S-ROM 52-3422Recalls & reported problems

Depuy (Ireland) · S-ROM Total Hip System FEMORAL STEM Standard 36 STD NECK +8 11/13 TAPER 22x17x165 Lateral · Class 3 · Implant

All models of S-ROM (185)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
15
Deaths reported
0
Injuries reported
15
Recalls naming it
0

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 10603295171058), received between Jun 29, 2017 and May 17, 2024 — classified as 15 injuries.

The most frequently reported specific device problem is “Loss of Osseointegration” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem8
  • Loss of Osseointegration3
  • Biocompatibility1
  • Degraded1
  • Insufficient Device Problem Information1
  • Loss of or Failure to Bond1
  • Naturally Worn1
  • Osseointegration Problem1

Patient problems

  • Pain11
  • No Code Available7
  • Test Result7
  • Discomfort4
  • Inadequate Osseointegration4
  • Unspecified Infection4
  • Foreign Body Reaction3
  • Loss of Range of Motion3

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 17, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT HAS A RIGHT TOTAL HIP REPLACEMENT DUE TO SEVERE, DEGENERATIVE ARTHRITIS. PATIENT WAS REVISED TO ADDRESS PAIN AND INFECTED HIP. ASR CONSTRUCT REVISED. IT WAS ALSO INDICATED THAT THERE WAS LOOSENING AT THE BONE TO IMPLANT INTERFACE. DOR: (B)(6) 2018; RIGHT HIP.

    Patient problems: Adhesion(s); Foreign Body Reaction; Pain; Unspecified Infection

    MAUDE report 19339765 ↗
  • InjuryDec 15, 2021Degraded; Degraded

    CLINICAL NOTIFICATION RECEIVED FOR REVISION OF RIGHT TOTAL HIP TO ADDRESS HIP DISLOCATION AND POLYETHYLENE LINER WEAR. DATE OF IMPLANTATION: (B)(6) 2002. DATE OF REVISION: (B)(6) 2021. (RIGHT HIP). TREATMENT: REVISION OF ACETABULAR LINER AND FEMORAL HEAD.

    Patient problems: Joint Laxity; Swelling/ Edema; Synovitis

    MAUDE report 13013862 ↗
  • InjurySep 1, 2021Osseointegration Problem; Osseointegration Problem

    ADDITIONAL INFORMATION RECEIVED INDICATED THAT THERE WAS LYSIS AROUND THE PROXIMAL SECTION OF THE FEMORAL COMPONENT. IT WAS DIFFICULT TO REMOVE AND IT FELT LIKE THE FEMORAL COMPONENT WAS "WIDESCREEN WIPERING" IN THE FEMUR.

    Patient problems: Inadequate Osseointegration; Osteolysis

    MAUDE report 12407418 ↗
  • InjuryMay 12, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IN ADDITION TO WHAT WERE PREVIOUSLY ALLEGED, LITIGATION REPORTED THAT THE PLAINTIFFS REVISION INVOLVED, IN PART, A FINDING OF AN ADVERSE TISSUE REACTION RELATED TO THE METAL-ON-METAL ARTICULATION, RELEASE OF A LARGE AMOUNT OF BROWNISH FLUID, AND DEBRIDEMENT AND REMOVAL OF BROWNISH FRIABLE TISSUE THROUGHOUT THE ENTIRE HIP JOINT, ALL DUE TO TOXIC HEAVY METAL/METAL ION COBALT AND CHROMIUM POISONING.

    Patient problems: Ambulation or Postural Difficulties; Anxiety; Arrhythmia; Depression; Discomfort; Distress; Dyspnea; Extraskeletal ossification; Fall; Fatigue; Foreign Body Reaction; Hypersensitivity/Allergic reaction; Limb Fracture; Loss of Range of Motion; Muscle/Tendon Damage; Numbness; Pain; Paresthesia; Swelling/ Edema; Unspecified Tissue Injury

    MAUDE report 11814330 ↗
  • InjurySep 8, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED DUE TO PAIN SECONDARY TO POSSIBLE INFECTION. DOI: UNKNOWN, DOR: (B)(6) 2020, AFFECTED SIDE: LEFT HIP.

