SIGMA 9505-01-121Recalls & reported problems
Depuy (Ireland) · SIGMA HIGH PERFORMANCE PATELLA RESECTION GUIDE · Class 1
All models of SIGMA (380)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 10603295224815), received between Jan 23, 2018 and Feb 22, 2023 — classified as 17 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (11 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Device-Device Incompatibility11
- Break5
- Naturally Worn2
- Component Missing1
- Difficult to Insert1
- Material Twisted/Bent1
- Mechanical Jam1
- Positioning Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- No Consequences Or Impact To Patient5
- Not Applicable3
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 22, 2023Break; Break
IT WAS REPORTED THE IMPACTORS ARE DAMAGED. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16420299 ↗MalfunctionFeb 21, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THE RATCHET MECHANISM IS NO LONGER WORKING. NO SURGICAL DELAY. NO ADVERSE AFFECTS ON THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16408993 ↗MalfunctionNov 21, 2022Device-Device Incompatibility; Naturally Worn; Device-Device Incompatibility; Naturally Worn
IT WAS REPORTED THAT THE PATELLA RESECTION CLAMP WON'T STAY CLAMPED. TEETH ARE WORN. THERE WAS NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15833554 ↗MalfunctionFeb 9, 2021Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT PATELLA CUTTER CLASP WILL NOT HOLD. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11305438 ↗MalfunctionFeb 9, 2021Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT PATELLA CUTTER CLASP WILL NOT HOLD. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11305435 ↗MalfunctionFeb 9, 2021Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT PATELLA CUTTER CLASP WILL NOT HOLD. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11305434 ↗MalfunctionFeb 9, 2021Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT PATELLA CUTTER CLASP WILL NOT HOLD. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11305430 ↗MalfunctionFeb 9, 2021Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT PATELLA CUTTER CLASP WILL NOT HOLD. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11305427 ↗MalfunctionFeb 9, 2021Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT PATELLA CUTTER CLASP WILL NOT HOLD. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11305426 ↗MalfunctionFeb 13, 2020Break; Component Missing; Break; Component Missing
IT WAS REPORTED THAT THE DEPTH PADDLE BROKE OFF OF THE SIGMA PATELLA CUTTING GUIDE. THE PIECE WENT MISSING IN THE WASHER STERILIZER. THE CLAMP IS BEING RETURNED WITHOUT THE BROKEN PIECE. NO SURGICAL DELAY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9709914 ↗MalfunctionJan 23, 2020Device-Device Incompatibility; Naturally Worn; Device-Device Incompatibility; Naturally Worn
IT WAS REPORTED THAT DURING TKA WITH SIGMA HP INSTRUMENTS THE DOCTOR STATED THAT THE PATELLA CLAMP WAS WORN AND NOT HOLDING. ANOTHER SET WAS OPENED AND THE DOCTOR FELT THAT THE CLAMP WAS ALSO LOOSE. BOTH NEED TO BE REPLACED DUE TO WEAR.
Patient problems: Not Applicable
MAUDE report 9622235 ↗MalfunctionDec 20, 2019Device-Device Incompatibility; Material Twisted/Bent; Device-Device Incompatibility; Material Twisted/Bent
IT WAS REPORTED THAT THE SURGEON WAS USING THE PATELLA RESECTION GUIDE TO CUT THE PATELLA, BUT IT KEPT UNLOCKING CAUSING THE DOCTOR TO READ JUST THE INSTRUMENT. THERE WASN¿T ANY SIGNIFICANT DELAY TO THE SURGERY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9505813 ↗MalfunctionMay 29, 2019Device-Device Incompatibility; Mechanical Jam; Device-Device Incompatibility; Mechanical Jam
IT WAS REPORTED THAT DURING TKA, THE PATELLA CLAMP WAS NOT CATCHING TO HOLD THE PATELLA FOR A RESECTION CUT.
Patient problems: Not Applicable
MAUDE report 8649485 ↗MalfunctionMar 28, 2019Difficult to Insert; Positioning Problem; Difficult to Insert; Positioning Problem
(B)(4). SIGMA HP PAT RES GUIDE - NOT WORKING. THIS WAS DISCOVERED TO BE FAULTY DURING A CASE YESTERDAY. THE SURGERY WAS NOT SLOWED DOWN AND THE PATIENT WAS NOT AFFECTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8460813 ↗MalfunctionMar 27, 2019Break; Break
IT WAS REPORTED THAT THE PATELLA RESECTION DEVICE WHICH IS THE CLAW FROM SIGMA SNAPS OPEN AFTER BEING CLAMPED ON PATELLA. THIS HAPPEN AS SOON AS THE SAW VIBRATES IT.
Patient problems: Not Applicable
MAUDE report 8456290 ↗MalfunctionFeb 5, 2018Break; Break
IT WAS REPORTED THAT THE PATELLA CLAMP TIP BROKE IN THE RATCHET MECHANISM. NO SURGICAL DELAY. ALL PIECES WERE RETRIEVED FROM THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7240869 ↗MalfunctionJan 23, 2018Break; Break
IT WAS REPORTED THAT THE PATELLA CUTTING GUIDE HAD PIECE BREAK OFF IN THE RATCHET MECHANISM AND ALL PIECES WERE RETRIEVED. NO SURGICAL DELAY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7211816 ↗
Other TEMPLATE devices
- ATTUNE 2544-01-014Depuy (Ireland) · TEMPLATE802 reports
- ATTUNE 2545-00-546Depuy (Ireland) · TEMPLATE1 recall301 reports
- ARTICUL/EZE 2531-51-000Depuy (Ireland) · TEMPLATE323 reports
- DELTA XTEND 2307-95-000Depuy (Ireland) · TEMPLATE315 reports
- ATTUNE 2545-00-547Depuy (Ireland) · TEMPLATE1 recall231 reports
- ATTUNE 2544-00-520Depuy (Ireland) · TEMPLATE279 reports
- ARTICUL/EZE 2530-91-000Depuy (Ireland) · TEMPLATE271 reports
- ATTUNE 2545-00-545Depuy (Ireland) · TEMPLATE1 recall221 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page