Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10603295216926), received between Mar 27, 2020 and Mar 21, 2023 — classified as 6 injuries.
The most frequently reported specific device problem is “Loss of or Failure to Bond” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem4
- Loss of or Failure to Bond2
- Fracture1
Patient problems
- Unspecified Infection4
- Pain2
- Swelling/ Edema2
- Ambulation or Postural Difficulties1
- Fluid Discharge1
- Hypersensitivity/Allergic reaction1
- Insufficient Information1
- No Code Available1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryMar 21, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT HAD PAIN AND STERLING AND WAS TESTED FOR NICKEL ALLERGY. THE TEST CAME BACK POSITIVE SO THE SURGEON TOOK IT ALL OUT AND PUT IN AN AESCULAP REVISION CONSTRUCT. DOI: UNKNOWN. DOR: (B)(6) 2023. AFFECTED SIDE: RIGHT KNEE.
Patient problems: Hypersensitivity/Allergic reaction; Pain
MAUDE report 16584905 ↗InjuryFeb 21, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT RECEIVED AN I&D WITH A REVISION OF THE LEFT KNEE LPS HINGED INSERT TO TREAT PAIN AND SWELLING SECONDARY TO (B)(6) INFECTION. THE PATIENT WAS PLACED ON IV ANTIBIOTICS. THE PROCEDURE WAS COMPLETED WITHOUT COMPLICATIONS. DOI: (B)(6) 2015, DOR: (B)(6) 2016, (LEFT KNEE).
Patient problems: Swelling/ Edema; Unspecified Infection
MAUDE report 13567743 ↗InjuryFeb 17, 2022Loss of or Failure to Bond; Loss of or Failure to Bond
AMPUTATION. THE PATIENT RECEIVED AN ABOVE THE KNEE AMPUTATION DUE TO SIGNIFICANT PAIN, INFECTION, SWELLING, DRAINING, LYMPHEDEMA AND INABILITY TO AMBULATE. UPON ENTERING THE JOINT, THE CONSTRUCT WAS NOTED TO BE GROSSLY LOOSE AND INFECTED. THE AMPUTATION WAS COMPLETED WITHOUT INDICATION OF COMPLICATION BY THE SURGEON. DOI: (B)(6) 2017; DOR: (B)(6) 2021; UNKNOWN KNEE . THIS COMPLAINT IS ALSO LINKED TO (B)(4) WHERE THE TIBIAL TRIAL WAS REPORTED. ANOTHER PC IS LINKED TO THIS COMPLAINT UNDER (B)(4) WHERE THE EVENT HAPPENED IN BETWEEN APR 2017 ¿ MAR 2019.
Patient problems: Ambulation or Postural Difficulties; Fluid Discharge; Pain; Swelling/ Edema; Unspecified Infection
MAUDE report 13549622 ↗InjuryFeb 17, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED AMPUTATION WAS DONE. DOI - (B)(6) 2015; DOR - (B)(6) 2016; UNKNOWN KNEE SIDE.
Patient problems: Unspecified Infection
MAUDE report 13549339 ↗InjuryMar 29, 2021Loss of or Failure to Bond; Fracture; Loss of or Failure to Bond; Fracture
SURGEON REPORTED TO BFARM: "THE KNEE ENDOPROSTHESIS WAS IMPLANTED IN 2002. SINCE 2013, LOOSENING OF THE PROSTHESIS AND FRACTURE OF THE TIBIAL STEM IS KNOWN. DUE TO MULTIPLE INTERNAL DISEASES, THE PATIENT WAS NOT ABLE TO UNDERGO SURGERY UP TO NOW. NOW, KNEE-TEP REMOVAL WAS PERFORMED BECAUSE OF KNEE-TEP LOOSENING AND STEM-FRACTURE."
Patient problems: Insufficient Information
MAUDE report 11584827 ↗InjuryMar 27, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS INFECTED. SURGEON REMOVED EVERYTHING. DOI: (B)(6) 2019, (B)(6) 2019 (INSERT). DOR: (B)(6) 2020.
Patient problems: No Code Available; Unspecified Infection
MAUDE report 9891432 ↗
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page