RECLAIM 2975-00-100Recalls & reported problems
Depuy (Ireland) · RECLAIM PROXIMAL REAMER GUIDE POST · Class 1
All models of RECLAIM (149)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 10603295155850), received between Apr 12, 2018 and Jan 26, 2026 — classified as 13 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device-Device Incompatibility4
- Material Twisted/Bent3
- Naturally Worn3
- Appropriate Device Problem Term/Code Not Available2
- Break2
- Material Deformation2
- Corroded1
- Material Discolored1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- Insufficient Information3
- No Consequences Or Impact To Patient2
- No Code Available1
- Not Applicable1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 26, 2026Material Deformation
IT WAS REPORTED THAT DURING A HIP SURGERY, THE RECLAIM MODULAR INSTRUMENTS WERE STRIPPED OUT AND NON-FUNCTIONAL. THE SURGEON HAD TO TRIAL OFF REAL IMPLANTS. THERE WAS NO PATIENT HARM, BUT THE PROCEDURE WAS DELAYED BY NINETY MINUTES DUE TO THE INABILITY TO PROPERLY TRIAL AND STRUGGLE WITH THE INSTRUMENTS. ACTIVITY LEVEL ¿ YES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24182995 ↗MalfunctionMay 8, 2024Break; Break
IT WAS REPORTED THAT THE INSTRUMENT WAS BROKEN.
Patient problems: Insufficient Information
MAUDE report 19268522 ↗MalfunctionDec 14, 2022Device-Device Incompatibility; Naturally Worn; Device-Device Incompatibility; Naturally Worn
IT WAS REPORTED THAT THE ITEMS WILL NOT THREAD, SURGEON COMPLAIN ABOUT THREADS STRIPPING ON ALL ITEMS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15983825 ↗MalfunctionDec 14, 2022Device-Device Incompatibility; Naturally Worn; Device-Device Incompatibility; Naturally Worn
IT WAS REPORTED THAT THE ITEMS WILL NOT THREAD, SURGEON COMPLAIN ABOUT THREADS STRIPPING ON ALL ITEMS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15983820 ↗MalfunctionAug 11, 2022Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT DURING AN OFF TABLE DIRECT ANTERIOR TOTAL HIP THE ACTIS BROACH PERFORATED THE FEMORAL SHAFT, CAUSING A TROCH FRACTURE. THE DOCTOR ELECTED TO CABLE THE FEMUR AND SWITCH TO A RECLAIM STEM. THE INCISION WAS EXTENDED. THE EXPOSURE WAS MADE DIFFICULT DUE TO THE PATIENTS LARGE ABDOMEN. THE FEMUR WAS PREPPED WITH THE RECLAIM INSTRUMENTS AND A 15X140 STEM WAS PLACED. WHEN REMOVING THE STEM FROM THE INSERTER THE STEM THREADS WERE STRIPPED AND CROSS THREADED. MULTIPLE ATTEMPTS WHERE MADE TO ATTACH THE STEM EXTRACTOR WITH THE RECLAIM INSTRUMENTS AND A UNIVERSAL EXTRACTOR SET. THESE ALL FAILED. THE DOCTOR THEN ELECTED TO SEAT THE PROXIMAL BODY BUT WAS UNABLE TO THREAD THE ASSEMBLY TOOL INTO THE STEM. SURGEON ELECTED TO PLACE THE PROXIMAL BODY AND IMPACT IT MANUALLY. THE LOCKING BOLT WAS ATTEMPTED. IT WAS UNABLE TO THREAD INTO THE STEM. THE DOCTOR REQUESTED THAT ALL THE THREADED INSTRUMENTS THAT WERE USED BE REPLACED. SURGERY TIME WAS EXTENDED BY 2 HOURS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15209841 ↗MalfunctionJan 11, 2022Naturally Worn; Naturally Worn
IT WAS REPORTED THAT THE RECLAIM DISTAL STEM INSERTER WAS CROSS THREADED. RECLAIM PROX REAMER GUIDE POST WAS CROSS THREADED. SURGICAL DELAY OF 3 MINUTES AND NEEDS REPLACEMENT. NO ADDITIONAL INFORMATION.
Patient problems: Insufficient Information
MAUDE report 13226213 ↗MalfunctionAug 17, 2021Device-Device Incompatibility; Device-Device Incompatibility
ADDITIONAL INFORMATION RECEIVED INDICATED THAT THERE WERE NO PIECES BROKEN OFF FROM THE INSTRUMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12331601 ↗MalfunctionJan 8, 2021Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT THE SEVERAL RECLAIM INSTRUMENTS WITH THREADS WERE DAMAGED DURING NORMAL USE. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11142667 ↗MalfunctionJan 8, 2021Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT THE SEVERAL RECLAIM INSTRUMENTS WITH THREADS WERE DAMAGED DURING NORMAL USE. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11141666 ↗MalfunctionJan 8, 2021Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT THE SEVERAL RECLAIM INSTRUMENTS WITH THREADS WERE DAMAGED DURING NORMAL USE. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11141646 ↗MalfunctionDec 12, 2018Material Deformation; Material Deformation
IT WAS REPORTED THAT THE INSTRUMENTS WON¿T THREAD PROPERLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8155025 ↗MalfunctionSep 6, 2018Break; Corroded; Material Discolored; Appropriate Device Problem Term/Code Not Available; Break; Corroded; Material Discolored; Appropriate Device Problem Term/Code Not Available
AFTER RECEIVING 2 COMPLAINTS ABOUT A HIP KIT, ORTHOKIT MANAGER CHECKED THE BUILD 5 IN THE WAREHOUSE, AND SAW RUST ON ALMOST EVERY PART OF THE KIT. FOR SOME PARTS IT'S ALSO MISSING COLOR CODES (USUALLY RED, YELLOW, GREEN, BLUE). THIS KIT WENT TO A CUSTOMER IN (B)(6) 2016 AND WAS NEW AT THE TIME. IT WAS CHECKED WHEN IT CAME BACK IN 2016 AND WAS LET IN STORAGE SINCE THEN.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7850156 ↗MalfunctionApr 12, 2018Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
DER STATES THAT THE GUIDE POST WOULDN¿T PROPERLY THREAD INTO THE DISTAL RECLAIM STEM, AND THE SURGEON FELT IT DAMAGED THE THREADS INSIDE THE STEM. THE ASSEMBLY TOOL ALSO WOULDN¿T THREAD INTO THE DISTAL STEM PROPERLY, SO THE SURGEON DECIDED TO PUT IN A NEW DISTAL STEM. THE 18X140 WAS COMPROMISED AND COULDN¿T BE USED. NO SURGICAL DELAY.
Patient problems: Insufficient Information; No Code Available; Not Applicable
MAUDE report 7424706 ↗
Other TEMPLATE devices
- ATTUNE 2544-01-014Depuy (Ireland) · TEMPLATE802 reports
- ATTUNE 2545-00-546Depuy (Ireland) · TEMPLATE1 recall301 reports
- ARTICUL/EZE 2531-51-000Depuy (Ireland) · TEMPLATE323 reports
- DELTA XTEND 2307-95-000Depuy (Ireland) · TEMPLATE315 reports
- ATTUNE 2545-00-547Depuy (Ireland) · TEMPLATE1 recall231 reports
- ATTUNE 2544-00-520Depuy (Ireland) · TEMPLATE279 reports
- ATTUNE 2545-00-545Depuy (Ireland) · TEMPLATE1 recall221 reports
- ARTICUL/EZE 2530-91-000Depuy (Ireland) · TEMPLATE271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page