S-ROM 53-6318Recalls & reported problems
Depuy (Ireland) · S-ROM HIP STEM TRIAL BODY 18mm · Class 1
All models of S-ROM (185)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10603295172802), received between Jun 1, 2018 and Oct 28, 2024 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device-Device Incompatibility2
- Bent1
- Break1
- Crack1
- Material Deformation1
- Naturally Worn1
- Scratched Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Insufficient Information1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 28, 2024Break; Crack; Material Deformation; Naturally Worn; Break; Crack; Material Deformation; Naturally Worn
IT WAS REPORTED THAT THE TRIAL BROKE DURING HIP SURGERY. IT DID NOT BREAK INTO PIECES, IT DID NOT DELAY SURGERY, NOTHING WAS LEFT INSIDE THE PATIENT.
Patient problems: Insufficient Information
MAUDE report 20544752 ↗MalfunctionFeb 21, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT AN UNKNOWN SURGERY WAS PLANNED ON (B)(6) 2023. THE IMPRINT ON THE STEM BODY TRIAL WAS MISALIGNED. THIS EVENT WAS NOTICED ON (B)(6) 2023, BEFORE THE DEVICE WAS DELIVERED TO THE HOSPITAL. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16408481 ↗MalfunctionJan 5, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON (B)(6), 2022, THE S-ROM TRIAL STEM BODY18 IN QUESTION WAS OUT OF MIDLINE WHEN MATED WITH NECK TRIAL. THE DEVICE WAS ORDERED FOR THE SURGERY DATED (B)(6), 2022, IT WAS BEFORE USE. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16102492 ↗MalfunctionJun 1, 2018Bent; Scratched Material; Bent; Scratched Material
NEW REP AT (B)(6) MEDICAL CENTER. WENT THROUGH INSTRUMENT INVENTORY ON (B)(6) 2018, AND NOTICED A FEW INSTRUMENTS THAT WERE IN AN UNACCEPTABLE CONDITION. THIS INSTRUMENT WAS NOT USED DURING A SURGERY AND NO STAFF MEMBER HAD MENTIONED THE DAMAGE. THIS IS A SROM 18 PROXIMAL BODY (COLOR: GREEN). THE GROOVES WHERE THE NECK TRIAL SITS ARE DAMAGED (BENT, CHIPPED). ASKING THAT THESE ARE REPLACED AND SHIPPED TO: (B)(6). THE DAMAGED INSTRUMENT WILL BE RETURNED TO (B)(4). LOT CODE: AH0604. ON THE COMPLAINT TOOL; YOU ARE UNABLE TO ADD THE LOT CODE. THE PRODUCT WON'T ADD/UPLOAD WHEN THE LOT IS ALSO TYPED. PATIENT CONSEQUENCE? NO.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7560361 ↗
Other TEMPLATE devices
- ATTUNE 2544-01-014Depuy (Ireland) · TEMPLATE802 reports
- ATTUNE 2545-00-546Depuy (Ireland) · TEMPLATE1 recall301 reports
- ARTICUL/EZE 2531-51-000Depuy (Ireland) · TEMPLATE323 reports
- DELTA XTEND 2307-95-000Depuy (Ireland) · TEMPLATE315 reports
- ATTUNE 2545-00-547Depuy (Ireland) · TEMPLATE1 recall231 reports
- ATTUNE 2544-00-520Depuy (Ireland) · TEMPLATE279 reports
- ATTUNE 2545-00-545Depuy (Ireland) · TEMPLATE1 recall221 reports
- ARTICUL/EZE 2530-91-000Depuy (Ireland) · TEMPLATE271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page