P.F.C. SIGMA 86-4180Recalls & reported problems
Depuy Orthopaedics, Inc. · P.F.C. SIGMA TIBIAL TRAY FIXED BEARING KEEL TITANIUM 2 CEMENTED · Class 2 · Implant
All models of P.F.C. SIGMA (18)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10603295216421), received between Jul 20, 2018 and Apr 23, 2024 — classified as 8 injuries.
The most frequently reported specific device problem is “Loss of or Failure to Bond” (3 reports).
7 other recalls cover devices cleared under the same FDA submission (K830927, K961685); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Loss of or Failure to Bond3
- Adverse Event Without Identified Device or Use Problem2
- Loose or Intermittent Connection2
- Migration2
- Fracture1
Patient problems
- No Code Available4
- Insufficient Information2
- Not Applicable2
- Pain2
- Unspecified Infection2
- Osteolysis1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryApr 23, 2024Loss of or Failure to Bond; Loss of or Failure to Bond
A DEPUY SYNTHES IMPLANT WAS REVISED AND REVIEWED FOR ANALYSIS. REASON FOR REVISION: LOOSE (UNKNOWN COMPONENTS OR INTERFACE). AT REVISION TIBIAL INSERT SHOWED MINOR YELLOWING AND SEVERAL POLY LOSS.
Patient problems: Insufficient Information
MAUDE report 19166005 ↗InjuryAug 29, 2022Loss of or Failure to Bond; Migration; Loss of or Failure to Bond; Migration
PRIMARY PFC FB OPERATION APPROX 15 YEARS AGO. TIBIA NEEDED REVISING. CR FEMUR REMAINS INSITU - TIBIA REMOVED AND MBT TRAY WITH SLEEVE AND STEM INSERTED DOI: JAN 2007. DOR: (B)(6) 2022. UNKNOWN SIDE.
Patient problems: Insufficient Information
MAUDE report 15312021 ↗InjuryDec 20, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
INFECTED LEFT TOTAL KNEE.
Patient problems: No Code Available; Unspecified Infection
MAUDE report 9504387 ↗InjuryOct 28, 2019Fracture; Fracture
IT WAS REPORTED THAT THE AFFECTED SIDE IS THE RIGHT KNEE. REMOVED SIGMA PLI GVF INS 2 8MM, PFC*SIGMA/OV DOME PAT 3 PEG,32, DEPUY CMW 1G 40G AND PFC*SIGMA C/R NPOR FEM RT SZ 2 FROM D11 FIELD.
Patient problems: No Code Available
MAUDE report 9247172 ↗InjuryJun 18, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED TO ADDRESS INFECTION. COMPETITOR CEMENT WAS USED. IT WAS ALSO REPORTED THAT IT WAS A (B)(6). COMPONENTS WERE REMOVED AND ANTIBIOTIC ARTICULATING SPACER WAS IMPLANTED. ONLY CATALOG AND LOT NUMBER ON TIBIA COMPONENT AVAILABLE. FEMUR WAS A 2RIGHT CR CEMENTED. 2X10 XLK CR FB INSERT. DOI: (B)(6) 2015, DOR: (B)(6) 2019, RIGHT KNEE.
Patient problems: No Code Available; Unspecified Infection
MAUDE report 8707746 ↗InjuryMay 15, 2019Loss of or Failure to Bond; Migration; Loss of or Failure to Bond; Migration
THE PRIMARY SURGERY WAS PERFORMED ON (B)(6) 2008 VIA TKA. IT WAS REPORTED THAT THE REVISION SURGERY WAS PERFORMED ON (B)(6) 2019, BY REPLACING THE TRAY (P/N: 864180) AND THE INSERT (P/N: 960420) WITH A SLEEVE (P/N: UNKNOWN) OF M.B.T. REVISION SYSTEM DUE TO A LOOSENING ON TIBIAL SIDE. THE TIBIAL SIDE HAD TILTED IN MEDIAL DIRECTION. NO LOOSENING WAS OBSERVED ON FEMORAL SIDE AND THERE WAS NO INFECTION. THERE WAS NO SURGICAL DELAY.
Patient problems: No Code Available
MAUDE report 8611688 ↗InjuryAug 15, 2018Loose or Intermittent Connection; Loose or Intermittent Connection
CLINICAL NOTIFICATION RECEIVED FOR REVISION RELATED TO LOOSENING OF TIBIAL AND FEMORAL COMPONENTS AND OSTEOLYSIS. DOI: (B)(6) 2000. DOR: (B)(6) 2018. (LEFT KNEE).
Patient problems: Not Applicable; Osteolysis; Pain
MAUDE report 7784071 ↗InjuryJul 20, 2018Loose or Intermittent Connection; Loose or Intermittent Connection
THE PRIMARY SURGERY WAS PERFORMED TO IMPLANT THE IMPLANTS TO THE PATIENT¿S LEFT KNEE JOINT (DATE OF PRIMARY SURGERY WAS UNKNOWN). THE PATIENT COMPLAINED THE PAIN (ABOUT A YEAR AGO), SO THAT THE SURGEON FOUND THE LOOSENING OF THE KNEE IMPLANTS. IT WAS REPORTED THAT THE REVISION SURGERY WAS PERFORMED ON (B)(6) 2018 DUE TO THE LOOSENING. NO FURTHER INFORMATION WAS PROVIDED BY THE HOSPITAL.
Patient problems: Not Applicable; Pain
MAUDE report 7706427 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K830927, K961685. One submission often covers many models, so check the product and lot details in each record.
- Z-0057-2007 · Sep 14, 2006 · TerminatedDepuy PFC Oval Dome Patella (Part Number 960100)3-Peg, 32 mm, sterile; REF 96-0100.
- Z-1319-05 · May 26, 2005 · TerminatedDepuy SIGMA curved plus insert; polyethylene tibial insert, Sz 5, 10.0 Ref 97-0371.
- Z-1318-05 · May 26, 2005 · TerminatedDepuy SIGMA curved plus insert; polyethylene tibial insert, Sz 4, 10 mm; Ref 97-0361.
- Z-1317-05 · May 26, 2005 · TerminatedDepuy SIGMA curved plus insert; polyethylene tibial insert, Sz 3, 15.0 mm; Ref 97-0353.
- Z-1315-05 · May 26, 2005 · TerminatedDepuy SIGMA curved plus insert; polyethylene tibial insert, Sz 2.5, 8 mm; Ref 97-0340.
- Z-1320-05 · May 26, 2005 · TerminatedDepuy SIGMA curved plus insert; polyethylene tibial insert, Sz 5, 15.0 mm; Ref 97-0373.
- Z-1316-05 · May 26, 2005 · TerminatedDepuy SIGMA curved plus insert; polyethylene tibial insert, Sz 3, 12.5 mm; Ref 97-0352.
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page