S-ROM 56-3517Recalls & reported problems
Depuy (Ireland) · S-ROM Total Hip System FEMORAL STEM STANDARD 36 STD NECK +6 11/13 TAPER 16x11x150 Lateral · Class 3 · Implant
All models of S-ROM (185)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This DePuy S-ROM hip stem is a metal implant placed in the thigh bone to replace the hip joint. Of its 129 reports, 127 are injuries and 2 are malfunctions, with no deaths; most describe patient pain, and several describe the stem loosening or its neck breaking off, followed by repeat surgery. Some describe a soft tissue mass around the joint, and others describe infection or dislocation. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 129
- Deaths reported
- 0
- Injuries reported
- 127
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 129 adverse event reports that list this exact device (DI 10603295178200), received between Dec 17, 2014 and Feb 26, 2026 — classified as 127 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Loss of Osseointegration” (19 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem55
- Loss of Osseointegration19
- Appropriate Device Problem Term/Code Not Available13
- Fracture13
- Corroded8
- Osseointegration Problem8
- Degraded7
- Biocompatibility6
Patient problems
- Pain73
- No Code Available57
- Test Result48
- Discomfort30
- Tissue Damage28
- Inadequate Osseointegration25
- Foreign Body Reaction18
- Inflammation14
Recent adverse event reports
The 25 most recent of 129 reports, in the reporter's own words as published by the FDA.
InjuryFeb 26, 2026Fracture
PATIENT WAS REVISED DUE TO THE SROM NECK SHEERED OFF OF STEM. ORIGINAL SURGERY DATE WAS (B)(6) 2018. DOI: 2018, DOR: (B)(6) 2026, AFFECTED SIDE: LEFT HIP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24455734 ↗InjuryNov 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ADDITIONAL INFORMATION RECEIVED STATES THAT THE PATIENT WAS REVISED DUE TO PSEUDOTUMOR FORM METAL BEARING SURFACE. REVISION NOTES REPORTED OF SEVERE EROSIVE CHANGES AND EXTENSIVE SOFT TISSUE MASS CONSISTENT WITH PSEUDOTUMOR. THE ANTERIOR CAPSULE WAS INCISED WITH RELEASE EXTENSIVE AMOUNT OF BROWN DISCOLORED FLUID. THERE WAS ALSO SIGNIFICANT TENDINOSIS VISIBLE AND BONE EROSION. CLINICAL VISITS REPORTED OF PAIN, WEAKNESS, OSTEOLYSIS, AND BONE LOSS.
Patient problems: Arthritis; Bone Notching; Cyst(s); Fatigue; Osteolysis; Pain; Pocket Erosion; Swelling/ Edema
MAUDE report 23582020 ↗InjuryJun 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT'S IMPLANTS REMOVED DUE TO INFECTED HIP, PLACED PROSTHETIC, STEM, AND ALL POLY LINER. DOI: UNKNOWN. DOR:(B)(6) 2025. AFFECTED SIDE: UNKNOWN.
Patient problems: Unspecified Infection
MAUDE report 22358975 ↗InjuryJan 15, 2025Fracture; Fracture
IT WAS REPORTED THAT IMPLANT BROKE AT NECK JUNCTION. A NEW IMPLANT WAS NEEDED. IT WAS UNKNOWN IF THERE WAS A SURGICAL DELAY. NO FURTHER DETAILS WERE PROVIDED.
Patient problems: Insufficient Information
MAUDE report 21157013 ↗InjuryOct 15, 2024Fracture; Fracture
ADDITIONAL INFORMATION RECEIVED STATES THAT; UPDATE RECEIVED FOR: DATE OF REVISION: (B)(6) 2024 TREATMENT/IMPACT: READMITTED ON (B)(6) 2024 FOR REVISION.
