Skip to content

S-ROM 2576-20-100Recalls & reported problems

Depuy (Ireland) · S-ROM ZT TRIAL SLEEVE SIZE 20 B SMALL · Class 1

All models of S-ROM (185)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10603295146735), received between Dec 11, 2019 and Mar 23, 2026 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Illegible Information” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Illegible Information3
  • Material Twisted/Bent3
  • Scratched Material3
  • Device-Device Incompatibility2
  • Material Deformation2
  • Naturally Worn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Consequences Or Impact To Patient4

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 23, 2026Illegible Information

    IT WAS REPORTED THAT THE TOPS OF THE SLEEVE TRIALS HAVE THE SIZES SCRAPED OFF. THIS DID NOT OCCUR INTRAOPERATIVELY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24665309 ↗
  • MalfunctionJun 16, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE INSTRUMENT WAS DIFFICULT TO EXTRACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22210893 ↗
  • MalfunctionJan 11, 2024Material Deformation; Scratched Material; Material Deformation; Scratched Material

    THIS INSTRUMENT IS COMPOSED OF A 1 PART. THE ZT TRIAL SLEEVES HAVE A FEW DENTS WITH SHARP SCRATCHES. THE PRODUCT DAMAGE DOCUMENTED IN THIS PC HAS BEEN IDENTIFIED DURING LOAN KIT INSPECTION, BY THE LOAN KIT TECHNICIAN. THE EVENT DATE AND ALERT DATE ARE THE DATE THAT THE INSPECTION TOOK PLACE. THERE IS NO SURGEON, PROCEDURE, OR PATIENT DETAILS AVAILABLE. NO FURTHER INFORMATION CAN BE OBTAINED AS THE CASE WAS NOT REPORTED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18493676 ↗
  • MalfunctionJun 9, 2023Device-Device Incompatibility; Naturally Worn; Device-Device Incompatibility; Naturally Worn

    IT WAS REPORTED THAT THE BOTTOM PORTION OF THE SLEEVE TRIAL HAS WORN, MAKING IT DIFFICULT TO REMOVE WITH THE SLEEVE TRIAL EXTRACTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17098421 ↗
  • MalfunctionFeb 27, 2020Material Twisted/Bent; Scratched Material; Illegible Information; Material Twisted/Bent; Scratched Material; Illegible Information

    IT WAS REPORTED THAT THE S-ROM SLEEVE TRIALS ARE DIFFICULT TO IMPOSSIBLE TO READ SIZING DUE TO WEAR AND TEAR AND NORMAL EVERYDAY USE. THE NEED TO BE REPLACED SO THAT THE SURGEON IS ABLE TO SAFELY DETERMINE THE SIZE OF THE TRIAL BEING USED. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9765950 ↗
  • MalfunctionFeb 25, 2020Material Twisted/Bent; Illegible Information; Material Twisted/Bent; Illegible Information

    IT WAS REPORTED THAT THE S-ROM SLEEVE TRIALS ARE EITHER DIFFICULT TO IMPOSSIBLE TO READ THE SIZE. THIS IS A SAFETY CONCERN FOR THE SURGEON WHO CAN'T EASILY DETERMINE THE SIZE OF THE TRIAL SLEEVE BEING INSERTED. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9752931 ↗
  • MalfunctionJan 2, 2020Scratched Material; Scratched Material

    IT WAS REPORTED THAT THERE WERE CUTS IN SLEEVE FOUND WHEN SET RETURNED TO THE OFFICE FOR INSPECTION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9540439 ↗
  • MalfunctionDec 11, 2019Material Deformation; Material Twisted/Bent; Material Deformation; Material Twisted/Bent

    IT WAS REPORTED THAT THE SLEEVE TRIALS ALONG THE DISTAL EDGE HAD BEEN WORN AND BENT WHERE THEY CANNOT BE EXTRACTED AND GETTING STUCK IN THE FEMUR. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9450491 ↗

Other TEMPLATE devices

All TEMPLATE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page