Skip to content

ATTUNE 2545-00-724Recalls & reported problems

Depuy (Ireland) · ATTUNE FEMORAL TRIAL POSTERIOR STABLIZED SIZE 4 RIGHT · Class 1

All models of ATTUNE (1475)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy ATTUNE femoral trial post is a sizing template used during hip replacement surgery to measure the femoral canal before implanting the real parts. The reports mostly describe the plastic template breaking, cracking or chipping during impaction, plus scratches, discoloration and worn surfaces; in most cases surgery continued with another tray and no lasting harm. There are 49 reports, 1 injury and no deaths, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
49
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 49 adverse event reports that list this exact device (DI 10603295133964), received between Oct 10, 2017 and Jul 1, 2026 — classified as 1 injury, 48 malfunctions.

The most frequently reported specific device problem is “Break” (18 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 3 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break18
  • Scratched Material13
  • Crack11
  • Naturally Worn4
  • Material Discolored3
  • Appropriate Device Problem Term/Code Not Available2
  • Device-Device Incompatibility2
  • Illegible Information2

Patient problems

  • No Consequences Or Impact To Patient25
  • No Clinical Signs, Symptoms or Conditions18
  • Insufficient Information3
  • No Information2
  • Bone Fracture(s)1
  • Not Applicable1

Recent adverse event reports

The 25 most recent of 49 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 1, 2026Break

    IT WAS REPORTED THAT DURING IMPACTION OF THE FEMORAL TRIAL, THE FEMORAL TRIAL BROKE. THIS DID NOT IMPACT THE LENGTH OF THE SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25680550 ↗
  • MalfunctionDec 21, 2025Break; Break

    WAS THE PRODUCT BEING USED IN A CLINICAL TRIAL? NO. DID THE EVENT HAPPEN DURING A PROCEDURE? YES. WERE YOU IN THE PROCEDURE AT THE TIME OF THE EVENT? NO. EVENT OUTCOME/HOW WAS IT MANAGED? ANOTHER TRAY NEEDED TO BE OPENED WAS THERE ANY CONSEQUENCE TO THE PATIENT DUE TO THE EVENT? NO. WAS THE SURGERY PROLONGED DUE TO THE EVENT? BY 3 MINUTES HAS THE REPORTER FACILITY INDICATED THERE MAY BE LEGAL ACTION? NO. PLEASE GIVE A DETAILED EXPLANATION OF EVENT: THE POSTERIOR CONDYLE OF THE FEMUR TRIAL BROKE OF DURING IMPACTION ONTO THE FEMUR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23861277 ↗
  • MalfunctionOct 31, 2025Break; Break

    IT WAS REPORTED THAT THE INSTRUMENTS WERE BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23435850 ↗
  • MalfunctionSep 27, 2024No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE ALLEGATION IS UNKNOWN. THE EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20321628 ↗
  • MalfunctionJul 23, 2024Break; Naturally Worn; Break; Naturally Worn

    IT WAS REPORTED THAT ALL THE FEMORAL TRIALS ARE BECOMING CHIPPED AND WORN FROM TOO MUCH USE. UNKNOWN SURGICAL DELAY.

    Patient problems: Insufficient Information

    MAUDE report 19806954 ↗
  • MalfunctionJul 23, 2024Break; Naturally Worn; Break; Naturally Worn

    IT WAS REPORTED THAT ALL THE FEMORAL TRIALS ARE BECOMING CHIPPED AND WORN FROM TOO MUCH USE. UNKNOWN SURGICAL DELAY.

    Patient problems: Insufficient Information

    MAUDE report 19806931 ↗
  • MalfunctionJul 23, 2024Break; Naturally Worn; Break; Naturally Worn

    IT WAS REPORTED THAT ALL THE FEMORAL TRIALS ARE BECOMING CHIPPED AND WORN FROM TOO MUCH USE. UNKNOWN SURGICAL DELAY.

    Patient problems: Insufficient Information

    MAUDE report 19806819 ↗
  • MalfunctionJul 27, 2023Corroded; Material Discolored; Contamination /Decontamination Problem; Corroded; Material Discolored; Contamination /Decontamination Problem

    DEVICE REPORT FROM DEPUY SYNTHES AUSTRALIA REPORTS AN EVENT AS FOLLOWS:DURING A LOAN KIT INSPECTION, THE INSTRUMENTS WERE FOUND TO HAVE BLACK SPOTS ALL OVER THE THEM. NO OTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17409924 ↗
  • MalfunctionJul 21, 2022Break; Crack; Break; Crack

    IT WAS REPORTED THAT THE FEMORAL TRIAL HAS BROKEN IN HALF. THERE WAS A CRACK AND BROKEN THROUGH UPON IMPACTION. ALL PIECES WERE RECOVERED. NO ADVERSE AFFECTS ON THE PATIENT. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15073099 ↗
  • MalfunctionOct 5, 2021Break; Break