Patient problems: Fall; Foreign Body Reaction; Pain
MAUDE report 20451494 ↗InjuryDec 4, 2023Osseointegration Problem; Osseointegration Problem
PATIENT WAS REVISED DUE TO LOOSENING OF STEM. DOI: (B)(6) 2021 DOR: NOVEMBER 15, 2023 AFFECTED SIDE: RIGHT HIP
Patient problems: Inadequate Osseointegration
MAUDE report 18256628 ↗InjuryDec 22, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE SURGEON REVISED DUE TO LEG LENGTH DISCREPANCY FROM A DYSPLASTIC HIP. SURGEON REMOVED STEM AND FEMORAL HEAD AND THEN REMOVED A 3CM SECTION OF THE FEMUR. CANAL WAS REAMED WITH FLEXIBLE REAMERS. HE THEN TRIALED A 11X16 36 STD +6 STEM AND A 36 +9 FEMORAL HEAD. THOSE SIZES WERE THEN IMPLANTED. NO SURGICAL DELAY WAS REPORTED. DOI: (B)(6) 2022. DOR: (B)(6) 2022. AFFECTED SIDE: LEFT HIP.
Patient problems: Physical Asymmetry
MAUDE report 16033247 ↗InjuryOct 14, 2022Use of Device Problem; Use of Device Problem
PATIENT WAS REVISED DUE TO DISLOCATION. DOI: (B)(6) 2022. DOR: (B)(6) 2022. AFFECTED SIDE: LEFT HIP.
Patient problems: Insufficient Information
MAUDE report 15606406 ↗InjuryJul 12, 2022Fracture; Fracture
PATIENT WAS REVISED AS THE STEM BROKE ABOVE THE SLEEVE. THERE WAS NO SURGICAL DELAY REPORTED. THERE WAS NO LOOSENING REPORTED. DOI: (B)(6) 2004; DOR: (B)(6) 2022; AFFECTED SIDE: RIGHT HIP.
Patient problems: Insufficient Information
MAUDE report 14985793 ↗InjuryJul 11, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED ON (B)(6) 2021. SALES REP WAS UNABLE TO PULL UP REVISION RECORDS. PATIENT HAS HAD SOME HIP INSTABILITY ISSUES. PATIENT HAS EHLOR DANLOS SYNDROME. PATIENT COMPLAINED THAT THE LEG FELT TOO LONG. THE SURGEON FELT THAT IT WAS NEEDED TO HAVE THE HIP LENGTH AND OFFSET REDUCED AND THE GREATER TROCH TRANSFERRED TO TIGHTEN THE SOFT TISSUE. TRIAL REDUCTION WAS PERFORMED AND FELT TO BE STABLE. NEW STEM AND FEMORAL HEAD WERE IMPLANTED. SLEEVE AND ACETABULAR COMPONENTS WERE MAINTAINED FROM PREVIOUS SURGERY. THERE WAS NO SURGICAL DELAY. DOI: (B)(6) 2021. DOR: (B)(6) 2022. AFFECTED SIDE: LEFT HIP.
Patient problems: Joint Laxity; Physical Asymmetry
MAUDE report 14980948 ↗InjuryJun 29, 2022Degraded; Adverse Event Without Identified Device or Use Problem; Degraded; Adverse Event Without Identified Device or Use Problem
PINNACLE LITIGATION COMPLAINT RECEIVED AD (B)(6) 2022. PATIENT WAS REVISED WAS DUE TO HIGH HEAVY METAL POISONING. RECORDS ALSO ALLEGES PAIN, INJURY, TOXIC HEAVY COBALT AND CHROMIUM DEBRIS, METALLOSIS, METAL WEAR, EMOTIONAL TRAUMA AND DISTRESS. DOI: (B)(6) 2012, DOR: (B)(6) 2021 UNK HIP.
Patient problems: Discomfort; Distress; Fibrosis; Foreign Body Reaction; Hypersensitivity/Allergic reaction; Loss of Range of Motion; Metal Related Pathology; Pain; Physical Asymmetry; Swelling/ Edema; Unspecified Tissue Injury
MAUDE report 14855646 ↗InjuryDec 7, 2021Fracture; Fracture
QUESTION: A. WAS THE PATIENT EXPERIENCING ANY ADVERSE SYMPTOMS THAT LED TO THE REVISION SURGERY? ANSWER: THE BROKEN STEM I¿M SURE CAUSED ISSUES BUT DON¿T KNOW EXACTLY WHAT THAT MIGHT BE! IT WAS NEVER STATED.