    Patient problems: No Code Available; Pain; Unspecified Infection

    MAUDE report 10504857 ↗
  • InjuryMar 19, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AFTER REVIEW OF MEDICAL RECORDS PATIENT WAS REVISED TO ADDRESSED FAILED RIGHT METAL ON METAL TOTAL HIP ARTHROPLASTY WITH METALLOSIS. OPERATIVE NOTES INDICATED PAIN, ELEVATED COBALT AND CHROMIUM LEVELS. THERE WAS ABUNDANT SCAR TISSUE AROUND THE JOINT THAT WAS DEBRIDED. EVIDENCE OF METALLOSIS WITH SOME BROWN STAINING OF TISSUE IN THE SYNOVIUM AND FLUID IN THE JOINT. THERE WERE NO SIGNS OF INFECTION. DOI: (B)(6) 2006, DOR: (B)(6) 2015, (RIGHT HIP).

    Patient problems: Adhesion(s); Edema; Foreign Body Reaction; No Code Available; Pain; Test Result

    MAUDE report 9856861 ↗
  • InjurySep 30, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED BECAUSE OF PAIN. UPDATE REC'D 9/19/2014 - LITIGATION ALLEGES THE PATIENT SUFFERS FROM PAIN. UPDATE 21 AUG 2018 (B)(4) HAS BEEN RE-OPENED DUE TO RECEIPT OF PPF AND MEDICAL RECORDS. IN ADDITION TO WHAT PREVIOUSLY ALLEGED, PPF ALLEGED ELEVATED METAL IONS. AFTER REVIEW OF MEDICAL RECORDS, NO REVISION NOTES REPORTED. DOI: (B)(6) 2003. DOR: (B)(6) 2013; LEFT HIP.

    Patient problems: No Code Available; Pain; Test Result

    MAUDE report 9134365 ↗
  • InjuryJun 20, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REASON FOR ORIGINAL COMPLAINT.- LITIGATION PAPERS ALLEGE THE PATIENT WAS REVISED ON IN (B)(6) OF 2008 AND AGAIN ON (B)(6) 2010. ORIGINAL LITIGATION INDICATED THAT THE PATIENT WAS IMPLANTED WITH ASR; HOWEVER, AMENDED LITIGATION ALLEGES THAT THE PATIENT WAS IMPLANTED WITH PINNACLE. LITIGATION ALLEGES PAIN AND MOBILITY ISSUES. THE REVISION OPERATIVE NOTE INDICATED PAIN AND NO METALLOSIS. THE STEM WAS REVISED SO THE ACETABULAR COMPONENT WAS ABLE TO BE VISUALIZED. IN ADDITION TO WHAT WAS PREVIOUSLY ALLEGED, PPF ALLEGES ELEVATED METAL IONS AND METAL WEAR. STEM HAS BEEN ADDED DUE TO THE ALLEGED BLOOD HEAVY METAL INCREASED. DOI: (B)(6) 2002; DOR: (B)(6) 2011; LEFT HIP. PATIENT HAS BILATERAL HIP IMPLANTS.

    Patient problems: Loss of Range of Motion; No Code Available; Pain; Test Result

    MAUDE report 8717744 ↗
  • InjuryFeb 7, 2019Naturally Worn; Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED TO ADDRESS PAIN, LEG LENGTH DISCREPANCY, SYNOVITIS. UPDATE PPF ALLEGES PSEUDOTUMOR, METAL WEAR, METALLOSIS AND ELEVATED METAL IONS. DOI: (B)(6) 2009; DOR: (B)(6) 2012; (LEFT HIP).

    Patient problems: Limited Mobility Of The Implanted Joint; No Code Available; Pain; Test Result; Tissue Damage

    MAUDE report 8315891 ↗
  • InjuryJan 22, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    LITIGATION ALLEGES ELEVATED METAL IONS. DOI: (B)(6) 2003 - DOR: (B)(6) 2011 (RIGHT HIP).

    Patient problems: Discomfort; Inflammation; No Code Available; Pain; Test Result

    MAUDE report 8267365 ↗
  • InjuryDec 17, 2018Loss of Osseointegration; Biocompatibility; Loss of Osseointegration; Biocompatibility

    LITIGATION ALLEGES THAT THE PATIENT SUFFERS FROM LARGE AMOUNTS OF TOXIC AMOUNTS OF COBALT CHROMIUM METAL IONS AND PARTICLES TO BE RELEASED INTO THE PATIENTS BLOOD, TISSUE, AND BONE SURROUNDING THE IMPLANT. IT IS ALSO ALLEGED THAT THE PATIENT SUFFERS FROM PAIN, DISCOMFORT, AND WEAKNESS. UPDATE AD 04 MAY 2018: RECEIPT OF PPF AND MEDICAL RECORD. PPF ALLEGES LOOSENING OF STEM. AFTER REVIEW OF MEDICAL RECORDS, PATIENT WAS REVISED TO ADDRESS FAILED TOTAL HIP ARTHROPLASTY. ADDED STEM DUE TO ALLEGED LOOSENING OF STEM.

