Skip to content

ATTUNE 2545-00-691Recalls & reported problems

Depuy (Ireland) · ATTUNE SHIM SIZE 9 5mm · Class 1

All models of ATTUNE (1475)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Attune shim is a surgical sizing tool used during knee replacement. Of its 90 reports, 1 involves an injury and no deaths are recorded. Reports mostly describe the instrument breaking, cracking, or parts coming loose, and most describe no clinical signs or impact on the patient. There were 2 reports in the last 12 months, down from 8 before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
90
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 90 adverse event reports that list this exact device (DI 10603295133636), received between Jun 30, 2017 and Apr 3, 2026 — classified as 1 injury, 89 malfunctions.

The most frequently reported specific device problem is “Break” (53 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 4 reportsApr 2024: 1 reportsMay 2024: 4 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 4 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break53
  • Crack22
  • Detachment of Device or Device Component18
  • Device-Device Incompatibility6
  • Component Missing3
  • Adverse Event Without Identified Device or Use Problem1
  • Appropriate Device Problem Term/Code Not Available1
  • Device Dislodged or Dislocated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions43
  • No Consequences Or Impact To Patient36
  • Insufficient Information7
  • No Information4
  • No Code Available2

Recent adverse event reports

The 25 most recent of 90 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 3, 2026Break

    IT WAS REPORTED THAT THE INSTRUMENTS HAVE BEEN BROKEN OVER TIME. THEY WERE NOT USED, THERE WAS NO DELAY TO SURGERY, AND THE PATIENT WAS NOT AFFECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24780623 ↗
  • MalfunctionMar 5, 2026Break

    IT WAS REPORTED THAT THE BITS FROM ATTUNE KIT ARE BROKEN. THERE WAS NO SURGERY DELAY OR PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24521833 ↗
  • MalfunctionSep 4, 2025Break; Break

    IT WAS REPORTED THAT THE SHIMM HAD BROKEN WHEN PROCESSING TRAY. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22964694 ↗
  • MalfunctionJun 27, 2025Break; Break

    IT WAS REPORTED THAT THE ATTUNE SHIM SZ9 5MM WAS BROKEN/ CHIPPED. THE BASEPLATE INSERTER ROD WAS BROKEN. THE EXCEL T-HANDLE WAS BROKEN. THE ACTIS RETAINING STEM INSERTER WAS BROKEN AT BOTTOM. THE PINNACLE STRAIGHT GRATER HNDL CAME APART. THIS EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22346916 ↗
  • MalfunctionJun 19, 2025Break; Break

    IT WAS REPORTED THE INSTRUMENT BROKE INTO TWO (2) PIECES PRIOR TO SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22292781 ↗
  • MalfunctionJun 6, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE ATTUNE SHIM SZ9 5MM WAS MISSING THE METAL RING. SHIM CANNOT CONNECT TO INSERT TOP WHEN SURGEON IS TRIALLING IMPLANTS FOR KNEE REPLACEMENT. NOTICED DURING SURGERY, HOWEVER INSTRUMENT WAS NOT NEEDED FOR THE PROCEDURE SO THERE WAS NO IMPACT TO THE PATIENT. THERE WAS NO DELAY TO SURGERY, NO IMPACT ON PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22161857 ↗
  • MalfunctionJun 2, 2025Break; Break

    IT WAS REPORTED THAT A SHIM FROM A CONSIGNMENT TRAY WAS BROKEN; AT THE RAISED SILVER DISC FOR ATTACHMENT. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22126366 ↗
  • MalfunctionMay 6, 2025Break; Break

    IT WAS REPORTED THAT THE 9/10 5MM SHIM BROKE DURING TRIALING. THERE WAS NO PATIENT HARM AND NO SURGICAL DELAY.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21971426 ↗
  • MalfunctionApr 3, 2025Break; Break

    IT WAS REPORTED THAT THE INSTRUMENTS BROKE. NO FURTHER DETAILS WERE PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 21761155 ↗
  • MalfunctionDec 2, 2024Crack; Detachment of Device or Device Component; Crack; Detachment of Device or Device Component

    ADDITIONAL INFORMATION WAS RECEIVED AND STATED THAT THE INSTRUMENT WAS BROKEN INTO 2 PIECES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20818323 ↗
  • MalfunctionMay 16, 2024Break; Break

    IT WAS REPORTED THAT INSTRUMENT BROKE ON BACK TABLE WHILE SNAPPING TOGETHER. THIS WAS NOT DONE NEAR PATIENT. ALL 3 BROKEN PIECES REMOVED FROM BACK TABLE. THERE WAS NO DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19331806 ↗
  • MalfunctionMay 13, 2024Break; Break

