Quick summary

This DePuy CORAIL cementless hip stem is used in total hip replacement. The 54 reports mostly describe the stem loosening or failing to grow into the bone, along with pain, infection, and metal debris, often leading to revision surgery; 50 injuries and 1 death are reported. There are no recalls naming this device, and the report count is flat at 1 in each of the last two 12-month periods.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
54
Deaths reported
1
Injuries reported
50
Recalls naming it
0

Summary

The FDA's MAUDE database contains 54 adverse event reports that list this exact device (DI 10603295168898), received between Jul 21, 2017 and Dec 1, 2025 — classified as 1 death, 50 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Osseointegration Problem” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem28
  • Osseointegration Problem6
  • Biocompatibility5
  • Appropriate Device Problem Term/Code Not Available4
  • Insufficient Device Problem Information4
  • Loss of Osseointegration4
  • Degraded2
  • Tear, Rip or Hole in Device Packaging2

Patient problems

  • No Code Available21
  • Pain20
  • Test Result12
  • Discomfort10
  • Inadequate Osseointegration10
  • Unspecified Infection9
  • Tissue Damage6
  • Foreign Body Reaction5

Recent adverse event reports

The 25 most recent of 54 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 1, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO INFECTION. DOI: (B)(6) 2023, DOR: (B)(6) 2025, AFFECTED SIDE: UNKNOWN HIP.

    Patient problems: Unspecified Infection

    MAUDE report 23685766 ↗
  • MalfunctionFeb 26, 2025Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THERE WAS DAMAGE TO PLASTIC PACKAGING ON PRODUCT. OUTER PACKAGING WITHOUT DAMAGE. NO IMPACT ON PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21472179 ↗
  • InjurySep 30, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ON (B)(6) 2012, PATIENT UNDERWENT A RIGHT TOTAL HIP ARTHROPLASTY DUE TO DEGENERATIVE JOINT DISEASE. ON (B)(6) 2023, PATIENT UNDERWENT A REVISION DUE TO PAIN AND LOSS OF FUNCTION SECONDARY TO METALLOSIS.

    Patient problems: Foreign Body Reaction; Loss of Range of Motion; Pain

    MAUDE report 20338977 ↗
  • InjuryJun 7, 2024Osseointegration Problem; Osseointegration Problem

    IT WAS REPORTED THAT THE EXTRACTION OF TOTAL HIP IMPLANTS WAS PERFORMED DUE TO LOOSENING. DOI: (B)(6) 2021 DOR: (B)(6) 2024 AFFECTED SIDE: RIGHT HIP.

    Patient problems: Inadequate Osseointegration

    MAUDE report 19492401 ↗
  • InjuryOct 2, 2023Osseointegration Problem; Osseointegration Problem

    REVISION OF CORAIL STEM DUE TO LOOSENING. THE SURGEON BELIEVES THAT THE STEM WAS INITIALLY OVERSIZED, POTTED DISTALLY WITHOUT FULLY SEATING PROXIMALLY, AND EVENTUALLY CAME LOOSE. A STEM ONLY REVISION SURGERY WAS PERFORMED. STEM AND HEAD WERE REMOVED. DOI: (B)(6) 2019. DOE: (B)(6) 2023. AFFECTED SIDE: RIGHT.

    Patient problems: Inadequate Osseointegration; Pain

    MAUDE report 17851448 ↗
  • InjuryFeb 13, 2023Osseointegration Problem; Osseointegration Problem

    LOOSE STEM AND ACETABULAR IMPLANTS DUE TO SUSPECTED INFECTION, ALL IMPLANTS REMOVED AND TEMPORARY CEMENT SPACER IMPLANTED, THIS WILL STAY IN PLACE UNTIL INFECTION HAS BEEN TREATED, SECOND STAGE REVISION WILL THEN TAKE PLACE. PRIMARY OPERATION WAS DONE ON (B)(6) 2021. DOR: (B)(6) 2023.

