SIGMA 9505-01-263Recalls & reported problems
Depuy (Ireland) · SIGMA HIGH PERFORMANCE FIXED REFERENCE FEMORAL SIZER · Class 1
All models of SIGMA (380)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 10603295225669), received between Dec 22, 2017 and Sep 19, 2024 — classified as 13 malfunctions.
The most frequently reported specific device problem is “Material Twisted/Bent” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Twisted/Bent3
- Mechanical Jam3
- Appropriate Device Problem Term/Code Not Available2
- Failure to Disconnect1
- Incorrect Measurement1
- Material Deformation1
- Material Discolored1
- Naturally Worn1
Patient problems
- No Consequences Or Impact To Patient7
- No Clinical Signs, Symptoms or Conditions5
- No Code Available1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 19, 2024Incorrect Measurement; Incorrect Measurement
FAULTY FEMORAL SIZING GUIDE NO FURTHER INFORMATION IS AVAILABLE AT PRESENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20255264 ↗MalfunctionJan 31, 2024Naturally Worn; Naturally Worn
IT WAS REPORTED THAT THE SIZE BROKE DUE TO WEAR AND TEAR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18615462 ↗MalfunctionJun 23, 2022Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT THE STYLUS ON THE HP SIZER GUIDE IS COLD WELDED AND WILL NOT MOVE. NO PIECES BROKE OFF OF THE INSTRUMENT. THERE WAS NO DELAY SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14790647 ↗MalfunctionApr 28, 2021Material Deformation; Material Deformation
IT WAS REPORTED THAT ONE OF THE POSTERIOR FEET OF THE FIXED REFERENCE SIZER WAS FOUND BENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11738652 ↗MalfunctionOct 27, 2020Physical Resistance/Sticking; Physical Resistance/Sticking
DURING CASE THE FEMORAL SIZER AND THE FEMORAL POSTERIOR REFERENCING FEMORAL ROTATION GUIDES WERE NOT GOING TOGETHER EASILY. THEN ONCE WE GOT IT ON IT WAS EXTREMELY DIFFICULT TO TAKE OFF. HAD TO USE A HOHMAN TO REMOVE THE FEMORAL ROTATION GUIDE.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Code Available
MAUDE report 10741443 ↗MalfunctionApr 29, 2020Material Discolored; Material Twisted/Bent; Material Discolored; Material Twisted/Bent
INSTRUMENTS SHOW SIGNS OF WEAR AND DISCOLOURATIONS. NOTICED DURING CLEANING PROCESS. NO SURGICAL AFFECT. NO PATIENT AFFECT. INSTRUMENTS ARE FURTHER IN USE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10012585 ↗MalfunctionMay 22, 2019Material Twisted/Bent; Material Twisted/Bent
STYLUS FOR FIXED REF FEMORAL SIZER WILL NOT SCREW DOWN. PATIENT STATUS/OUTCOME/CONSEQUENCES: NO.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8631743 ↗MalfunctionFeb 22, 2019Mechanical Jam; Mechanical Jam
THE INSTRUMENTS MOVABLE PART / SENSING ELEMENT BECAME STUCK AND IS NO LONGER POSSIBLE TO MOVE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8361520 ↗MalfunctionFeb 11, 2019Failure to Disconnect; Failure to Disconnect
THE INSTRUMENT IS NO LONGER POSSIBLE TO DISASSEMBLE AS THE SCREW WAS TIGHTENED TOO TIGHTLY
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8327998 ↗MalfunctionJan 10, 2019Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT DURING AN INSTRUMENT DEMO WITH THE SURGEON, THE FEMORAL SIZER WAS NOTED TO BE BENT AND DAMAGED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8235242 ↗MalfunctionJan 9, 2018Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT THE INSTRUMENT WAS DEFECTIVE ON THE KIT WHEN IT ARRIVED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7176343 ↗MalfunctionJan 9, 2018Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
THE INSTRUMENT´S WINDING BUTTON IS JAMMED AND IS NOT POSSIBLE TO MOVE ANY LONGER. NO ADVERSE EFFECT FOR PATIENT, NO SURGERY PROLONGATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7174970 ↗MalfunctionDec 22, 2017Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE KNOB WILL NOT LOOSEN.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7142543 ↗
Other TEMPLATE devices
- ATTUNE 2544-01-014Depuy (Ireland) · TEMPLATE802 reports
- ATTUNE 2545-00-546Depuy (Ireland) · TEMPLATE1 recall301 reports
- ARTICUL/EZE 2531-51-000Depuy (Ireland) · TEMPLATE323 reports
- DELTA XTEND 2307-95-000Depuy (Ireland) · TEMPLATE315 reports
- ATTUNE 2545-00-547Depuy (Ireland) · TEMPLATE1 recall231 reports
- ATTUNE 2544-00-520Depuy (Ireland) · TEMPLATE279 reports
- ARTICUL/EZE 2530-91-000Depuy (Ireland) · TEMPLATE271 reports
- ATTUNE 2545-00-545Depuy (Ireland) · TEMPLATE1 recall221 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page