DELAYED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23496035 ↗
  • MalfunctionNov 6, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THE SPRINGS SPANNED OFF OF THE ATTUNE ARTICULATION SURFACE FIXED BEARING CRUCIATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23489871 ↗
  • MalfunctionNov 6, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THE SPRINGS FELL OFF IN COUPLE OF PROCEDURES. (B)(4), IS RELATED TO (B)(4) AND (B)(4). (B)(4), IS THE MOTHER COMPLAINT. (B)(4), AND (B)(4), WERE CREATED TO CAPTURE THE ADDITIONAL IPS OF THE MOTHER COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23485750 ↗
  • MalfunctionOct 27, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT SSD NOTED THAT THE THE SPRING WAS MISSING ON THE DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23392393 ↗
  • MalfunctionSep 11, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE ITEM WAS FOUND TO HAVE THE SPRINGS MISSING BEFORE THE CASE. THERE WAS NO RISK OF THEM BEING NEAR THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23030535 ↗
  • MalfunctionSep 2, 2025Break; Break

    IT WAS REPORTED THAT THE ITEMS WERE BROKEN ON AN UNKNOWN DATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22939819 ↗
  • MalfunctionJul 30, 2025Detachment of Device or Device Component; Material Deformation; Detachment of Device or Device Component; Material Deformation

    IT WAS REPORTED DURING THE PROCEDURE, THE METAL SPRINGS FROM THE UNDERSIDE OF THE ARTICULATING SURFACE HAVE BECOME DETACHED/FALLEN OFF. THE SPRINGS ARE MISSHAPEN AND STRETCHED AND SO CANNOT BE REATTACHED SECURELY. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22664179 ↗
  • MalfunctionJul 9, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THERE WERE BROKEN INSTRUMENTS. THE SPRINGS WERE NOT ATTACHED ANYMORE. ISSUE OCCURRED DURING SURGERY, BUT THERE WERE NO ISSUES AS THE SURGICAL TEAM MANAGED TO TAKE THE SPRING OUT. THE PATIENT WAS NOT AFFECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22460425 ↗
  • MalfunctionJun 6, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE INSTRUMENT BROKE DURING KNEE SURGERY. THE SPRING CAME OFF THE BACK OF THE ARTICULATING SURFACE, IT DID NOT BREAK IN HALF, NOTHING WAS LEFT INSIDE THE PATIENT AND IT DID NOT DELAY SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22161931 ↗
  • MalfunctionMay 6, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THE BELOW INSTRUMENT BROKE DURING KNEE SURGERY THIS MORNING AT HOSPITAL. IT IS MISSING ONE OF ITS SPRINGS, THERE WERE NO DELAYS TO SURGERY, NOTHING WAS LEFT INSIDE THE PATIENT. THE MISSING SPRING WAS NOTED ON OPENING THE INSTRUMENT TRAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21969549 ↗
  • MalfunctionApr 22, 2025Break; Break

    IT WAS REPORTED THAT PATIENT UNDERWENT TOTAL KNEE REPLACEMENT ON (B)(6) 2025. THE ATTUNE ARTICULAR SURFACE TRIAL HAD A CHIPPED TAB. NO AFFECT ON PATIENT. NO DELAY TO SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21881539 ↗
  • MalfunctionFeb 20, 2025Break; Break

    IT WAS REPORTED THAT BAL SPRINGS WERE BROKEN.

    Patient problems: Insufficient Information

    MAUDE report 21426919 ↗
  • MalfunctionDec 30, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT ON (B)(6) 2024, A PINNACLE INSTRUMENT WAS BROKEN. ALSO, AN ATTUNE ARTICULATE SURFACE WAS MISSING A SPRING. THERE WAS NO REPORTED PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21037120 ↗
  • MalfunctionDec 12, 2024Break; Break

    IT WAS REPORTED THAT THE INSTRUMENTS ARE BROKEN. THE BREAKAGE DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20918693 ↗
  • MalfunctionOct 2, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ADDITIONAL INFORMATION WAS RECEIVED: THERE WAS NO PATIENT HARM AND NO DELAYS TO SURGERY.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20353086 ↗
  • MalfunctionAug 22, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT THERE IS A SPRING MISSING FROM THE ATTUNE ARTICULATING SURFACE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20030090 ↗
  • MalfunctionJul 17, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE SPRING WAS MISSING FROM THE ARTICULATING SURFACE SO THE SHIMS ARE NOT SNAPPING TOGETHER PROPERLY. THERE WAS NO DELAY IN SURGERY AND NO NEGATIVE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19764853 ↗
  • MalfunctionJul 11, 2024Detachment of Device or Device Component; Material Deformation; Detachment of Device or Device Component; Material Deformation

    IT WAS REPORTED THAT THE METAL SPRING AROUND THE PEG GOT SNAPPED AND BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19720670 ↗
  • MalfunctionJun 4, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THE INSTRUMENT BROKE DURING KNEE SURGERY AT HOSPITAL. NO TIME WAS ADDED TO THE CASE, NO ADVERSE EVENTS WERE CAUSED TO THE PATIENT. THE SPRING CAME OFF THE ARTICULATING SURFACE, NO PARTS WERE LEFT IN THE PATIENT. IT DID NOT DELAY THE CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19458740 ↗
  • MalfunctionMay 1, 2024Break; Detachment of Device or Device Component; Material Deformation; Break; Detachment of Device or Device Component; Material Deformation

    THE ARTICULATING SURFACE TRIAL PEG AND SPRING WERE DAMAGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19226478 ↗

Other TEMPLATE devices

All TEMPLATE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page