SIGMA 94-0022Recalls & reported problems
Depuy (Ireland) · SIGMA FEMORAL CRUCIATE RETAINING POROCOAT SIZE 2 RIGHT · Class 2 · Implant
All models of SIGMA (380)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10603295223467), received between Aug 16, 2017 and Sep 14, 2021 — classified as 3 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Insufficient Device Problem Information1
- Loss of Osseointegration1
Patient problems
- Pain3
- No Code Available2
- Inadequate Osseointegration1
- Joint Swelling1
- Osteolysis1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjurySep 14, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT HAS A RIGHT SIGMA TOTAL KNEE, DONE ¿SOMETIME IN 2012¿ PER SURGEON. PT WAS REVISED FOR PAIN, COMPONENTS WERE WELL FIXED, SURGEON CONVERTED TO POSTERIOR STABILIZED FEMUR COMPONENT. THE TIBIA COMPONENT WAS RETAINED AND NOT REMOVED. DOI: 2012, DOR: (B)(6) 2021, RIGHT KNEE.
Patient problems: Pain
MAUDE report 12471049 ↗InjuryMay 20, 2020Loss of Osseointegration; Loss of Osseointegration
REVISION TOTAL KNEE REPLACEMENT: THE PATIENT PRESENTED TO THE SURGEON WITH A PAINFUL KNEE. BONE SCAN SHOWED THE FEMUR AND PATELLA WERE LOOSE. INFECTION MARKERS ALL WITHIN NORMAL RANGE. WHEN IMPLANTS WERE REMOVED THERE WERE SIGNS OF OSTEOLYSIS BEHIND THE FEMORAL COMPONENT CENTRAL TO THE ANTERIOR CHAMFER REGION. NO OBVIOUS SIGNS OF OSTEOLYSIS ON THE POLY OF THE INSERT OR PATELLA. PHOTOS TAKEN OF IMPLANTS. IMPLANTS DISCARDED. NO AE REPORTED. NO DELAY IN THE SURGERY.
Patient problems: Inadequate Osseointegration; No Code Available; Osteolysis; Pain
MAUDE report 10075192 ↗InjuryAug 16, 2017Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT THERE WAS A FAILURE OF THE INSERT, DUE TO "CLUNK" NOISE COMING FROM THE KNEE. UPDATE 5-1-2017: MEDICAL RECORDS HAVE BEEN REVIEWED. PRIMARY OPERATIVE NOTE (B)(6) 2010 INDICATES THE PATIENT RECEIVED A RIGHT PFC TOTAL KNEE REPLACEMENT. NO COMPLICATIONS NOTED BY THE SURGEON. CLINICAL NOTES (B)(6) 2017 INDICATE THE PATIENT WAS BEING SEEN FOR INSTABILITY, PAIN, SWELLING AND CLUNKING OF THE RIGHT TOTAL KNEE REPLACEMENT. IT CAN BE GATHERED THAT THE REPORTED "CLUNK" IS MORE OF A SENSATION THAN A NOISE. REVISION OPERATIVE NOTES (B)(6) 2017 INDICATE DURING THE SURGERY THE SURGEON NOTED SEVERE POSTERIOR POLYETHYLENE WEAR, NO EVIDENCE OF INFECTION, COMPONENTS WERE WELL FIXED. THERE WAS ALSO NATURAL PATELLA BAJA WITH CHONDRAL DAMAGE. THE REVISION WAS COMPLETED WITHOUT ANY COMPLICATION NOTED BY THE SURGEON. REPORTABILITY HAS BEEN UPDATED. UPDATED 8-16-2017.
Patient problems: Joint Swelling; No Code Available; Pain
MAUDE report 6797710 ↗
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page