Skip to content

P.F.C. SIGMA 96-1042Recalls & reported problems

Depuy (Ireland) · P.F.C. SIGMA CRUCIATE SUBSTITUTING BOX TRIAL SZ 2 · Class 1

All models of P.F.C. SIGMA (487)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10603295234708), received between Mar 20, 2018 and Oct 17, 2023 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Material Twisted/Bent” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Twisted/Bent4
  • Material Deformation3
  • Appropriate Device Problem Term/Code Not Available2
  • Device-Device Incompatibility2
  • Naturally Worn2
  • Mechanical Jam1

Patient problems

  • No Consequences Or Impact To Patient6
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 17, 2023Device-Device Incompatibility; Mechanical Jam; Device-Device Incompatibility; Mechanical Jam

    IT WAS REPORTED THAT THE SCREW IS ROLLED AND WILL NOT PROPERLY ASSEMBLE TO MATING DEVICE. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17948578 ↗
  • MalfunctionSep 27, 2022Device-Device Incompatibility; Naturally Worn; Device-Device Incompatibility; Naturally Worn

    IT WAS REPORTED THAT THE SIGMA PS SIZE 2 STB NOTCH TRIAL IS STRIPPED. IT WILL NOT THREAD INTO THE FEMORAL COMPONENT TRIAL. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15494602 ↗
  • MalfunctionNov 5, 2019Material Deformation; Material Twisted/Bent; Material Deformation; Material Twisted/Bent

    IT WAS REPORTED THAT THE FEM BOX SCREW STRIPPED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9281010 ↗
  • MalfunctionNov 5, 2019Material Deformation; Material Twisted/Bent; Material Deformation; Material Twisted/Bent

    IT WAS REPORTED THAT THE FEM BOX SCREW STRIPPED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9281008 ↗
  • MalfunctionOct 25, 2019Material Twisted/Bent; Naturally Worn; Material Twisted/Bent; Naturally Worn

    IT WAS REPORTED THAT THE SCREW OF THE PFC*SIGMA C S(PS)BOX TRIAL SZ3 WAS STRIPPED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9239305 ↗
  • MalfunctionJul 2, 2019Material Deformation; Material Twisted/Bent; Material Deformation; Material Twisted/Bent

    IT WAS REPORTED THAT THE BOX TRIAL SCREW WAS STRIPPED DUE TO OUR TIGHTENING DURING OPERATION. TRIAL WAS NEITHER USED OR NEEDED FOR OPERATION. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8752883 ↗
  • MalfunctionApr 25, 2018Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE SCREW ON THE SIGMA SZ 5 BOX TRIAL WAS STUCK WOULD NOT LOOSEN. THERE WAS NO DELAY IN THE SURGERY AS THIS WAS IDENTIFIED BEFORE THE CASE STARTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7457671 ↗
  • MalfunctionMar 20, 2018Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE SCREW STRIPPED. DOES NOT LOCK INTO THE FEMORAL TRIAL WELL. IT WAS NOTICED UPON INSPECTION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7352634 ↗

Other TEMPLATE devices

All TEMPLATE devices →

Sources and updates