Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10603295146346), received between Jul 21, 2017 and Apr 11, 2026 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility3
  • Appropriate Device Problem Term/Code Not Available2
  • Corroded1
  • Device Damaged by Another Device1
  • Illegible Information1
  • Material Deformation1
  • Material Twisted/Bent1
  • Naturally Worn1

Patient problems

  • No Consequences Or Impact To Patient6
  • No Clinical Signs, Symptoms or Conditions3
  • Insufficient Information1
  • No Information1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 11, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THESE PRODUCT HAD REACHED END OF LIFE DUE TO WEAR AND TEAR. THE SURGEONS COMPLAINED THAT INSTRUMENTS WERE DULL, ILLEGIBLE, AND GENERALLY WORN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24845929 ↗
  • MalfunctionApr 19, 2023Corroded; Corroded

    DEVICES APPEAR RUSTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16767561 ↗
  • MalfunctionSep 21, 2021Naturally Worn; Naturally Worn

    QUESTION: 1. WAS SURGERY TIME EXTENDED? IF YES, WHAT IS THE DURATION OF THE DELAY?" ANSWER: ON (B)(6) 2021 ADDITIONAL INFORMATION (2) : "SURGERY TIME WAS NOT EXTENDED." ON (B)(6) 2021 ADDITIONAL INFORMATION (3) : "SURGERY TIME WAS ONLY USED EXTENDED 2 MIN.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 12510958 ↗
  • MalfunctionFeb 20, 2020Illegible Information; Illegible Information

    IT WAS REPORTED THAT SROM INSTRUMENTS ARE DIFFICULT OR IMPOSSIBLE TO READ THE SIZING. SURGEON IS CONCERNED THIS IS A PATIENT SAFETY ISSUE AND INSTRUMENTS SHOULD BE REPLACED. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9734656 ↗
  • MalfunctionJul 2, 2019Device-Device Incompatibility; Material Deformation; Material Twisted/Bent; Scratched Material; Device-Device Incompatibility; Material Deformation; Material Twisted/Bent; Scratched Material

    ADDITIONAL INFORMATION RECEIVED INDICATING THAT THE FEATURE THAT WAS WORN AND NOT MAKING IT FUNCTION AS INTENDED WAS THE TOP PORTION THAT ACCEPTS THE NECK SEGMENT DOES NOT FUNCTION CORRECTLY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8753993 ↗
  • MalfunctionMay 20, 2019Device-Device Incompatibility; Device-Device Incompatibility

    ADDITIONAL INFORMATION RECEIVED INDICATING THAT NO BROKEN PIECES WERE LEFT IN THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8624797 ↗
  • MalfunctionFeb 8, 2019Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE TRIAL NECK WAS NO LONGER ATTACHED TO THE STEM TRIAL. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8323532 ↗
  • MalfunctionAug 21, 2018Device Damaged by Another Device; Device Damaged by Another Device

    IT WAS REPORTED THAT THE TRIAL 15MM STEM AND THE 36+8 NECK SEGMENT ARE BOTH STRIPPED AND WOULD NOT ENGAGE. THE STEM/SLEEVE WEDGES DAMAGED. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7802402 ↗
  • MalfunctionJul 11, 2018Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE NECK TRIAL WOULD NOT SECURELY FIT ON THE STEM TRIAL. NECK TRIAL WOULD SPIN AROUND ON THE STEM TRIAL. NO SURGICAL.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7676922 ↗
  • MalfunctionJul 21, 2017Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE THREADS ON STEM TRIALS ARE DAMAGED. IT WAS ALSO STATED THAT THE NECK TRIALS DO NOT LOCK DOWN EASY.

    Patient problems: No Information

    MAUDE report 6731174 ↗