S-ROM ZTT 55-0728Recalls & reported problems
Depuy Orthopaedics, Inc. · S-ROM Total Hip System ZTT Oversize Proximal Sleeve POROUS COATED 20 F LRG FITS: 18 x 13 STEM Cone: F Triangle: LRG 18 INTERNAL DIAMETER · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10603295175582), received between Mar 9, 2020 and Mar 9, 2020 — classified as 1 injury.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- Fatigue1
- Foreign Body Reaction1
- Injury1
- Metal Related Pathology1
- No Code Available1
- Osteolysis1
- Pain1
- Test Result1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryMar 9, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AFTER REVIEW OF MEDICAL RECORDS, THE PATIENT WAS REVISED DUE TO ADVERSE TISSUE RESPONSE. LAB WORK SHOWED AN INCREASE IN COBALT CHROMIUM LEVELS AS SHE BEGAN HAVING PAIN. OPERATIVE NOTES INDICATED THAT THE PATIENT HAD A LARGE FLUID-FILLED SOFT TISSUE MASS OVER THE GREATER TROCHANTER AND EXPANSION OF HER TROCHANTERIC BURSA. THE BURSA WAS FILLED WITH A CLOUDY WHITE FLUID CONSISTENT WITH THAT SEEN IN METAL WEAR REACTION. THERE WAS A GREAT DEAL OF LYTIC WEAR DEBRIS SPREAD THROUGHOUT THIS LARGE SAC. THE GREATER TROCHANTER HAD AREAS OF LYSIS, AS DID THE MEDIAL NECK. THERE WAS EVIDENCE OF FRETTING CORROSION ON THE TAPER. THERE WERE LARGE AREAS OF LYSIS PRESENT BEHIND THE CUP. THE LARGE HEAD BALL WAS DISIMPACTED FROM THE TAPER AND FRETTING CORROSION WAS VISIBLE. ADDITIONALLY, IT WAS ALSO REPORTED THAT DURING THE PRIMARY HIP REPLACEMENT SURGERY, A SMALL CRACK OCCURRED ANTERIORLY IN THE NECK DURING IMPACTION OF THE STEM, IT WAS ELECTED TO PASS A 2 MM CERCLAGE CABLE AROUND THE FEMUR TO PREVENT CRACK EXTENSION.
Patient problems: Fatigue; Foreign Body Reaction; Injury; Metal Related Pathology; No Code Available; Osteolysis; Pain; Test Result; Weakness
MAUDE report 9807006 ↗
Other Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented devices
- BIOLOX DELTA 1365-36-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented839 reports
- BIOLOX DELTA 1365-36-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented643 reports
- PINNACLE POROCOAT 1217-22-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented335 reports
- PINNACLE POROCOAT 1217-22-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented263 reports
- BIOLOX DELTA 1365-36-330Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented255 reports
- PINNACLE POROCOAT 1217-22-050Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented253 reports
- PINNACLE POROCOAT 1217-22-056Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented220 reports
- SUMMIT POROCOAT 1570-11-120Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented1 death204 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page