    ADDITIONAL INFORMATION WAS RECEIVED: A. WAS THERE ANY SURGICAL DELAY RELATED TO THE EVENT? IF YES, WHAT IS THE DURATION OF THE DELAY? NO DELAY. B. WAS THERE ANY PATIENT HARM/CONSEQUENCE RELATED TO THE EVENT? NO HARM. C. DID THE INSTRUMENT BROKE INTO 2 OR MORE PIECES? TIBIAL TRIAL WAS IN 2 PIECES. BOTH RETRIEVED EASILY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19298325 ↗
  • MalfunctionMay 8, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT BOTH SIZE 9/10 BY 5MM SHIMS HAD THE METAL RING THAT CLIPS INTO THE POLY TRIAL FALL OUT RENDERING THEM UNUSABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19274120 ↗
  • MalfunctionMay 8, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT BOTH SIZE 9/10 BY 5MM SHIMS HAD THE METAL RING THAT CLIPS INTO THE POLY TRIAL FALL OUT RENDERING THEM UNUSABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19274115 ↗
  • MalfunctionApr 11, 2024Break; Break

    IT WAS REPORTED THAT THE METAL RING FELL OUT OF SHIM WHEN TAKING IT OFF OF THE SPACER BLOCK. SHIM WAS DISCARDED BY FACILITY. ALL PIECES WERE RETRIEVED. NOTHING FURTHER TO REPORT.

    Patient problems: Insufficient Information

    MAUDE report 19087683 ↗
  • MalfunctionMar 27, 2024Crack; Crack

    IT WAS REPORTED THAT THERE WAS A CRACK ON THE ATTUNE SHIM AND THE KNOB OF THE INSTRUMENT HAS SNAPPED OFF PATIENT STATUS/ OUTCOME / CONSEQUENCES , WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED: UNKNOWN, IS THE PATIENT PART OF A CLINICAL STUDY UNKNOWN, (B)(4) DEVICE PROPERTY OF NONE, DEVICE IN POSSESSION OF NONE, (B)(4) DEVICE PROPERTY OF NONE, DEVICE IN POSSESSION OF NONE BY CHECKING THIS BOX I CERTIFY THAT ALL INFORMATION THAT ARE KNOWN/AVAILABLE HAS BEEN DISCLOSED. IF ANY NEW INFORMATION WILL BE MADE AVAILABLE, THE ADDITIONAL INFORMATION WILL BE SUBMITTED THROUGH CST. TRUE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18985655 ↗
  • MalfunctionMar 25, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE METAL RING ON SHIM FELL OUT UPON INSERTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18970237 ↗
  • MalfunctionMar 11, 2024Crack; Detachment of Device or Device Component; Crack; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE SHIMS WERE CRACKED/MISSING METAL CONNECTING PIECES.

    Patient problems: Insufficient Information

    MAUDE report 18874766 ↗
  • MalfunctionMar 11, 2024Crack; Detachment of Device or Device Component; Crack; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE SHIMS WERE CRACKED/MISSING METAL CONNECTING PIECES.

    Patient problems: Insufficient Information

    MAUDE report 18874591 ↗
  • MalfunctionFeb 9, 2024Break; Break

    IT WAS REPORTED THAT THERE IS A BROKEN/CHIPPED PEG ON THE 41 MM OVAL DOME PATELLA TRIAL AND CHIPPED SHIM SIZE 9/10 - 5MM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18675685 ↗
  • MalfunctionSep 29, 2023Break; Crack; Break; Crack

    IT WAS REPORTED THAT SHIM BROKE INTO TWO DURING EXTRACTIONS. NOTHING WAS LEFT IN PATIENT. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17844977 ↗
  • MalfunctionJul 28, 2023Break; Break

    IT WAS REPORTED THAT THE SHIM WAS FOUND BROKEN BEFORE CASE STARTED. IT DID NOT AFFECT THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17413641 ↗
  • MalfunctionJul 11, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE METAL SLEEVE INTERFACE IN THE SHIM WAS MISSING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17300334 ↗
  • MalfunctionApr 11, 2023Break; Crack; Break; Crack

    IT WAS REPORTED THAT DEVICE WAS DAMAGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16716614 ↗
  • MalfunctionApr 6, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE WHITE TRIAL POSTS WERE BOTH DAMAGED AND THE SHIMS DON'T HOLD TO INSERT TRIAL WELL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16690518 ↗

Other TEMPLATE devices

All TEMPLATE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page