    IT WAS REPORTED THAT ALL ATTUNE FEMORAL TRIALS HAVE FRACTURED FROM REPEATED IMPACTION. 4 OF EACH SIDE, 24 TOTAL REPLACEMENT TRIAL FEMURS. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12578104 ↗
  • MalfunctionOct 5, 2021Break; Break

    IT WAS REPORTED THAT ALL ATTUNE FEMORAL TRIALS HAVE FRACTURED FROM REPEATED IMPACTION. 4 OF EACH SIDE, 24 TOTAL REPLACEMENT TRIAL FEMURS. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12577763 ↗
  • MalfunctionOct 5, 2021Break; Break

    IT WAS REPORTED THAT ALL ATTUNE FEMORAL TRIALS HAVE FRACTURED FROM REPEATED IMPACTION. 4 OF EACH SIDE, 24 TOTAL REPLACEMENT TRIAL FEMURS. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12577748 ↗
  • MalfunctionOct 5, 2021Break; Break

    IT WAS REPORTED THAT ALL ATTUNE FEMORAL TRIALS HAVE FRACTURED FROM REPEATED IMPACTION. 4 OF EACH SIDE, 24 TOTAL REPLACEMENT TRIAL FEMURS. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12577489 ↗
  • MalfunctionSep 7, 2021Crack; Crack

    HOSPITAL STAFF NOTICED CROUTINE INSPECTION. NO SURGICAL DELAYLS DURING ROUTINE INSPECTION. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12431997 ↗
  • MalfunctionSep 7, 2021Crack; Crack

    HOSPITAL STAFF NOTICED CROUTINE INSPECTION. NO SURGICAL DELAYLS DURING ROUTINE INSPECTION. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12431988 ↗
  • MalfunctionSep 7, 2021Crack; Crack

    HOSPITAL STAFF N DURING ROUTINE INSPECTION. NO SURGICAL DELAYLS DURING ROUTINE INSPECTION. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12431987 ↗
  • MalfunctionSep 7, 2021Crack; Crack

    HOSPITAL STAFF N DUCED CROUTINE INSPECTION. NO SURGICAL DELAYLS DURING ROUTINE INSPECTION. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12431971 ↗
  • MalfunctionJul 8, 2021Break; Break

    IT WAS REPORTED THE INSTRUMENT WAS BROKEN WHILE IMPACTING. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12135807 ↗
  • MalfunctionMay 14, 2021Scratched Material; Illegible Information; Scratched Material; Illegible Information

    IT WAS REPORTED THAT THE NOTCH GUIDES HAVE ALL LOST THE WHITE ON THE NUMBERS AND ALSO HAVE WEAR/GROOVES ON THE METAL SURFACES. THE FEMORAL TRIALS ARE DISCOLORED AND HAVE GOUGES ON THE SURFACE. NO REPORTED SURGICAL DELAYS OR PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11825995 ↗
  • MalfunctionApr 7, 2021Scratched Material; Scratched Material

    IT WAS REPORTED THAT THE FOLLOWING ATTUNE FEMORAL TRIALS WERE REQUESTED TO BE REPLACED BY SPD AS THEY WERE BEAT UP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11629427 ↗
  • MalfunctionJan 4, 2021Material Integrity Problem; Material Integrity Problem

    IT WAS STATED THAT FEMORAL TRIALS ARE DAMAGED. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11113630 ↗
  • MalfunctionOct 8, 2020Break; Break

    ADDITIONAL INFORMATION RECEIVED INDICATED THAT THE ORANGE RING WAS CHIPPED AND CRACKED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10650207 ↗
  • MalfunctionSep 24, 2020Crack; Crack

    IT WAS REPORTED THAT THERE WERE CRACKED ON FEMUR TRIALS. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10583048 ↗
  • MalfunctionAug 21, 2020Break; Break

    IT WAS REPORTED THAT THE SIZE 4 RIGHT FEMORAL TRIAL IS BROKEN. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10436594 ↗
  • InjuryJul 22, 2020Break; Break

    DURING IMPACTION THE LATERAL CONDYLE OF THE ATTUNE FEMORAL TRIAL BROKE OFF (254500724) . AS THE PIECES OF THE BROKEN TRIAL WERE REMOVED IT WAS FOUND THAT THE PATIENT HAD DEVELOPED A NON DISPLACED LATERAL FEMORAL CONDYLAR FRACTURE STARTING AT THE ANTERIOR PORTION OF THE BOX RESECTION AND EXTENDING TO THE POSTERIOR PORTION OF THE FEMUR. THIS WAS REPAIRED USING TWO 4.0MM CANNULATED SCREWS AND THE SURGERY WAS FINISHED WITH NO FURTHER INCIDENT.

    Patient problems: Bone Fracture(s); Not Applicable

    MAUDE report 10309167 ↗