Skip to content

CORAIL 3L92505Recalls & reported problems

Depuy (Ireland) · CORAIL HIP SYSTEM CEMENTLESS FEMORAL STEM 12/14 AMT 135 DEGREES KA SIZE 15 HA COATED STANDARD COLLAR · Class 2 · Implant

Quick summary

This DePuy CORAIL cementless hip stem is used in hip replacement surgery. Most of the 27 reports describe the stem failing to grow properly into the bone, or infection and pain that led surgeons to remove or revise the implant, along with a fracture and one report of the stem arriving damaged. Reports mention 24 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
27
Deaths reported
0
Injuries reported
24
Recalls naming it
0

Summary

The FDA's MAUDE database contains 27 adverse event reports that list this exact device (DI 10603295168713), received between Jul 26, 2017 and Apr 17, 2026 — classified as 24 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Loss of Osseointegration” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem13
  • Loss of Osseointegration4
  • Insufficient Device Problem Information2
  • Osseointegration Problem2
  • Biocompatibility1
  • Corroded1
  • Degraded1
  • Device-Device Incompatibility1

Patient problems

  • No Code Available10
  • Unspecified Infection9
  • Pain7
  • Inadequate Osseointegration6
  • Test Result3
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • Osteolysis2
  • Tissue Damage2

Recent adverse event reports

The 25 most recent of 27 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 17, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT REMOVED COMPONENTS DUE TO FX, DISLOCATION, AND NON COMPLIANCE. DOI- (B)(6) 2026. DOR-UNKNOWN. AFFECTED SIDE-RIGHT HIP.

    Patient problems: Limb Fracture

    MAUDE report 24915200 ↗
  • InjuryOct 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT REMOVAL DUE TO INFECTION. AFFECTED SIDE: LEFT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 23367209 ↗
  • InjuryJun 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT UNDERWENT REVISION AND COMPONENTS WERE REVISED DUE TO INFECTION. DOI: (B)(6) 2024. DOR: (B)(6) 2024. AFFECTED SIDE: UNKNOWN HIP.

    Patient problems: Unspecified Infection

    MAUDE report 22292695 ↗
  • InjuryJan 15, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT DEVELOPED INFECTION APPROXIMATELY TWO MONTHS POSTOP. THEY REMOVED ALL COMPONENTS AND IMPLANTED CEMENT SPACERS. DOI: UNKNOWN. AFFECTED SIDE: LEFT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 21160044 ↗
  • InjuryDec 21, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ADDITIONAL INFORMATION RECEIVED STATES THAT THERE WAS NO SURGICAL DELAY.

    Patient problems: Unspecified Infection

    MAUDE report 18371831 ↗
  • InjuryOct 19, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO PAIN. IT WAS UNKNOWN, IF THERE WAS A SURGICAL DELAY. DOI- (B)(6) 2022. DOR- (B)(6) 2023. AFFECTED SIDE- LEFT HIP.

    Patient problems: Pain

    MAUDE report 17967252 ↗
  • MalfunctionDec 9, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT THE PRODUCT WAS RECEIVED DAMAGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15950340 ↗
  • MalfunctionJul 18, 2022Difficult to Open or Remove Packaging Material; Difficult to Open or Remove Packaging Material

    IT WAS REPORTED THAT ON (B)(6) 2022, THE PATIENT UNDERWENT AN UNKNOWN SURGERY WITH THE CORAIL AMT STEM. THE SURGEON COMPLAINED REGARDING THE STYROFOAM ATTACHED TO THE NECK OF THE STEM AS PROTECTION. THE STYROFOAM WAS ATTACHED FIRMLY THAT IT COULD NOT BE EASILY REMOVED BY HAND AND IS DIFFICULT TO REMOVE. IN ADDITION, IT IS DIFFICULT TO REMOVE THE NECK WHEN THE SURGEON¿S HAND IS SLIPPERY DURING SURGERY, AND THERE IS A RISK OF DAMAGING THE NECK PART WHEN PICKING IT UP WITH A COCKEREL AND OTHER INSTRUMENTS. THEREFORE, THE SURGEON ASKED THE SALES REP TO IMPROVE ALL THE IMPLANTS PROTECTION. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 15045034 ↗
  • InjuryAug 24, 2021Osseointegration Problem; Osseointegration Problem