    Patient problems: Inadequate Osseointegration; Unspecified Infection

    MAUDE report 16366952 ↗
  • InjuryDec 14, 2022Osseointegration Problem; Osseointegration Problem

    CORAIL LOSENING. UNNDERSIZING ? FIRST IMPLANTATION 2019.

    Patient problems: Inadequate Osseointegration

    MAUDE report 15978165 ↗
  • InjuryAug 29, 2022Osseointegration Problem; Osseointegration Problem

    PATIENT WAS REVISED DUE TO LOOSE FEMORAL STEM. THE LOOSENING IS NON CEMENTED. THERE WAS NO SURGICAL DELAY. DOI: (B)(6) 2018 DOR: (B)(6) 2022. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Inadequate Osseointegration

    MAUDE report 15313221 ↗
  • InjuryAug 11, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SURGEON HAD REMOVED IMPLANTS FOR INFECTION AT BRISBANE PRIVATE HOSPITAL. DR [SURGEON NAME] DOES NOT REQUIRE ANY CONTACT AS REMOVAL WAS FOR INFECTION ONLY. ORIGINAL SURGERY WAS DONE IN 2020. ORIGINAL SURGERY SURGEON WAS DR [SURGEON NAME]. ORIGINAL SURGERY HOSPITAL WAS AT (B)(6) HOSPITAL.

    Patient problems: Unspecified Infection

    MAUDE report 15214697 ↗
  • InjuryMar 8, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED DUE TO INFECTION. DOI: (B)(6) 2021 (CUP, SCREWS, STEM), DOI: (B)(6) 2021(HEAD, LINER), DOR: FEB 21, 2022, AFFECTED SIDE: LEFT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 13693393 ↗
  • InjuryOct 1, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED DUE TO INFECTION. DOI: UNKNOWN, DOR: (B)(6) 2021, AFFECTED SIDE: UNKNOWN HIP.

    Patient problems: Unspecified Infection

    MAUDE report 12563141 ↗
  • InjurySep 28, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED DUE TO 2ND STAGE IMPLANTATION BECAUSE OF INFECTION. IT WAS ALSO INDICATED THAT THERE WAS LOOSENING OF THE UNKNOWN COMPONENT AT THE BONE TO IMPLANT INTERFACE. DOI: (B)(6) 2019, DOR: (B)(6) 2021, (LEFT HIP).

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 12540307 ↗
  • InjuryJul 22, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED DUE TO PAIN. DOI: UNKNOWN - DOR: (B)(6) 2021 (RIGHT HIP).

    Patient problems: Pain

    MAUDE report 12211221 ↗
  • InjuryJun 15, 2021Degraded; Adverse Event Without Identified Device or Use Problem; Degraded; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT COMPLAINED OF HIP PAIN. XRAYS WERE OBTAINED AND SHOWN TO BE A ASR CUP THAT HAS BEEN RECALLED. A BLOOD TEST WAS COMPLETED AND SHOWN TO HAVE HIGH METAL IONS IN HIS BLOOD WORK. A MRI AND CT WERE ORDERED TO SHOW HE HAD A PSEUDO TUMOR. INTRA-OP WOULD INDICATE THAT WAS TRUE. THERE WAS METALOSIS, AND BLACKENING OF TISSUE IN HIS HIP. CULTURES AND SPECIMEN WERE SENT OFF FOR FURTHER TESTING. THE CUP WAS REMOVED, THE HIP WAS DEBRIDED, IRRIGATED AND A NEW CUP WAS PLACED. AFTER TRIALING USING A 40 MM HEAD AND POLY IT WAS DETERMINED THIS PATIENT HAD A STABLE HIP WITH FULL RANGE OF MOTION. CATALOG AND LOT NUMBERS ARE NOT AVAILABLE DUE TO AN UPGRADE IN THE HOSPITALS SOFTWARE. NO OP REPORT WAS AVAILABLE. THE ORIGINAL SURGEON THAT DID THE PRIMARY HAS PAST AWAY SEVERAL YEARS AGO. NO OTHER INFORMATION IS AVAILABLE FROM WHAT HAS BEEN STATED. DOI: (B)(6) 2009 - DOR: (B)(6) 2021 (LEFT HIP).

