ATTUNE 2545-00-745Recalls & reported problems
Depuy (Ireland) · ATTUNE SHIM SIZE 9 7mm · Class 1
All models of ATTUNE (1475)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 10603295134176), received between Jan 19, 2018 and May 11, 2026 — classified as 15 malfunctions.
The most frequently reported specific device problem is “Break” (11 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break11
- Crack7
- Detachment of Device or Device Component4
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Consequences Or Impact To Patient7
- Insufficient Information1
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 11, 2026Break
IT WAS REPORTED THAT THE PATIENT HAD A RIGHT TOTAL KNEE. DURING TRIALING, THE TIBIAL INSERT SPACER SHIM BROKE INTO TWO PIECES. ALL PIECES WERE FOUND.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25136861 ↗MalfunctionMar 11, 2024Crack; Detachment of Device or Device Component; Crack; Detachment of Device or Device Component
IT WAS REPORTED THAT THE SHIMS WERE CRACKED/MISSING METAL CONNECTING PIECES.
Patient problems: Insufficient Information
MAUDE report 18875050 ↗MalfunctionAug 14, 2023Break; Crack; Break; Crack
IT WAS REPORTED THAT DEVICE WAS CRACKED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17541748 ↗MalfunctionMar 2, 2023Break; Break
IT WAS REPORTED THAT THE PLASTIC OF SHIM WAS BROKEN IN TWO PIECES. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16469816 ↗MalfunctionNov 21, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED DURING INSTRUMENT INSPECTION IT WAS NOTICED THAT THE METAL INSERTS IN THE SHIM HOLES ARE MISSING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15839144 ↗MalfunctionMay 17, 2021Crack; Detachment of Device or Device Component; Crack; Detachment of Device or Device Component
ATTUNE SHIM HAS CRACKED AROUND THE METAL INSERT AREA, ALSO ONE OF THE METAL LOCATION RINGS HAS COME OUT OF THE SHIM BODY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11832553 ↗MalfunctionDec 17, 2020Break; Break
EMAIL RECEIVED FROM CSSD DEPARTMENT AT SJOG MURDOCH WITH A LIST OF BROKEN INSTRUMENTS. THE INSTRUMENTS THAT ARE BROKEN WERE ALL FOUND TO BE BROKEN AFTER THE CLEANING PHASE OF USE. THE SURGERY THAT THEY WERE USED IN AND THE SURGEON USING THESE IS ALL UNKNOWN. THE DATE OF SURGERY AND THUS BREAKAGE IS ALSO UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11028971 ↗MalfunctionDec 15, 2020Break; Break
IT WAS REPORTED THAT SHIMS FOUND BROKEN IN STERILE PROCESSING. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11008981 ↗MalfunctionOct 22, 2020Break; Break
IT WAS REPORTED THAT THE SHIM HAS BROKEN INTO MULTIPLE PIECES AND THIS WAS UNUSABLE. IT WAS NOT USED ON A PATIENT AND IT WILL BE RETURNED TO DEPUY. NO SURGICAL DELAY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10720683 ↗MalfunctionApr 1, 2020Break; Crack; Break; Crack
ADDITIONAL INFORMATION RECEIVED CONFIRMED THAT THE DEVICES DID NOT BREAK. ONLY A SMALL CRACK WAS FOUND ON EACH DEVICES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9909182 ↗MalfunctionFeb 14, 2020Crack; Detachment of Device or Device Component; Crack; Detachment of Device or Device Component
IT WAS REPORTED THAT THE SHIMS CRACKED AND NEED TO BE REPLACED. ONE OF THE BUSHINGS FELL OUT OF ONE OF THEM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9711935 ↗MalfunctionDec 11, 2019Break; Crack; Break; Crack
ALL ATTUNE SHIMS HAVE CRACKED AROUND THE METAL INSERT AREA.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9450532 ↗MalfunctionJul 24, 2019Break; Break
ATTUNE SHIMS HAVE ALL CRACKED AROUND THE METAL INSERT AREA.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8821734 ↗MalfunctionMay 23, 2019Break; Crack; Break; Crack
AFTER BEING USED DURING SURGERY, BEING WASHED AND STERILIZED SEVERAL TIMES, THE ATTUNE SHIMS SEEM TO BREAK EASILY. THE ITEMS BELOW NEED TO BE REPLACED IN (B)(6) HOSPITAL. DESCRIBE THE 'DAMAGE' OR 'OTHER' OUTCOME: NO DAMAGE TO PATIENTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8636479 ↗MalfunctionJan 19, 2018Break; Break
THERE IS A FISSURE IN THE WHITE PLASTIC. NO ADVERSE CONSEQUENCES FOR PATIENT OR USER. NO SURGERY DELAY REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7205149 ↗
Other TEMPLATE devices
- ATTUNE 2544-01-014Depuy (Ireland) · TEMPLATE802 reports
- ATTUNE 2545-00-546Depuy (Ireland) · TEMPLATE1 recall301 reports
- ARTICUL/EZE 2531-51-000Depuy (Ireland) · TEMPLATE323 reports
- DELTA XTEND 2307-95-000Depuy (Ireland) · TEMPLATE315 reports
- ATTUNE 2545-00-547Depuy (Ireland) · TEMPLATE1 recall231 reports
- ATTUNE 2544-00-520Depuy (Ireland) · TEMPLATE279 reports
- ATTUNE 2545-00-545Depuy (Ireland) · TEMPLATE1 recall221 reports
- ARTICUL/EZE 2530-91-000Depuy (Ireland) · TEMPLATE271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.