Patient problems: Insufficient Information
MAUDE report 12953026 ↗InjuryOct 4, 2021Osseointegration Problem; Osseointegration Problem
ADDITIONAL INFORMATION RECEIVED INDICATED THAT BOTH STEM AND SLEEVE WERE LOOSE - UNDERSIZED. THE AFFECTED SIDE OF THE PATIENT WAS THE RIGHT HIP.
Patient problems: Inadequate Osseointegration; Pain
MAUDE report 12573912 ↗InjurySep 30, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ASR CUP WITH SROM STEM PRIMARY HIP DONE APPROX 15 YEARS AGO. PT PRESENTED WITH CARDIAC SYMPTOMS TO ED, INFECTION IN HIP NOTED. SAME REVISED TODAY. ALL IMPLANTS REMOVED, CEMENT SPACER INSERTED. DOI: 2006. DOR: (B)(6) 2021. UNKNOWN SIDE.
Patient problems: Unspecified Infection
MAUDE report 12554186 ↗InjurySep 29, 2021Degraded; Degraded
MEDICAL RECORDS RECEIVED. AFTER REVIEW OF MEDICAL RECORD, IT WAS REPORTED THAT THE PATIENT WAS REVISED FOR FAILED LEFT THA WITH RECALLED METAL ON METAL ON METAL DESIGN WITH MASSIVE PELVIC ASSOCIATED BONE LOOSE AND PSEUDOTUMOR. OPERATIVE NOTES REPORTED THAT THERE WAS A CHRONIC NON-UNION OF GREATER TROCHANTERIC FRACTURE. THERE WAS A DENSE DARK GRAY ORGANIZED MASS PROTRUDING FROM THE POSTERIOR PSEUDO CAPSULE. THERE WERE AREAS WITHIN THE MASS OF FRANK NECROSIS. DEBRIDED INTRACAPSULAR TISSUE THAT WAS SIMILARLY CONSISTENT WITH PSEUDOTUMOR. THERE WAS GROSS EVIDENCE OF TAPER JUNCTION CORROSION WITH BLACK DEBRIS ALONG THE TRUNNION. THERE WAS EXTENSIVE BLACK NECROTIC PASTRY DEBRIS BEHIND THE CUP. THE ANTERIOR COLUMN WAS LARGELY ABSENT DOWN INTO A SIGNIFICANT PORTION OF THE PUBIC ROOT. THERE WAS ESSENTIALLY NO ANTERIOR WALL. THERE WAS A SIGNIFICANT CAVITY EXTENDING UP THE ILIUM SIMILARLY FILLED WITH BLACK PASTY DEBRIS THAT WAS DEBRIDED. DOI: (B)(6) 2009, DOR: (B)(6) 2020, LEFT HIP.
Patient problems: Foreign Body Reaction; Hypersensitivity/Allergic reaction; Necrosis; Nonunion/Delayed-union Bone Fracture; Unspecified Tissue Injury
MAUDE report 12547657 ↗InjurySep 23, 2021Degraded; Migration; Degraded; Migration
(B)(4). LITIGATION PAPERS ALLEGE PAIN AND DISCOMFORT IN HER LEFT HIP CAUSING POTENTIAL UNNECESSARY AND ADDITIONAL SURGERIES, DIMINUTION IN EARNING CAPACITY, LOST WAGES, MEDICAL MONITORING EXPENSES, EMOTIONAL DISTRESS, LOSS OF ENJOYMENT OF LIFE, METALLOSIS EXPOSURE AND OTHER INJURIES PRESENTLY UNDIAGNOSED. FURTHER ALLEGED THE PATIENT WAS INJURED BY EXCESSIVE LEVELS OF CHROMIUM AND COBALT. DOI: (B)(6) 2008 - DOR: NONE REPORTED (LEFT HIP) PATIENT IS RESIDENT OF OR UPDATE: 12/6/2011 PLAINTIFF'S PRELIMINARY DISCLOSURE (PPD) FORM RECEIVED (B)(6) 2011. THERE WAS NO NEW INFORMATION RECEIVED THAT WOULD IMPACT THE OUTCOME OF THE INVESTIGATION. UPDATE AD 19 JUN 2019: (B)(4) HAS BEEN RE-OPENED UNDER (B)(4) DUE ASR PLAINTIFF DISCLOSURE FORM RECORDS RECEIVED. PDF INDICATED THAT THE PATIENT HAD UNDERWENT REVISION SURGERY. REPORTED REVISION DATE. DOI: (B)(6) 2008. DOR: (B)(6) 2019. LEFT HIP.