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Deformity/ Disfigurement; Discomfort; Distress; Foreign Body Reaction; Hypersensitivity/Allergic reaction; Metal Related Pathology; Pain; Swelling/ Edema; Unspecified Tissue Injury

    MAUDE report 12000512 ↗
  • InjuryApr 5, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT RECEIVED A REVISION DUE TO METALLOSIS AND METAL POISONING. THE HIP BONE AND SURROUNDING MUSCLE HAD SEVERE DEGENERATION DUE TO THE ELEVATED IONS. THE SUP, HEAD AND LINER WERE REVISED. THE STEM WAS RETAINED. DOI: (B)(6) 2009, DOR: (B)(6) 2020 (HEAD), DOR: (B)(6) 2020 (CUP, HEAD, STEM), LEFT HIP.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Foreign Body Reaction; Unspecified Tissue Injury

    MAUDE report 11612921 ↗
  • InjuryMar 26, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED DUE TO INFECTION. DOI: (B)(6) 2021; DOR: (B)(6) 2021; AFFECTED SIDE: LEFT HIP.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 11574839 ↗
  • InjuryJan 19, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    CLINICAL ADVERSE EVENT RECEIVED FOR TROCHANTERIC BURSITIS. EVENT IS NOT SERIOUS AND IS CONSIDERED MILD. EVENT IS DEFINITELY NOT RELATED DEVICE AND POSSIBLY RELATED TO PROCEDURE. DATE OF IMPLANT: (B)(6) 2019, DATE OF EVENT: (B)(6) 2021, (LEFT HIP).

    Patient problems: Inflammation; Pain

    MAUDE report 11195898 ↗
  • MalfunctionSep 10, 2020Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    THE IMPLANT WAS TOO BIG FOR ITS BROACH. COULD NOT FIT INSIDE THE FEMUR. HAD TO USE ANOTHER IMPLANT OF SAME SIZE.

    Patient problems: No Code Available; Not Applicable

    MAUDE report 10515797 ↗
  • MalfunctionJul 16, 2020Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IMPLANTS WERE RETURNED DUE TO WET PACKAGING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10283960 ↗
  • InjuryMay 26, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION CASE OF AN INFECTED CORAIL PINNACLE. IN SPEAKING WITH THE SURGEON BEFORE THE REVISION PROCEDURE, HE ADVISED THAT THE PATIENT HAD PRIMARY THR IN FEBRUARY THIS YEAR AND HAS PRESENTED TO ED WITH A DISSOCIATED ACETABULAR LINER AND INFECTED HIP. SURGEON UNABLE TO SAY WHY HE THOUGHT THE PATIENT HAD A DISSOCIATED LINER BUT DID MENTION THE PATIENT REPORTING A FALL AT SOME STAGE. REVISED HIP TO ZB-G7 DUAL MOBILITY CUP AND C-STEM (ATTACHMENT INCLUDED WITH PRODUCT USAGE FROM REVISION CASE COMPLETED TODAY). REVISION SURGERY WENT AS PLANNED. EXPLANTED PRODUCTS NOT AVAILABLE AS RETAINED BY HOSPITAL. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: Fall; Foreign Body Reaction; No Code Available; Unspecified Infection

    MAUDE report 10088019 ↗
  • InjuryMay 14, 2020Degraded; Adverse Event Without Identified Device or Use Problem; Degraded; Adverse Event Without Identified Device or Use Problem

    PCF AND MEDICAL RECORDS RECEIVED. AFTER REVIEW OF MEDICAL RECORDS, NO REVISION NOTES REPORTED. PATIENT ALLEGES CHRONIC PAIN, STIFFNESS, WALKING DIFFICULTY, SWELLING, DIFFICULTY IN WEARING SHOES AND SOCKS, DIFFICULTY WITH STAIRS, PAINFUL ANTALGIC GAIT, LIMP, ARTHRALGIAS, JOINT SWELLING, AND MYALGIAS. PHYSICAL EXAM STATED DECREASED STRENGTH, TENDERNESS, BONY TENDERNESS, SWELLING, AND CREPITUS. LABORATORY RESULTS STATED ELEVATED METAL IONS. DOI: (B)(6) 2008. DOR: NONE REPORTED; (LEFT HIP). PATIENT IS BILATERAL. SEE (B)(4) FOR THE RIGHT HIP.

