Skip to content

SUMMIT 2570-05-250Recalls & reported problems

Depuy (Ireland) · SUMMIT IMPLANT EXTRACTOR SLAP HAMMER · Class 1

Quick summary

This DePuy Summit slap hammer is a surgical tool used to remove implants. Its 102 FDA reports include 4 injuries and no deaths, and mostly describe threaded parts breaking off, sometimes inside other instruments, with some reports of bent parts and poor fit with other devices. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
100
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 100 adverse event reports that list this exact device (DI 10603295143239), received between Jul 21, 2017 and Apr 23, 2026 — classified as 4 injuries, 96 malfunctions.

The most frequently reported specific device problem is “Break” (66 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break66
  • Material Twisted/Bent18
  • Naturally Worn11
  • Device-Device Incompatibility10
  • Material Deformation5
  • Crack2
  • Adverse Event Without Identified Device or Use Problem1
  • Appropriate Device Problem Term/Code Not Available1

Patient problems

  • No Clinical Signs, Symptoms or Conditions54
  • No Consequences Or Impact To Patient39
  • No Code Available8
  • No Information5
  • Not Applicable4
  • Insufficient Information3

Recent adverse event reports

The 25 most recent of 100 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 23, 2026Break

    IT WAS REPORTED THAT THE SLAP HAMMER END WAS BROKE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24972003 ↗
  • MalfunctionMar 4, 2026Break

    IT WAS REPORTED THAT SLAP HAMMER BROKE OFF INSIDE CUP INSERTER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24505457 ↗
  • MalfunctionJan 6, 2026Break

    THE CLEAR OUT REAMER WAS DIFFICULT TO CHUCK UP AND HAS NOTICIBLE DAMAGE ON THE TIP. THE SLAP HAMMER BROKE OFF COMPLETELY WHEN WE TRIED TO REMOVE THE DISTAL STEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23974884 ↗
  • MalfunctionOct 21, 2025Break; Break

    IT WAS REPORTED THAT THE THREADED PART OF THE SLAP HAMMER WAS BROKEN OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23348334 ↗
  • MalfunctionOct 20, 2025Break; Break

    IT WAS REPORTED THAT THE THREADED PART OF THE SLAP HAMMER BROKE OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23335555 ↗
  • MalfunctionJun 30, 2025Break; Break

    IT WAS REPORTED THAT THE SLAP HAMMER ATTACHMENT POINT BROKE OFF INTO THE THREADED INSERTER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22356942 ↗
  • MalfunctionMar 19, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT RECLAIM STEM WAS REMOVED, UPON TRAY ASSEMBLY AND INSPECTION THE SLAP HAMMER AND RECLAIM STEM HANDLE WERE VISIBLY CROSS THREADED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21647609 ↗
  • MalfunctionFeb 19, 2025Break; Break

    IT WAS REPORTED THAT SLAP HANDLE FOR STEM EXTRACTOR SNAPPED AT THE THREADED PORTION. SURGICAL DELAY UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 21415102 ↗
  • MalfunctionFeb 19, 2025Break; Break

    IT WAS REPORTED THAT SLAP HANDLE FOR STEM EXTRACTOR SNAPPED AT THE THREADED PORTION. SURGICAL DELAY UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 21415098 ↗
  • MalfunctionDec 4, 2024Break; Break

    IT WAS REPORTED THAT THE SLAP HAMMER TIP BROKE OFF. UNKNOWN IF IT WAS ANY SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20848568 ↗
  • MalfunctionDec 4, 2024Break; Break

    IT WAS REPORTED THAT THE SLAP HAMMER TIP BROKE OFF. UNKNOWN IF IT WAS ANY SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20847724 ↗
  • MalfunctionNov 14, 2024Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    ADDITIONAL INFORMATION WAS RECEIVED AND STATED THAT THE THREAD SNAPPED OFF CLEAN AND THE REMAINING THREAD WAS INSIDE THE CORAIL BROACH HANDLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20685324 ↗
  • MalfunctionOct 22, 2024Break; Break

