S-ROM 55-6074Recalls & reported problems
Depuy Orthopaedics, Inc. · S-ROM SELF-TAPPING DOME SCREW - LOW PROFILE 6.5mm x 35mm · Class 2 · Implant
All models of S-ROM (92)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10603295175940), received between Jun 8, 2022 and Jul 22, 2025 — classified as 3 injuries.
The most frequently reported specific device problem is “Osseointegration Problem” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Osseointegration Problem3
- Fracture1
- Naturally Worn1
- Off-Label Use1
Patient problems
- Inadequate Osseointegration3
- Foreign Body Reaction2
- Joint Laxity1
- Pain1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJul 22, 2025Osseointegration Problem; Osseointegration Problem
THIS PC IS RELATED TO (B)(4) WHICH REPORTS THE CAUSE OF THE REVISION SURGERY. THIS PC REPORTS THE POOR FIT OF THE STEM FOUND DURING THE REVISION SURGERY. IT WAS REPORTED THAT ON (B)(6) 2008, THE PATIENT UNDERWENT AN INITIAL SURGERY PERFORMED WITH THE IMPLANTS IN QUESTION. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY. AFTER THE SURGERY ON (B)(6) 2025, A FEMORAL HEAD DISLOCATION WAS CONFIRMED. ON (B)(6) 2025, THE REVISION SURGERY WAS PERFORMED TO REPLACE THE LINER AND HEAD. DURING THE SURGERY, AN ACETABULAR CUP, POLO PRODUCT REPORTED ON (B)(4), LOOSENING WAS CONFIRMED AND THE CUP WAS REPLACED WITH A CEMENT CUP FROM OTHER COMPANY. WHEN THE HEAD WAS REPLACED WITH A NEW 9/10 TAPER HEAD, IT WAS FOUND THAT THE NECK TAPER WAS DISTORTED AND THERE WAS A POOR FIT, WHICH MAY HAVE BEEN THE CAUSE OF THE DISLOCATION, SO THE STEM WAS REMOVED AND REPLACED WITH A NEW CEMENTED STEM (C-STEM AMT). THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY. AFTER THE SURGERY, THE SURGEON CONCLUDED THAT THE REASON FOR THE CUP REVISION WAS THAT BLACK TISSUE HAD FORMED AT THE BOUNDARY SURFACE, WHICH HAD CAUSED BONE DISSOLUTION AND LOOSENING. THE REASON FOR THE STEM REVISION WAS THAT, JU
Patient problems: Inadequate Osseointegration
MAUDE report 22564621 ↗InjuryDec 8, 2022Naturally Worn; Osseointegration Problem; Naturally Worn; Osseointegration Problem
PATIENT WAS REVISED DUE TO LOOSENING OF A COMPETITOR CUP AND DEPUY SCREWS AT THE BONE TO IMPLANT INTERFACE, WEAR OF THE SCREWS, ARMD, AND UNSPECIFIED SUBLUXATION OF THE STEM RESULTING IN JOINT INSTABILITY. DOI: (B)(6) 2008. DOR: (B)(6) 2022. UNKNOWN HIP.
Patient problems: Foreign Body Reaction; Inadequate Osseointegration; Joint Laxity
MAUDE report 15942331 ↗InjuryJun 8, 2022Fracture; Off-Label Use; Osseointegration Problem; Fracture; Off-Label Use; Osseointegration Problem
IT WAS REPORTED THAT ON APRIL 9, 2007, THE PATIENT UNDERWENT A THA SURGERY FOR OA OF HIP. AFTER SURGERY, THE SCREWS BROKE DUE TO LOOSENING OF A CUP, AND THE PATIENT HAD PAIN. ON (B)(6) 2022, A REVISION SURGERY WAS PERFORMED WITH NO SURGICAL DELAY. IN THE REVISION SURGERY, THE STEM WAS KEPT USING AS IT WAS, AND THE CUP WAS REPLACED. THE LOOSENED CUP IS A COMPETITOR¿S PRODUCT. THE SURGEON COMMENTED THAT THE CAUSE OF LOOSENING OF THE CUP MIGHT HAVE BEEN DUE TO ALVAL (ASEPTIC LYMPHOCYTIC VASCULITIS-RELATED LESION) SINCE THE IMPLANTS IMPLANTED IN A STATE OF MOM. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Foreign Body Reaction; Inadequate Osseointegration; Pain
MAUDE report 14633732 ↗
Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices
- PINNACLE 1217-25-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED735 reports
- NA 1365-52-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED683 reports
- APEX 1246-03-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 deaths613 reports
- PINNACLE 1217-20-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 deaths590 reports
- PINNACLE 1217-30-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 death484 reports
- NA 1365-53-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED404 reports
- PINNACLE ALTRX 1221-36-054Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED388 reports
- PINNACLE ALTRX 1221-32-050Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED386 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page