Patient problems: Ambulation or Postural Difficulties; Anxiety; Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Discomfort; Distress; Limb Fracture; Loss of Range of Motion; Muscular Rigidity; Pain; Physical Asymmetry
MAUDE report 12518573 ↗InjurySep 20, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PPF AND IMPLANT RECORD RECEIVED. PATIENT HAD AN INFECTION AND DISLOCATION AFTER 1ST REVISION. DOI: (B)(6) 2011. DOR: (B)(6) 2017, (LEFT HIP).
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 12495768 ↗InjuryJul 8, 2021Fracture; Fracture
PATIENT PRESENTED WITH A SROM HIP THAT WAS DONE ABOUT 10 YEARS AGO. ONE OF THE DISTAL TIPS HAD FRACTURED FROM THE STEM. SURGEON REMOVED STEM AND SLEEVE AND CHANGED THE INSERT. ALL PARTS WERE RETRIEVED FROM THE PATIENT. DOI: (B)(6) 2011 - (B)(6) 2021 (RIGHT HIP)
Patient problems: Insufficient Information
MAUDE report 12135655 ↗InjuryJun 15, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE ORIGINAL IMPLANT DATE UNKNOWN. PATIENT SUFFERED A HIP DISLOCATION OF SROM HIP AND CONSTRAINED LINER. SURGEON WAS NOT AWARE OF THE PATIENTS PREVIOUS SURGICAL DATES OR REASONS FOR SURGERY. SURGEON FELT THIS PATIENT WAS DEALING WITH SOME HIP IMPINGEMENT OF THEIR +4/10 DEGREE CONSTRAINED LINER ON THE NECK OF THE SROM STEM. HE FELT THEY WOULD BENEFIT FROM CONVERTING THE LINER TO JUST A +4 NEUTRAL CONSTRAINED LINER. HE ALSO FELT THE PATIENT WOULD BENEFIT FROM REPLACING THE STEM SO HE COULD ADJUST THE VERSION A BIT. A TRIAL REDUCTION WAS PERFORMED WITH THE NEW STEM AND A +4 NEUTRAL CONSTRAINED LINER AND A 28MM +3 HEAD WAS FOUND TO BE SATISFACTORY. THE REAL LINER AND HEAD WERE OPENED AND IMPLANTED. THE HIP WAS REDUCED AND FOUND TO BE STABLE. THE CLOSING PROCESS BEGAN. DOI: UNKNOWN; DOR: (B)(6) 2021 (RIGHT HIP).
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 12003216 ↗InjuryJun 8, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT HAD RISING METAL ION LEVELS AND IT WAS SUGGESTED THAT SHE HAVE HER ASR CUP REMOVED AND REPLACE WITH A STANDARD CUP AND POLY LINER. RIGHT HIP. THE ASR CUP WAS REMOVED AND A COMPETITOR MDM SYSTEM WAS REPLACED. SURGEON STATED THAT THE HIP LOOKED PRISTINE WHEN HE WENT IN AND WAS ABLE TO EASILY REMOVE THE CUP AND ASR HEAD AND REPLACED WITH THE COMPETITOR MDM SYSTEM. HE PLACED AN SROM +0 CERAMIC 28MM HEAD. DOI: UNK, DOR: (B)(6) 2021, RIGHT HIP.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 11957955 ↗InjuryApr 15, 2021Fracture; Fracture
HIP REVISION TO REMOVE BROKEN SROM AND PATIENT REPORT OF PAIN.
Patient problems: Pain
MAUDE report 11670674 ↗InjuryFeb 25, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS REVISED DUE TO HIP AND LEG PAIN. STEM AND HEAD REMOVED AND REVISED WITH COMPETITORS COMPONENT. COMPETITOR CUP WAS NOT REVISED. NO SURGICAL DELAY. DOI: (B)(6) 2017. DOR: (B)(6) 2021; RIGHT HIP.