    ADDITIONAL INFORMATION RECEIVED AND STATING THAT: A. WHAT WAS THE EXACT ALLEGATION AGAINST THE DEVICE? WAS IT WORN, CRACKED, BROKEN INTO PIECES, BENT, STRIPPED, CROSS THREADED OR ANY DEVICE INTERACTION? T-HANDLE = CRACKED HANDLE; IMPLANT EXTRACTOR SLAPHAMMER = BROKEN THREAD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20507621 ↗
  • MalfunctionSep 26, 2024Break; Break

    ADDITIONAL INFORMATION WAS RECEIVED: B. WAS/WERE THERE ANY ADVERSE CONSEQUENCE/S THAT AFFECTED THE PATIENT BECAUSE OF THE REPORTED EVENT? NO. C. DID THE EVENT OCCUR DURING THE SURGERY? IF YES, WAS THERE A SURGICAL DELAY? WHAT IS THE DURATION OF THE DELAY? YES, BUT NO DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20311289 ↗
  • MalfunctionSep 16, 2024Break; Break

    ADDITIONAL INFORMATION WAS RECEIVED: WAS THERE ANY PATIENT CONSEQUENCES OR NOT? IF YES, PLEASE PROVIDE DETAILS. NO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20225484 ↗
  • MalfunctionJun 17, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE THREADS WERE DAMAGED ON THE SLAP HAMMER. SURGICAL DELAY WAS UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 19547011 ↗
  • MalfunctionApr 15, 2024Naturally Worn; Naturally Worn

    ADDITIONAL INFORMATION WAS RECEIVED. A. WAS THERE ANY ADVERSE CONSEQUENCES THAT AFFECTED THE PATIENT, BECAUSE OF THE REPORTED EVENT? NO. B. WAS THERE A SURGICAL DELAY? IF YES, WHAT IS THE DURATION OF THE DELAY? NO. C. WERE THERE ANY PIECES BROKEN OFF FROM THE INSTRUMENT? IF YES, DID IT BREAK INTO 2 OR MORE PIECES? NO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19112394 ↗
  • MalfunctionApr 8, 2024Break; Break

    THE THREADED PART OF THE SLAP HAMMER BROKE OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19057954 ↗
  • MalfunctionMar 7, 2024Break; Break

    IT WAS REPORTED THAT THE THREADED KNOB BROKE OFF THE SLAP HAMMER IN THE INSERTER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18855993 ↗
  • MalfunctionFeb 21, 2024Break; Break

    IT WAS REPORTED THAT SLAP HAMMER SCREW IN BROKE OFF. ALL BROKEN PIECES WERE RETRIEVED. THIS EVENT DID NOT SLOW DOWN THE SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18747968 ↗
  • MalfunctionOct 18, 2023Break; Break

    IT WAS REPORTED THAT THE THREADS FOR SLAP HAMMER BROKE OFF INSIDE THE THREADED STEM INSERTER WHILE TRYING TO EXTRACT A STEM. THE SLAP HAMMER BROKE INTO 2 PIECES AND ALL PIECES WERE RETRIEVED. THERE WERE NO ADVERSE EFFECTS FOR THE PATIENT AS A RESULT OF THIS EVENT AND SURGERY WAS NOT DEFINITIVELY EXTENDED AS A RESULT OF THIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17956415 ↗
  • MalfunctionOct 13, 2023Break; Break

    IT WAS REPORTED THAT CORAIL THREADED INSERTER AND SLAP HAMMER BROKE DURING EXTRACTION SURGERY. SURGICAL DELAY UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17929897 ↗
  • MalfunctionJul 24, 2023Break; Break

    IT WAS REPORTED THAT THE THREADED TIP BROKE OFF INSIDE THREADED EXTRACTION HANDLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17381772 ↗
  • MalfunctionJun 20, 2023Break; Break

    IT WAS REPORTED THAT THE SLAP HAMMER BROKE AND THE BROKEN THREADS WERE STUCK IN THE BROACH HANDLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17164440 ↗
  • MalfunctionJun 1, 2023Break; Break

    IT WAS REPORTED THAT THE THREADED END OF SLAP HAMMER WAS SHEARED OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17041138 ↗

Other HAMMER, SURGICAL devices

All HAMMER, SURGICAL devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.