    THE PATIENT WAS REVISED TO ADDRESS LOOSENING OF THE STEM AT THE BONE TO IMPLANT INTERFACE. THE PATIENT COMPLAINS OF PAIN IN THE FEMUR. XRAYS SHOW STEM WELL-FIXED DISTALLY, BUT LUCENT LINES PROXIMALLY INDICATING LOOSE STEM. PATIENT WAS TAKEN TO OR TO FIND STEM LOOSE PROXIMALLY AND WELL-FIXED DISTALLY. AN OSTEOTOMY WAS REQUIRED TO REMOVE AFTER SEVERAL FAILED ATTEMPTS TO REMOVE FROM ABOVE. FROZEN SECTION DID NOT SHOW SIGNS OF INFECTION. SURGEON ALSO CHOOSE TO CONVERT TO DUAL MOBILITY CONSTRUCT TO ENHANCE STABILITY DUE TO INCREASED RISK AFTER REVISION PROCEDURE. DOI: (B)(6) 2020, DOR: (B)(6) 2021, AFFECTED SIDE: RIGHT HIP.

    Patient problems: Inadequate Osseointegration; Pain

    MAUDE report 12362097 ↗
  • InjuryApr 22, 2021Adverse Event Without Identified Device or Use Problem; Osseointegration Problem; Adverse Event Without Identified Device or Use Problem; Osseointegration Problem

    ADDITIONAL INFORMATION RECEIVED INDICATED THAT THE HIP STEM WAS LOOSE AND LOOSENING WAS DUE TO INFECTION.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Inadequate Osseointegration

    MAUDE report 11709900 ↗
  • InjuryApr 7, 2021Degraded; Degraded

    22 MAR 2021. MEDICAL RECORD REVIEWED. MEDICAL RECORDS RECEIVED WERE REVIEWED TO IDENTIFY PATIENT HARMS/PRODUCT ISSUES. ON (B)(6) 2020, THE PATIENT UNDERWENT A RIGHT HIP REVISION TO ADDRESS OSTEOLYSIS, ELEVATED METAL LEVELS (VALUES NOT GIVEN), EARLY PSEUDOTUMOR, AND LIMB ASYMMETRY. THE SURGEON NOTED SYNOVITIS AND REACTIVE CHANGE IN THE SOFT TISSUE, AS WELL AS GRAY DISCOLORATION CONSISTENT WITH METAL DEBRIS. THE TRUNNION WAS CLEANED WITH A SCRATCH PAD TO CLEAR ALL CORROSION. MASSIVE OSTEOLYSIS WAS NOTED AROUND THE FEMUR, HOWEVER, THE FEMUR WAS VERY WELL INTACT AND RETAINED ALONG WITH THE CUP. THE HEAD AND LINER WERE REVISED. THERE WERE NO INDICATED INTRA-OPERATIVE COMPLICATIONS. TASK CREATED TO UPDATE PC.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Hypersensitivity/Allergic reaction; Osteolysis; Physical Asymmetry; Synovitis

    MAUDE report 11631645 ↗
  • InjuryNov 6, 2020Loss of Osseointegration; Loss of Osseointegration

    IT WAS REPORTED THAT THE PATIENT RECENTLY HAD SURGERY AND DEVELOPED AN INFECTION SHORTLY AFTER. SHE CAME BACK IN FOR AN I&D, POTENTIAL POLY/HEAD SWAP. UPON OPENING THE JOINT AND TRYING TO REMOVE THE HEAD THE STEM BECAME LOOSE AT BONE TO IMPLANT INTERFACE. AT THIS POINT THE SURGEON DECIDED TO REMOVE ALL IMPLANTS AND PUT A TEMPORARY SPACER UNTIL THE INFECTION CLEARS. DOI: (B)(6) 2020 - DOR: (B)(6) 2020 (LEFT HIP).

    Patient problems: Inadequate Osseointegration; No Code Available; Unspecified Infection

    MAUDE report 10801912 ↗
  • InjuryJul 20, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT FEMUR FRACTURED AND DISLOCATED SO THE DOCTOR REPLACES HEAD AND STEM WITH NEW REVISION STEM AND HEAD. DOI: (B)(6) 2020, DOR: (B)(6) 2020, RIGHT SIDE.

    Patient problems: Bone Fracture(s); No Code Available

    MAUDE report 10295861 ↗
  • InjuryOct 22, 2019Loss of Osseointegration; Loss of Osseointegration

    THE PATIENT WAS REVISED TO ADDRESS PERIPROSTHETIC FRACTURE AFTER A FALL AND LOOSENING OF THE STEM AT THE BONE TO IMPLANT INTERFACE. DOI: (B)(6) 2019; DOR: (B)(6) 2019; LEFT HIP.

