Skip to content

S-ROM 2576-36-008Recalls & reported problems

Depuy Orthopaedics, Inc. · S-ROM NECK TRIAL 36 STD +8 LAT · Class 1

All models of S-ROM (92)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
18
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 10603295147077), received between Jan 24, 2018 and Nov 5, 2024 — classified as 18 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility9
  • Illegible Information5
  • Naturally Worn5
  • Scratched Material4
  • Material Twisted/Bent3
  • Material Deformation2
  • Connection Problem1
  • Device Damaged by Another Device1

Patient problems

  • No Consequences Or Impact To Patient10
  • No Clinical Signs, Symptoms or Conditions6
  • Insufficient Information1
  • No Code Available1
  • Not Applicable1

Recent adverse event reports

The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 5, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE NECK SEGMENTS DOES NOT SCREW INTO STEM PROPERLY. THERE WAS NO ADVERSE CONSEQUENCE AND NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20613338 ↗
  • MalfunctionSep 12, 2023Device-Device Incompatibility; Naturally Worn; Device-Device Incompatibility; Naturally Worn

    IT WAS REPORTED THAT 36+8 SROM NECK TRIAL WILL NOT PROPERLY THREAD ON TO THE STEM TRIAL. THREADS APPEAR TO BE WORN OR STRIPPED. SURGICAL DELAY UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 17729012 ↗
  • MalfunctionJul 6, 2022Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE NECK NO LONGER TIGHTENS DOWN ON THE STEM, OTHER NECKS WORK ON THE STEM BUT THIS ONE DOESN'T WORK ON ANY. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14935322 ↗
  • MalfunctionFeb 21, 2022Device-Device Incompatibility; Device-Device Incompatibility

    NECK TRIAL WOULD NOT CONNECT TO THE STEM TRIAL. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13574058 ↗
  • MalfunctionSep 30, 2021Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE TRIAL NECK DOESN¿T LOCK DOWN TO THE STEM. BOTH THE STEM AND NECK HAVE BEEN REQUESTED TO BE REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12557709 ↗
  • MalfunctionJul 9, 2021Device-Device Incompatibility; Material Twisted/Bent; Naturally Worn; Device-Device Incompatibility; Material Twisted/Bent; Naturally Worn

    IT WAS REPORTED THAT THE NECK TRIAL IS STRIPPED. SURGEON COULD NOT GET THE NECK TRIAL TO LOCK DOWN TO THE STEM TRIAL. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12145497 ↗
  • MalfunctionMar 25, 2021Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE NECK TRIALS NO LONGER ATTACH TO THE STEMS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11570724 ↗
  • MalfunctionApr 23, 2020Material Twisted/Bent; Naturally Worn; Illegible Information; Material Twisted/Bent; Naturally Worn; Illegible Information

    IT WAS REPORTED THAT THESE INSTRUMENTS HAVE EXCESSIVE WEAR AND ARE DIFFICULT TO READ. REP INDICATED "THAT THEY NEED TO BE REPLACED". NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9992124 ↗
  • MalfunctionApr 3, 2020Scratched Material; Illegible Information; Scratched Material; Illegible Information

    IT WAS REPORTED THAT THE INSTRUMENTS ARE SCRATCHED UP OR DAMAGED AND SIZES CAN¿T BE READ OR IDENTIFIED EASILY. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9919387 ↗
  • MalfunctionApr 3, 2020Scratched Material; Illegible Information; Scratched Material; Illegible Information

    IT WAS REPORTED THAT THE INSTRUMENTS ARE SCRATCHED UP OR DAMAGED AND SIZES CAN¿T BE READ OR IDENTIFIED EASILY. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9919386 ↗
  • MalfunctionApr 3, 2020Scratched Material; Illegible Information; Scratched Material; Illegible Information

    IT WAS REPORTED THAT THE INSTRUMENTS ARE SCRATCHED UP OR DAMAGED AND SIZES CAN¿T BE READ OR IDENTIFIED EASILY. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9919385 ↗
  • MalfunctionFeb 19, 2020Material Twisted/Bent; Naturally Worn; Scratched Material; Illegible Information; Material Twisted/Bent; Naturally Worn; Scratched Material; Illegible Information

    IT WAS REPORTED THAT THE INSTRUMENT WERE DIFFICULT OR IMPOSSIBLE TO READ ETCHING DUE TO WEAR AND TEAR OVER THE YEARS. THE SURGEON IS CONCERNED ABOUT THE PATIENT SAFETY ISSUE AND THE INSTRUMENTS NEED TO BE REPLACED.

    Patient problems: No Code Available; No Consequences Or Impact To Patient

    MAUDE report 9729053 ↗
  • MalfunctionJan 23, 2020Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE SROM MACH 1 36+8 LAT NECK TRIAL ITEM WILL NOT LOCK DOWN TIGHTLY ON TRIAL BODY - POSSIBLY DAMAGED. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9624942 ↗
  • MalfunctionJul 1, 2019Device-Device Incompatibility; Device-Device Incompatibility

    THREAD ON THE INSIDE OF THE S-ROM 36+8 NECK TRIAL WAS FAULTY AND WOULDN'T ENGAGE WITH THE STEM TRIAL.

    Patient problems: Not Applicable

    MAUDE report 8750480 ↗
  • MalfunctionJan 8, 2019Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE SROM 36+8 TRIAL NECK IS STRIPPED AND WON'T TIGHTEN DOWN ON TRIAL STEM. NEEDS REPLACING. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8227446 ↗
  • MalfunctionAug 21, 2018Device Damaged by Another Device; Device Damaged by Another Device

    IT WAS REPORTED THAT THE TRIAL 15MM STEM AND THE 36+8 NECK SEGMENT ARE BOTH STRIPPED AND WOULD NOT ENGAGE. THE STEM/SLEEVE WEDGES DAMAGED. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7802426 ↗
  • MalfunctionMar 23, 2018Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE 36+8 TRIAL NECK ON SROM WILL NOT TIGHTEN PROPERLY. NO PIECES BROKEN AND NO PATIENT INVOLVEMENT. NEEDS TO BE REPLACED. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7365127 ↗
  • MalfunctionJan 24, 2018Connection Problem; Material Deformation; Connection Problem; Material Deformation

    IT WAS REPORTED THAT THE THREADS ON 36STD+8LAT NECK TRIAL ARE STRIPPED. NEEDS REPLACING. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7217004 ↗

Other TEMPLATE devices

All TEMPLATE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page