Skip to content

P.F.C. 86-6887Recalls & reported problems

Depuy Orthopaedics, Inc. · P.F.C. TIBIAL FLUTED ROD TRIAL 20mm x 115mm · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10603295217695), received between Jul 25, 2018 and Oct 24, 2019 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break1
  • Detachment of Device or Device Component1

Patient problems

  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 24, 2019Break; Break

    BROKEN CONNECTOR.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9234322 ↗
  • MalfunctionJul 25, 2018Detachment of Device or Device Component; Detachment of Device or Device Component

    PRODUCT NAME: PFC STEM TRIAL EXTRACTOR; PRODUCT CODE: 865226; LOT/BATCH/EXP: TBC; WAS THE PRODUCT BEING USED IN A CLINICAL TRIAL? NO; DID THE EVENT HAPPEN DURING A PROCEDURE? YES; WERE YOU IN THE PROCEDURE AT THE TIME OF THE EVENT? YES; EVENT OUTCOME/HOW WAS IT MANAGED? BROKEN INSTRUMENT REMOVED USING COLD CHISEL AND MOLEGRIPS. WAS THERE ANY CONSEQUENCE TO THE PATIENT DUE TO THE EVENT? WAS THE SURGERY PROLONGED DUE TO THE EVENT? IF YES, CONFIRM HOW MANY MINUTES DELAY ¿ SURGERY DELAY WAS APPROXIMATELY 20 MINUTES. HAS THE REPORTER FACILITY INDICATED THERE MAY BE LEGAL ACTION? NO. IS THE PRODUCT AVAILABLE FOR RETURN? YES AFTER CLEANING AND DECON. PLEASE GIVE A DETAILED EXPLANATION OF THE EVENT: THE OPERATION WAS A DISTAL FEMUR TO SLEEVE CASE. THE SURGEON HAS BEEN USING THE KIT FOR OVER 5 YEARS AND IS A KOL AND VISITATION FOR US. AFTER THE TRIAL THE SURGEON REMOVED THE TRIAL COMPONENTS AND, AS IF OFTEN THE CASE, IN THESE PATIENTS THE 31MM SLEEVE TRIAL AND 20 X 115 STEM TRIAL DISASSOCIATED FROM THE TRIAL DISTAL FEMUR. THE SURGEON INSERTED THE STEM TRIAL EXTRACTOR, ATTACHED THE SLAP HAMMER AND TRIED TO REMOVE THE TRIAL SLEEVE AND STEM. THE END OF THE INSTRUMENT SHEAR OFF AT THE POINT WHE

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7717762 ↗

Other TEMPLATE devices

All TEMPLATE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page