    Patient problems: Inadequate Osseointegration; No Code Available

    MAUDE report 9218036 ↗
  • InjuryJul 19, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED TO ADDRESS INFECTION. DOI: (B)(6) 2019; DOR: (B)(6) 2019, RIGHT HIP.

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 8807066 ↗
  • InjuryJun 18, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ADDITIONAL INFORMATION RECEIVED THAT THE SURGERY DATE WAS UNKNOWN AND THE AFFECTED SIDE INVOLVED WAS THE LEFT SIDE.

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 8708725 ↗
  • InjuryJun 7, 2019Loss of Osseointegration; Loss of Osseointegration

    IT WAS REPORTED THAT THE PATIENT WAS PRESENTED WITH A PAINFUL HIP. SURGEON REMOVED A LOOSE CORAIL STEM, CUP WAS WELL FIXED. SURGEON REPLACED STEM AND CHANGED LINER WHILE IN SITU. DOI: (B)(6) 2012, DOR: (B)(6) 2019, LEFT SIDE.

    Patient problems: Inadequate Osseointegration; No Code Available; Pain

    MAUDE report 8678950 ↗
  • InjuryMay 23, 2019Corroded; Corroded

    CLINICAL NOTIFICATION RECEIVED FOR REVISION OF RIGHT TOTAL HIP TO ADDRESS ADVERSE LOCAL TISSUE REACTION (ALTR). DOI: (B)(6) 2008; DOR: (B)(6) 2019, (RIGHT HIP).

    Patient problems: No Code Available; Osteolysis; Pain; Tissue Damage

    MAUDE report 8636930 ↗
  • InjuryMay 16, 2019Fracture; Fracture

    CORAIL STEM BREAKAGE. PATIENT WEIGHT: (B)(6); PRIMARY SURGERY IN 2016; STEM SIZE: 15; X-RAY UPLOADED TO ECM.

    Patient problems: No Code Available

    MAUDE report 8616959 ↗
  • InjuryMar 8, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED TO ADDRESS CONFIRMED FOR INFECTION. ALL HIP HEMI IMPLANTS WERE REMOVED AND AN ANTIBIOTIC SPACER WAS PLACED. DOI: (B)(6) 2019; DOR: (B)(6) 2019; RIGHT HIP.

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 8404732 ↗
  • InjuryDec 28, 2018Biocompatibility; Biocompatibility

    LITIGATION ALLEGES THE PATIENT SUFFERS FROM HIGHLY ELEVATED BLOOD METAL LEVELS, PAIN, POPPING, SQUEAKING AND DIFFICULTY AMBULATING. UPDATE AD 07 MAY 2018 RECEIPT OF PPF AND STICKER SHEETS. IN ADDITION TO WHAT WERE PREVIOUSLY ALLEGED, PPF ALLEGES METAL WEAR, METALLOSIS AND ELEVATED METAL IONS.

    Patient problems: Pain; Test Result; Tissue Damage

    MAUDE report 8202306 ↗
  • InjuryDec 3, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED TO ADDRESS PAIN. THIS IS A CLINICAL PATIENT, REPORT STATES THAT THE PATIENT HAD METALLOSIS, AND PSEUDOTUMOR. UPDATE PPF ALLEGES PSEUDOTUMOR AND ELEVATED METAL IONS. DOI: (B)(6) 2008 - DOR: (B)(6) 2012 (RIGHT HIP).

    Patient problems: No Code Available; Test Result

    MAUDE report 8125282 ↗
  • MalfunctionNov 2, 2018Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE TWO PIECE STEM INSERTER, WHICH THREADS INTO STEM, COULD NOT UN-THREAD ONCE THE STEM WAS IMPACTED. WHEN THE STEM WAS EXTRACTED, IT WAS NOT POSSIBLE TO REMOVE THE STEM FROM THE INSERTER.

    Patient problems: Not Applicable

    MAUDE report 8034193 ↗
  • InjurySep 25, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PPF AND IMPLANT RECORDS RECEIVED. PPF ALLEGED ELEVATED METAL IONS AFTER FIRST REVISION. DOI: (B)(6) 2013 - DOR: NONE REPORTED (LEFT HIP).

    Patient problems: Test Result

    MAUDE report 7905264 ↗
  • InjuryJul 23, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    THE PATIENT WAS REVISED TO ADDRESS ELEVATED METAL IONS AND PAIN. DOI: 2009; DOR: (B)(6) 2018; RIGHT HIP.

    Patient problems: Pain

    MAUDE report 7710279 ↗

Other Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate devices

All Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.