Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10603295175209), received between Nov 9, 2017 and May 2, 2023 — classified as 7 injuries.
The most frequently reported specific device problem is “Fracture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Fracture2
- Appropriate Device Problem Term/Code Not Available1
- Insufficient Device Problem Information1
- Loss of Osseointegration1
- Malposition of Device1
- Unstable1
- Use of Device Problem1
Patient problems
- No Code Available2
- Not Applicable2
- Pain2
- Ambulation or Postural Difficulties1
- Fibrosis1
- Foreign Body Reaction1
- Hypersensitivity/Allergic reaction1
- Insufficient Information1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryMay 2, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE PRIMARY SURGERY WAS PERFORMED WITH THE IMPLANTS IN QUESTION. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. THE REVISION SURGERY IS SCHEDULED ON (B)(6) 2023, DUE TO THE SUSPICION OF TRANIONOSIS. THE DETAILED INFORMATION WILL BE UPDATED AT A LATER DATE. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Foreign Body Reaction; Hypersensitivity/Allergic reaction; Swelling/ Edema
MAUDE report 16853792 ↗InjuryJan 11, 2023Fracture; Fracture
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE PATIENT UNDERWENT THE SURGERY VIA THA FOR RIGHT OA WITH S-ROM-A STEM IN QUESTION. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. AFTER THE SURGERY, IT WAS CONFIRMED ON (B)(6) 2022, THAT THE STEM BROKE ON BODY PART AND SLIT ROOT PART. IT CAN SEE THE BROKEN PARTS IN THE ATTACHED PHOTO. A REVISION SURGERY IS PLANNED ON (B)(6) 2023. THE SURGEON WANTS TO KNOW THE OCCURRENCE SITUATION OF THE EVENT WHERE THE STEM IS BROKEN IN TWO PLACES (STEM BODY PART AND SLIT BASE PART) LIKE THIS TIME. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Ambulation or Postural Difficulties; Insufficient Information
MAUDE report 16137617 ↗InjuryDec 20, 2019Fracture; Use of Device Problem; Malposition of Device; Fracture; Use of Device Problem; Malposition of Device
THE PRIMARY SURGERY WAS PERFORMED ON (B)(6) 2016 VIA THA. IT WAS REPORTED THAT IT WAS CONFIRMED THE STEM (P/N: 900525210) AND THE SLEEVE (P/N: 550504) FRACTURE BY X-RAY WHEN THE PATIENT VISITED TO THE HOSPITAL ON (B)(6) 2019, AND THE PATIENT¿S CONDITION WAS UNWALKABLE. THUS, THE REVISION SURGERY WAS SCHEDULED TO BE PERFORMED IN THE NEAR FUTURE (DATE OF SURGERY WAS NOT FIXED YET). POSSIBLE CAUSE: THE IMPLANTS IN QUESTION HAD BEEN PLACED IN A LITTLE BIT VARUS AT THE PRIMARY SURGERY. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Code Available; Not Applicable
MAUDE report 9505566 ↗InjuryDec 7, 2018Unstable; Appropriate Device Problem Term/Code Not Available; Unstable; Appropriate Device Problem Term/Code Not Available
InjuryOct 15, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PENDING PRODUCT LIABILITY, PFS ALLEGES PAIN, NEUROPATHY ON THE RIGHT FOOT, MUSCLE SPASM, AND MUSCLE ATROPHY. AFTER REVIEW OF MEDICAL RECORD, PATIENT WAS REVISED TO ADDRESS FAILED PAINFUL RIGHT TOTAL HIP ARTHROPLASTY. REVISION NOTES REPORTED THAT A FAIR AMOUNT OF BLOODY FIBROUS DEBRIS WAS CULTURED AND DEBRIDED. THE STEM WAS FOUND TO BE LOOSE AND BROKEN. INTRAOPERATIVE RADIOGRAPH SHOWED THAT PERFORATION HAD OCCURRED AT THE LATERAL CORTEX, SO AN ETO WAS PERFORMED. A CRACK WAS THEN NOTED AT THE DISTAL PART OF THE FEMUR, AND TWO CABLES WERE PASSED DISTALLY, AND ALSO TWO PROXIMALLY, TO REPAIR THE ETO.
Patient problems: Fibrosis; Muscle Spasm(s); Neuropathy; No Code Available; Pain; Tissue Damage
MAUDE report 7966422 ↗InjuryJun 13, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT THE REMOVAL SURGERY WAS PERFORMED (DATE OF THE PRIMARY SURGERY AND THE REMOVAL SURGERY WERE UNKNOWN) DUE TO THE INFECTION. THE SURGEON IMPLANTED CEMENT SPACER INTO THE PATIENT¿S BONE. NO FURTHER INFORMATION WAS PROVIDED BY THE HOSPITAL.
Patient problems: Not Applicable; Unspecified Infection
MAUDE report 7596069 ↗InjuryNov 9, 2017Loss of Osseointegration; Loss of Osseointegration
THE PATIENT WAS REVISED TO ADDRESS LOOSENING OF FEMORAL PART AT UNKNOWN INTERFACE. CEMENT MANUFACTURER IS UNKNOWN.
Patient problems: No Information
MAUDE report 7012905 ↗
Other Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented devices
- BIOLOX DELTA 1365-36-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented839 reports
- BIOLOX DELTA 1365-36-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented643 reports
- PINNACLE POROCOAT 1217-22-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented335 reports
- PINNACLE POROCOAT 1217-22-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented263 reports
- BIOLOX DELTA 1365-36-330Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented255 reports
- PINNACLE POROCOAT 1217-22-050Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented253 reports
- PINNACLE POROCOAT 1217-22-056Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented220 reports
- SUMMIT POROCOAT 1570-11-120Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented1 death204 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