Patient problems: Pain
MAUDE report 11378150 ↗InjuryFeb 5, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
CLINICAL NOTIFICATION RECEIVED FOR REVISION OF RIGHT TOTAL HIP ARTHROPLASTY TO ADDRESS DISLOCATION AND METALLOSIS. ON (B)(6) 2007, THE PATIENT HAD A RIGHT PRESS-FIT TOTAL HIP ARTHROPLASTY TO ADDRESS END-STAGE OSTEOARTHRITIS. DEPUY COMPONENTS WERE UTILIZED, INCLUDING S-ROM STEM. THE CUP WAS NOTED TO HAVE BEEN PLACED IN ABOUT 40-45 DEGREES OPEN ABDUCTION AND ABOUT 25 DEGREES ANTEVERSION THAT MATCHED HER VERSION ANATOMICALLY. THERE WERE NO COMPLICATIONS NOTED DURING THE SURGERY. ON (B)(6) 2021, THE PATIENT HAD A REVISION RIGHT TOTAL HIP ARTHROPLASTY WITH ACETABULAR CUP/LINER AND FEMORAL HEAD, TO ADDRESS FAILED RIGHT HIP, RECURRENT INSTABILITY, METALLOSIS WITH MILD ADVERSE REACTION TO METAL DEBRIS. THE PATIENT HAS HAD MULTIPLE DISLOCATION WHICH INCLUDED PAIN AND SWELLING PRIOR TO SURGERY. THE SURGEON INCLUDED IN HIS FINDINGS: MILD ADVERSE REACTION TO METAL DEBRIS, MOST LIKELY IMPINGEMENT OF THE NECK ON THE METAL LINER, MILD TRUNNIONOSIS ¿LITTLE BIT OF LYSIS AND/OR JUST STRESS SHIELDING IN THE BONE IN THE ACETABULUM, AND A PSEUDOTUMOR. THE ACETABULAR CUP WAS NOTED TO BE ¿SLIGHTLY MALPOSITIONED, SLIGHTLY VERTICAL, SLIGHTLY OVER-ANTEVERTED¿. DEPUY COMPONENTS WERE IMPLANTED DURING THIS PRO
Patient problems: Foreign Body Reaction; Hypersensitivity/Allergic reaction; Joint Laxity; Pain; Swelling/ Edema
MAUDE report 11287877 ↗InjuryFeb 5, 2021Fracture; Fracture
PATIENT PRESENTED TO EMERGENCY DEPARTMENT. FRACTURED SROM STEM JUST BELOW THE TOP OF THE SLEEVE - IMPLANTED 10 YEARS AGO AT (B)(6) HOSPITAL. THE REVISION SURGERY HAS BEEN CONFIRMED AS HAVING TAKEN PLACE ON (B)(6) 2021. THE PATIENT ((B)(6) YO) HAD TO HAVE THE STEM AND THE SLEEVE REMOVED AS THE SURGEON COULDN'T MANAGE TO REMOVE THE DISTAL STEM ON IT¿S OWN. ENOUGH BONE REMAINED TO USE AN SROM STEM FOR THE REVISION WITHOUT HAVING TO USE A DISTALLY FIXING STEM.
Patient problems: Insufficient Information
MAUDE report 11285663 ↗InjuryJan 4, 2021Migration; Migration
IT WAS REPORTED THAT THE STEM ON RIGHT SIDE BECAME INFECTED AND SUBSIDED SUBSTANTIALLY AND LOOSENING OF THE UNKNOWN COMPONENT AT BONE TO IMPLANT INTERFACE. WE REMOVED FEMORAL AND ACETABULAR CONSTRUCTS AND REPLACED WITH PROSTALAC. DOI: (B)(6) 2018 DOR: (B)(6) 2020 RIGHT HIP.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 11108289 ↗
Other Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented devices
- BIOLOX DELTA 1365-36-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented839 reports
- BIOLOX DELTA 1365-36-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented643 reports
- PINNACLE POROCOAT 1217-22-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented335 reports
- PINNACLE POROCOAT 1217-22-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented263 reports
- BIOLOX DELTA 1365-36-330Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented255 reports
- PINNACLE POROCOAT 1217-22-050Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented253 reports
- PINNACLE POROCOAT 1217-22-056Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented220 reports
- SUMMIT POROCOAT 1570-11-120Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented1 death204 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page