Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10603295146766), received between Sep 29, 2017 and Apr 17, 2026 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Illegible Information” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Illegible Information4
  • Appropriate Device Problem Term/Code Not Available3
  • Material Deformation2
  • Naturally Worn2
  • Scratched Material2
  • Device-Device Incompatibility1
  • Insufficient Device Problem Information1
  • Material Discolored1

Patient problems

  • No Consequences Or Impact To Patient6
  • No Clinical Signs, Symptoms or Conditions2
  • No Information2

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 17, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THESE PRODUCT HAVE REACHED END OF LIFE DUE TO WEAR AND TEAR. THE SURGEONS WERE COMPLAINING ABOUT INSTRUMENT BEING DULL, ILLEGIBLE, AND GENERALLY WORN. THERE WILL BE MULTIPLE PAGES ON THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24909355 ↗
  • MalfunctionMar 23, 2026Illegible Information

    IT WAS REPORTED THAT THE TOPS OF THE SLEEVE TRIALS HAVE THE SIZES SCRAPED OFF. THIS DID NOT OCCUR INTRAOPERATIVELY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24665391 ↗
  • MalfunctionApr 3, 2020Scratched Material; Illegible Information; Scratched Material; Illegible Information

    IT WAS REPORTED THAT THE SROM SLEEVES COULD NOT BE READ OR USED OR DAMAGED. NOT USED IN A CASE. LOT NUMBERS COULD NOT BE READ. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9924116 ↗
  • MalfunctionJul 1, 2019Device-Device Incompatibility; Material Deformation; Material Twisted/Bent; Scratched Material; Device-Device Incompatibility; Material Deformation; Material Twisted/Bent; Scratched Material

    IT WAS REPORTED THAT THE SLEEVE IS BENT AND WILL NOT SLIDE OVER THE STEM. THE STEM WILL NOT ALLOW THE NECK SEGMENT TO SCREW IN.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8748565 ↗
  • MalfunctionMar 29, 2019Material Discolored; Material Deformation; Naturally Worn; Material Discolored; Material Deformation; Naturally Worn

    IT WAS REPORTED THAT THE OFFICE SETS CANNOT READ THE NUMBERS. IT WAS WORN OUT. DID NOT EFFECT A SURGERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8466459 ↗
  • MalfunctionMar 29, 2019Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE OFFICE SETS CANNOT READ THE NUMBERS. IT WAS WORN OUT. DID NOT EFFECT A SURGERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8466455 ↗
  • MalfunctionApr 11, 2018Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT IDENTIFYING MARKINGS FOUND TO BE DAMAGED IN POST CLEANING INSPECTION, IT IS UNREADABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7421581 ↗
  • MalfunctionDec 11, 2017Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT ALL OF THE 10 SROM INSTRUMENTS WERE DEEMED TO BE UNUSABLE BECAUSE THE SIZE MARKINGS ARE NO LONGER VISIBLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7103843 ↗
  • MalfunctionSep 29, 2017Appropriate Device Problem Term/Code Not Available; Illegible Information; Appropriate Device Problem Term/Code Not Available; Illegible Information

    IT WAS REPORTED THAT THE ZT TRIAL SLEEVES ETCHING COMING OFF. IT WAS FOUND IN THE OFFICE SET.

    Patient problems: No Information

    MAUDE report 6903565 ↗
  • MalfunctionSep 29, 2017Appropriate Device Problem Term/Code Not Available; Illegible Information; Appropriate Device Problem Term/Code Not Available; Illegible Information

    IT WAS REPORTED THAT THE ZT TRIAL SLEEVES ETCHING COMING OFF. IT WAS FOUND IN THE OFFICE SET.

    Patient problems: No Information

    MAUDE report 6903536 ↗