    Patient problems: Ambulation or Postural Difficulties; Discomfort; Edema; Limited Mobility Of The Implanted Joint; Muscular Rigidity; No Code Available; Pain; Test Result

    MAUDE report 10055968 ↗
  • InjuryMay 1, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENTS' HIP WAS INFECTED. THE HIP HEMI CONSTRUCT WAS REMOVED AND AN ANTIBIOTIC SPACER WAS PLACED. DOI: (B)(6) 2020; RIGHT HIP.

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 10019022 ↗
  • InjuryMar 30, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    LITIGATION ALLEGES THAT PATIENT EXPERIENCED PAIN, WEAKNESS, DIFFICULTY WITH SIMPLE DAILY LIVING ACTIVITIES, INABILITY TO RETURN TO WORK, AND INCREASED METALLIC IONS IN HIS BLOODSTREAM. REVISION OPERATIVE NOTE REVIEWED AND CONFIRMED METALLOSIS. THE REVISION OPERATIVE NOTE ALSO MENTIONS ALVAL. IN ADDITION TO WHAT PREVIOUSLY ALLEGED, LITIGATION ALLEGED METAL WEAR, SKIN RASHES, WEIGHT LOSS, PAIN, INCREASE INCIDENCE OF ATRIAL FIBRILLATION, ANXIETY, THIRST, LOOSE OF TASTE, METALLIC TASTE IN SALIVA, AND DRY FRAGILE SKIN. AFTER REVIEW OF MEDICAL RECORDS IT WAS INDICATED THAT THE PATIENT PRESENT WITH PROGRESSIVE WORSENING LEFT HIP PAIN, DEMONSTRATED ELEVATED METAL IONS. THE PATIENT WAS THEN REVISED FOR PAINFUL METAL-SENSITIVITY REACTION TO NECROSIS (ALVAL). OPERATIVE NOTES REPORTED THAT THE CUP WAS LITTLE GENEROUSLY ANTEVERTED AND THERE WAS SOME EROSIVE DEFICIENCY NOTED OF THE ANTERIOR ACETABULAR BONY WALL, DUE TO INFLAMMATORY PROCESS.

    Patient problems: Anxiety; Atrial Fibrillation; Discomfort; Foreign Body Reaction; No Code Available; Polydipsia; Rash; Weakness

    MAUDE report 9899654 ↗
  • InjuryMar 10, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED TO ADDRESS PAIN. SWAPPED LINER LEFT CUP. DOI: (B)(6) 2017; DOR: (B)(6) 2020; LEFT HIP.

    Patient problems: No Code Available; Pain

    MAUDE report 9811835 ↗
  • InjuryFeb 25, 2020Use of Device Problem; Malposition of Device; Use of Device Problem; Malposition of Device

    CLINICAL ADVERSE EVENT RECEIVED FOR STEM ROTATION CAUSING PAIN : MALPOSITION/ NON-UNION TROCHANTER. EVENT IS SERIOUS AND IS CONSIDERED SEVERE. EVENT IS DEFINITELY RELATED TO DEVICE AND THERE IS A REMOTE POSSIBILITY THAT THE EVENT IS RELATED TO PROCEDURE. DATE OF IMPLANTATION: (B)(6) 2019. DATE OF EVENT (ONSET): (B)(6) 2020, (RIGHT HIP).

    Patient problems: Bone Fracture(s); Edema; No Code Available; Pain

    MAUDE report 9751743 ↗

Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices

All Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices →