Skip to content

P.F.C. SIGMA 96-1003Recalls & reported problems

Depuy (Ireland) · P.F.C. SIGMA CRUCIATE RETAINING FEMORAL TRIAL - NON POROUS SZ 3 LT 66mm M/L 61mm A/P · Class 1

All models of P.F.C. SIGMA (487)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10603295234579), received between Sep 8, 2017 and Apr 5, 2023 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device-Device Incompatibility4
  • Naturally Worn3
  • Appropriate Device Problem Term/Code Not Available2
  • Material Twisted/Bent2
  • Material Deformation1
  • Peeled/Delaminated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient4
  • No Information1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 5, 2023Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT TRIAL WAS WORN OUT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16682014 ↗
  • MalfunctionJan 17, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT BOX TRIAL SET SCREW IS STRIPPED, AND STUCK IN FEMUR TRIAL. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16186350 ↗
  • MalfunctionNov 15, 2022Device-Device Incompatibility; Naturally Worn; Device-Device Incompatibility; Naturally Worn

    IT WAS REPORTED THAT THE BOX TRIAL SCREW IS STRIPPED, CANNOT DISENGAGE FROM FEMUR TRIAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15796817 ↗
  • MalfunctionJul 13, 2021Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT THE FEMORAL TRIAL LUG HOLE HAS METAL SHARD CREATING PLASTIC INSERT DEBRIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12158769 ↗
  • MalfunctionJul 7, 2021Material Twisted/Bent; Naturally Worn; Material Twisted/Bent; Naturally Worn

    THE RECESS TO ATTACH THE TC3 BOX IS WORN SO IT NO LONGER LOCKS IN PLACE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12129910 ↗
  • MalfunctionMay 23, 2019Device-Device Incompatibility; Material Twisted/Bent; Device-Device Incompatibility; Material Twisted/Bent

    IT WAS REPORTED THAT THE SIGMA SIZE 3 LEFT PC FEMUR BOX IS CROSS THREADED AND STRIPPED INSIDE THE FEMUR AND WILL NOT COME APART. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8637557 ↗
  • MalfunctionMar 29, 2019Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    CUSTOMER´S CSSD IS COMPLAINING THE TRIALS NOT TO BE MARKED/ETCHED WITH AN ARTICLE NUMBER/REF NUMBER. SO IT IS NOT POSSIBLE TO IDENTIFY/CATEGORISE THE MEDICAL DEVICE. NO SURGERY DELAY, NO ADVERSE CONSEQUENCES FOR PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8463769 ↗
  • MalfunctionMar 13, 2019Device-Device Incompatibility; Device-Device Incompatibility

    HAD DIFFICULTLY MATING THE BOX TRIAL TO FEMORAL TRIAL.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8416241 ↗
  • MalfunctionFeb 4, 2019Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE SIZE 3 SIGMA BOX WOULD NOT UNSCREW FROM THE FEMORAL TRIAL. THE SCREW SEEMS TO BE STRIPPED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8300948 ↗
  • MalfunctionSep 8, 2017Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE FEMORAL TRIAL HAS STABILIZED BOX TRIAL ATTACHED AND WAS UNABLE TO DISASSEMBLE APPEARS SCREW HAS BEEN STRIPPED.

    Patient problems: No Information

    MAUDE report 6852891 ↗

Other TEMPLATE devices

All TEMPLATE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page