Skip to content

S-ROM ZTT 52-1403Recalls & reported problems

Depuy (Ireland) · S-ROM Modular Hip System ZTT Proximal Sleeve POROUS COATED 20 B SML FITS: 20 x 15 STEM Cone: B Triangle: Sml · Class 3 · Implant

All models of S-ROM ZTT (36)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10603295170570), received between Oct 9, 2018 and Dec 2, 2024 — classified as 6 injuries.

The most frequently reported specific device problem is “Osseointegration Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem4
  • Osseointegration Problem2
  • Insufficient Device Problem Information1
  • Malposition of Device1
  • Migration1
  • Use of Device Problem1

Patient problems

  • Pain5
  • Inadequate Osseointegration2
  • No Code Available2
  • Unspecified Infection2
  • Discomfort1
  • Foreign Body Reaction1
  • Hypersensitivity/Allergic reaction1
  • Osteolysis1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 2, 2024Osseointegration Problem; Osseointegration Problem

    IT WAS REPORTED THAT THE SURGEON REVISED FEMORAL COMPONENTS AND REPLACED WITH A COMPETITOR REVISION STEM. PATIENT HAD TRAUMA TO THE HIP AND THE STEM BECAME LOOSE. PROCEDURE WAS COMPLETED WITH NO SURGICAL DELAY. DOI: (B)(6) 2023. DOR: (B)(6) 2024. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Inadequate Osseointegration

    MAUDE report 20825214 ↗
  • InjuryMay 5, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ASR LITIGATION COMPLAINT RECEIVED AD 31 JAN 2022. LITIGATION RECORDS ALLEGES PAIN, INJURY ECONOMIC LOSS, PUNITIVE DAMAGES, EXCESSIVE LEVELS OF CHROMIUM AND COBALT. DOI: (B)(6) 2008. DOR: (B)(6) 2020, LEFT HIP.

    Patient problems: Hypersensitivity/Allergic reaction; Osteolysis; Pain; Pulmonary Embolism

    MAUDE report 14286631 ↗
  • InjuryJun 17, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PRIMARY SURGERY WAS PERFORMED ON (B)(6) 2007. IT WAS REPORTED THAT THE REVISION SURGERY WAS SCHEDULED ON (B)(6) 2020 DUE TO PAIN. THERE WAS SUSPICION OF ARMD. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Code Available; Pain

    MAUDE report 10162481 ↗
  • InjuryAug 9, 2019Use of Device Problem; Malposition of Device; Adverse Event Without Identified Device or Use Problem; Migration; Use of Device Problem; Malposition of Device; Adverse Event Without Identified Device or Use Problem; Migration

    LITIGATION ALLEGES PAIN, INSTABILITY AND ELEVATED METAL ION LEVELS. MEDICAL RECORDS REPORTED PAIN WITH WEIGHT BEARING AN A RADIOGRAPH REPORTEDLY SHOWED MILD VARUS SHIFT OF THE FEMORAL COMPONENT SUGGESTING LACK OF INGROWTH AND SEVERE OSTEOPOROSIS. UPDATE AD 25 JUL 2018:(B)(4) HAS BEEN RE-OPENED UNDER (B)(4) DUE TO THE RECEIPT OF PPF AND IMPLANT RECORD. PPF ALLEGES INFECTION, METALLOSIS AND METAL WEAR. UPDATED DOR, AGE, LAWYER AND LAW FIRM. ADDED IPS FOR THE PATIENT HARM. DOI: (B)(6) 2007 - DOR: (B)(6) 2018, (RIGHT HIP).

    Patient problems: Foreign Body Reaction; No Code Available; Pain; Unspecified Infection

    MAUDE report 8879281 ↗
  • InjuryApr 22, 2019Osseointegration Problem; Osseointegration Problem

    PPF ALLEGES LOOSENING OF STEM AND FRACTURE (BONE). DOI: (B)(6) 2009 - DOR: (B)(6) 2014 (LEFT HIP)

    Patient problems: Discomfort; Inadequate Osseointegration; Pain

    MAUDE report 8535258 ↗
  • InjuryOct 9, 2018Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    LITIGATION ALLEGES LOOSENING OF CUP AND INFECTION. DOI: (B)(6) 2008 - DOR: NONE REPORTED (LEFT HIP).

    Patient problems: Pain; Unspecified Infection; Weakness

    MAUDE report 7946565 ↗

Other Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented devices

All Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented devices →

Sources and updates