Skip to content

ATTUNE 2545-00-737Recalls & reported problems

Depuy (Ireland) · ATTUNE FEMORAL TRIAL POSTERIOR STABLIZED SIZE 7 LEFT · Class 1

All models of ATTUNE (1475)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy ATTUNE femoral trial is a sizing tool used in knee replacement surgery. The FDA records hold 96 reports, with no deaths and no injuries; many describe the trial breaking or cracking during use or inspection, and some describe scratched surfaces. Many reports say there was no patient harm. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
95
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 95 adverse event reports that list this exact device (DI 10603295134091), received between Jan 5, 2018 and May 19, 2026 — classified as 95 malfunctions.

The most frequently reported specific device problem is “Break” (48 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 2 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 3 reportsAug 2024: 1 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 4 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break48
  • Scratched Material25
  • Crack17
  • Naturally Worn9
  • Appropriate Device Problem Term/Code Not Available5
  • Material Twisted/Bent5
  • Material Integrity Problem4
  • Device-Device Incompatibility3

Patient problems

  • No Consequences Or Impact To Patient46
  • No Clinical Signs, Symptoms or Conditions43
  • Insufficient Information6
  • No Code Available1

Recent adverse event reports

The 25 most recent of 95 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 19, 2026Break

    IT WAS REPORTED THAT DURING OFFICE INSPECTION A FRACTURE WAS FOUND ON THE BACK PORTION OF FEMUR TRIALS. IT WAS REPORTED THAT THERE WAS NO SURGICAL DELAY. IT WAS REPORTED THAT THERE WAS NO PATIENT HARM. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25222619 ↗
  • MalfunctionMay 6, 2026Break

    IT WAS REPORTED THAT THE TRIAL BROKE AS THE DOCTOR WAS IMPACTING. THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25096841 ↗
  • MalfunctionMar 30, 2026Break; Crack

    IT WAS REPORTED THAT TRIALS WERE CRACKED OR BROKEN AND THE HOSPITAL WILL NOT ACCEPT THEM. THE ANGEL WING WAS BENT, AND THE TIBIA IMPACTOR HAS POPPED ITS SPRING. THE POLY LOST A SPRING ON ONE OF THE POSTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24729980 ↗
  • MalfunctionFeb 5, 2026Device-Device Incompatibility; Scratched Material

    IT WAS REPORTED THAT ATTUNE PS FEM TRIAL SZ 7 LT WAS SCRATCHED AND UNABLE TO ARTICULATE PROPERLY DURING A PROCEDURE. THE EVENT OCCURRED DURING A PROCEDURE, BUT THERE WAS NO ADVERSE OUTCOME OR CONSEQUENCE TO THE PATIENT, AND THE SURGERY WAS NOT PROLONGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24268525 ↗
  • MalfunctionNov 3, 2025Break; Break

    IT WAS REPORTED THAT INSTRUMENTS WERE BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23450922 ↗
  • MalfunctionOct 31, 2025Break; Break

    IT WAS REPORTED THAT THE INSTRUMENTS WERE BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23438869 ↗
  • MalfunctionOct 31, 2025Break; Break

    IT WAS REPORTED THE FEMORAL TRIAL WAS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23435720 ↗
  • MalfunctionOct 31, 2025Break; Break

    IT WAS REPORTED THE FEMORAL TRIAL WAS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23435719 ↗
  • MalfunctionOct 15, 2025Break; Break

    IT WAS REPORTED THAT DEVICE WAS BROKEN DURING TRIALING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23290485 ↗
  • MalfunctionJul 24, 2025Break; Break

    IT WAS REPORTED THAT THE INSTRUMENTS WERE FOUND WITHIN A BROKEN INSTRUMENTS TUB AT HOSPITAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22588519 ↗
  • MalfunctionJun 12, 2025Break; Break

    IT WAS REPORTED THERE WERE MULTIPLE BROKEN DEVICES.

    Patient problems: Insufficient Information

    MAUDE report 22197453 ↗
  • MalfunctionJan 28, 2025Break; Break

    IT WAS REPORTED THAT SURGEON BROKE THE INSTRUMENT WHILE TRIALING.

    Patient problems: Insufficient Information

    MAUDE report 21255405 ↗
  • MalfunctionJan 10, 2025Material Discolored; Illegible Information; Material Discolored; Illegible Information

    ADDITIONAL INFORMATION RECEIVED STATES THAT THE SALES REP WAS NOT SURE WHEN THE ISSUE OCCURRED BUT HE MENTIONED THAT IT WAS NOTICED DURING A CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21124855 ↗
  • MalfunctionSep 4, 2024No Apparent Adverse Event; No Apparent Adverse Event

    THE INSTRUMENTS WERE REPORTED FOR UNKNOWN REASON. DID NOT HAPPEN DURING SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20137714 ↗
  • MalfunctionSep 4, 2024No Apparent Adverse Event; No Apparent Adverse Event

    THE INSTRUMENTS WERE REPORTED FOR UNKNOWN REASON. DID NOT HAPPEN DURING SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20137711 ↗
  • MalfunctionAug 28, 2024Break; Break

    THE ATTUNE ARTICULAR SURFACE PEG ON THE UNDERSIDE OF THE INSTRUMENT BROKE OFF CLEANLY WHILE THE SURGEON WAS TRYING TO INSERT IT. ALL FRAGMENTS WERE RETRIEVED AND A DIFFERENT ARTICULAR SURFACE WAS QUICKLY USED WITHOUT ANY FURTHER ISSUES. THIS ONLY DELAYED THE SURGERY FOR A FEW MINUTES. SURGERY DELAYED 3 MINUTES DUE TO THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20084837 ↗
  • MalfunctionJul 23, 2024Break; Naturally Worn; Break; Naturally Worn

    IT WAS REPORTED THAT ALL THE FEMORAL TRIALS ARE BECOMING CHIPPED AND WORN FROM TOO MUCH USE. UNKNOWN SURGICAL DELAY.

    Patient problems: Insufficient Information

    MAUDE report 19807355 ↗
  • MalfunctionJul 23, 2024Break; Naturally Worn; Break; Naturally Worn

    IT WAS REPORTED THAT ALL THE FEMORAL TRIALS ARE BECOMING CHIPPED AND WORN FROM TOO MUCH USE. UNKNOWN SURGICAL DELAY.

    Patient problems: Insufficient Information

    MAUDE report 19807352 ↗
  • MalfunctionJul 23, 2024Break; Naturally Worn; Break; Naturally Worn

    IT WAS REPORTED THAT ALL THE FEMORAL TRIALS ARE BECOMING CHIPPED AND WORN FROM TOO MUCH USE. UNKNOWN SURGICAL DELAY.

    Patient problems: Insufficient Information

    MAUDE report 19807351 ↗
  • MalfunctionApr 9, 2024Break; Naturally Worn; Break; Naturally Worn

    IT WAS REPORTED THAT THE ATTUNE INSTRUMENTS ARE BROKEN AND WORN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19070026 ↗
  • MalfunctionJan 23, 2024Break; Scratched Material; Break; Scratched Material

    IT WAS REPORTED THAT THE FEMORAL TRIAL HAS A NICK IN THE MATERIAL. THE PRODUCT DAMAGE DOCUMENTED IN THIS PC HAS BEEN IDENTIFIED DURING LOAN KIT INSPECTION, BY THE LOAN KIT TECHNICIAN. THE EVENT DATE AND ALERT DATE ARE THE DATE THAT THE INSPECTION TOOK PLACE. THERE IS NO SURGEON, PROCEDURE, OR PATIENT DETAILS AVAILABLE. NO FURTHER INFORMATION CAN BE OBTAINED AS THE CASE WAS NOT REPORTED BY THE CUSTOMER. PATIENT STATUS/ OUTCOME / CONSEQUENCES , WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED: UNKNOWN, IS THE PATIENT PART OF A CLINICAL STUDY UNKNOWN, (B)(4) DEVICE PROPERTY OF NONE, DEVICE IN POSSESSION OF NONE BY CHECKING THIS BOX I CERTIFY THAT ALL INFORMATION THAT ARE KNOWN/AVAILABLE HAS BEEN DISCLOSED. IF ANY NEW INFORMATION WILL BE MADE AVAILABLE, THE ADDITIONAL INFORMATION WILL BE SUBMITTED THROUGH CST.TRUE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18563186 ↗
  • MalfunctionDec 7, 2023Break; Break

    IT WAS REPORTED THAT THE INSTRUMENTS BROKE.

    Patient problems: Insufficient Information

    MAUDE report 18282802 ↗
  • MalfunctionDec 1, 2023Break; Break

    ADDITIONAL INFORMATION RECEIVED FROM SALES REP STATES THAT THERE WAS NO DELAY, PATIENT WAS NOT AFFECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18252104 ↗
  • MalfunctionNov 15, 2023Break; Crack; Break; Crack

    IT WAS REPORTED THAT BOTH FEMORAL TRIALS HAVE GIANT CRACKS IN THEM. ONE IS ON THE BOTTOM PART OF THE BOX AND THE OTHER IS AT THE TOP PORTION OF THE BOX. NEITHER INSTRUMENT BROKE INTO TWO PIECES. NOTHING WAS LEFT IN THE PATIENT. NO ADVERSE EFFECTS. NO TIME ADDED TO CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18145944 ↗
  • MalfunctionNov 15, 2023Break; Crack; Break; Crack

    IT WAS REPORTED THAT BOTH FEMORAL TRIALS HAVE GIANT CRACKS IN THEM. ONE IS ON THE BOTTOM PART OF THE BOX AND THE OTHER IS AT THE TOP PORTION OF THE BOX. NEITHER INSTRUMENT BROKE INTO TWO PIECES. NOTHING WAS LEFT IN THE PATIENT. NO ADVERSE EFFECTS. NO TIME ADDED TO CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18145943 ↗

Other TEMPLATE devices