S-ROM 56-3120LRecalls & reported problems
Depuy Orthopaedics, Inc. · S-ROM Total Hip System FEMORAL STEM Left Long 36 STD NECK +8 11/13 TAPER 20x15x225 Lateral · Class 3 · Implant
All models of S-ROM (92)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10603295177845), received between Apr 6, 2020 and Apr 6, 2020 — classified as 1 injury.
The most frequently reported specific device problem is “Degraded” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Degraded1
Patient problems
- Adhesion(s)1
- Necrosis1
- No Code Available1
- Pain1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryApr 6, 2020Degraded; Degraded
UNF AND MEDICAL RECORDS RECEIVED. AFTER REVIEW OF MEDICAL RECORDS, PATIENT WAS REVISED TO ADDRESS SEVERE LEFT HIP PAIN. MRI SHOWED A HEAVY BURDEN OF INTRACAPSULAR PARTICLES. REVISION NOTES STATED THAT THE PATIENT WAS NOTED TO HAVE A SIGNIFICANT SCAR TISSUE, AND DEEP TO THE FASCIA, THERE WERE LARGE AREAS OF AVASCULAR NONVIABLE SCAR TISSUE. LARGE AREAS OF AVASCULAR NECROTIC TISSUE WERE REMOVED. THE PATIENT WAS NOTED TO HAVE A LARGE BARE AREAS OVERLYING THE GREATER TROCHANTER AND THE LATERAL ASPECT OF THE PROXIMAL FEMUR. THE GRANULATION TISSUE WAS ALSO IDENTIFIED AROUND THE ACETABULAR COMPONENT. EXTENSIVE BLACK MATERIAL AROUND THE TRUNNION WAS NOTED. IT APPEARED CONSISTENT WITH CORROSION. DOI: (B)(6) 2007. DOR: (B)(6) 2010, (LEFT HIP). DOI: (B)(6) 2006, (HEAD AND LINER). THIS PC IS FOR THE SECOND REVISION OF THE LEFT HIP. PLEASE SEE (B)(4) FOR THE FIRST REVISION AND (B)(4) FOR THE RIGHT HIP.
Patient problems: Adhesion(s); Necrosis; No Code Available; Pain
MAUDE report 9928760 ↗
Other Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented devices
- BIOLOX DELTA 1365-36-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented839 reports
- BIOLOX DELTA 1365-36-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented643 reports
- PINNACLE POROCOAT 1217-22-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented335 reports
- PINNACLE POROCOAT 1217-22-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented263 reports
- BIOLOX DELTA 1365-36-330Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented255 reports
- PINNACLE POROCOAT 1217-22-050Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented253 reports
- PINNACLE POROCOAT 1217-22-056Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented220 reports
- SUMMIT POROCOAT 1570-11-120Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented1 death204 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page