    Patient problems: Discomfort; Inadequate Osseointegration; No Code Available; Pain; Test Result; Weakness

    MAUDE report 8168750 ↗
  • InjurySep 19, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    PPF ALLEGES STROKE/HEART ATTACK AFTER REVISION. NO REVISION NOTE PROVIDED. DOI: (B)(6) 2011; DOR: NONE REPORTED; (RIGHT HIP).

    Patient problems: Infarction, Cerebral; Myocardial Infarction

    MAUDE report 7890193 ↗
  • InjuryJul 17, 2018Loss of Osseointegration; Loss of Osseointegration

    COMPLAINT DESCRIPTION: LITIGATION ALLEGES THE PATIENT SUFFERS FROM PAIN, DISCOMFORT, IMMOBILITY, INFLAMMATION AND HIGH LEVELS OF CHROMIUM AND COBALT. UPDATE AD (B)(6) 2018: (B)(4) HAS BEEN RE-OPENED UNDER (B)(4) DUE TO THE RECEIPT OF PPF AND STICKER SHEETS RECORD. PPF ALLEGES INFECTION REQUIRING IV ANTIBIOTICS AND LOOSENING OF STEM. UPDATED PRODUCT AND LOT NUMBER OF INSERT (121887356/2136865), HEAD ( 136531000/2193018), STEM ( 523422/2160316), CUP (121722056/A5WFX1000).

    Patient problems: Discomfort; Inadequate Osseointegration; Inflammation; Loss of Range of Motion; Pain; Test Result; Unspecified Infection

    MAUDE report 7694694 ↗
  • InjuryJul 16, 2018Loss of or Failure to Bond; Loss of or Failure to Bond

    PINNACLE PPF AND STICKER SHEETS RECEIVED. PPF ALLEGES INFECTION REQUIRING IV ANTIBIOTICS AND LOOSENING OF STEM. IT WAS ALSO INDICATED IN STICKERS SHEETS THAT PINNACLE SECTOR II ACETABULAR CUP, PINNACLE CANCELLOUS BONE SCREW, S-ROM TOTAL HIP SYSTEM FEMORAL STEM STANDARD, AND S-ROM TOTAL HIP SYSTEM ZTT PROXIMAL SLEEVE WERE IMPLANTED ON (B)(6) 2006. DOI: (B)(6) 2006 - DOR: (B)(6) 2010 (RIGHT HIP).

    Patient problems: Inadequate Osseointegration; Not Applicable; Unspecified Infection

    MAUDE report 7689481 ↗
  • InjuryJun 29, 2017Loss of Osseointegration; Loss of Osseointegration

    JUN 1, 2017. LEGAL MEDICAL RECORDS RECEIVED. LITIGATION ALLEGES PAIN IN PATIENT'S HIPS AND THIGHS. AFTER REVIEW OF MEDICAL RECORDS FOR MDR REPORTABILITY IT WAS REPORTED THAT THE PATIENT WAS REVISED TO ADDRESS MECHANICAL FAILURE OF RIGHT THR SECONDARY TO RECURRENT FEMORAL LOOSENING AND HETEROTOPIC OSSIFICATION. THERE WAS NO SIGNIFICANT SYNOVITIS OR CLINICALLY OBVIOUS SIGN OF INFECTION. THERE IS HETEROTOPIC OSSIFICATION AROUND THE FEMORAL ACETABULAR COMPONENTS, WHICH WAS EXTENSIVELY DEBRIDED. THERE WAS NO OVERT SIGN OF METALLOSIS, NO PSEUDOTUMOR. THE LEG LENGTH WAS FOUND TO BE SIGNIFICANTLY SHORT BY APPROX. 1 INCH WHICH WAS EQUALIZED. THE FEMORAL STEM WAS FOUND TO BE GROSSLY LOOSE AND WAS REMOVED. LABORATORY RESULT FOR COBALT CHROMIUM ARE BELOW 7 UG/L.

    Patient problems: Pain

    MAUDE report 6677902 ↗

Other Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented devices

All Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented devices →