Skip to content

CORAIL L20315Recalls & reported problems

Depuy (Ireland) · CORAIL HIP SYSTEM CEMENTLESS FEMORAL STEM 12/14 AMT 135 DEGREES KHO SIZE 15 HA COATED HIGH OFFSET NO COLLAR · Class 2 · Implant

Adverse event reports
19
Deaths reported
0
Injuries reported
18
Recalls naming it
0

Summary

The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 10603295258063), received between Jan 28, 2014 and Jan 29, 2024 — classified as 18 injuries, 1 malfunction.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem7
  • Appropriate Device Problem Term/Code Not Available3
  • Insufficient Device Problem Information3
  • Corroded2
  • Fracture2
  • Loss of Osseointegration1
  • Osseointegration Problem1
  • Tear, Rip or Hole in Device Packaging1

Patient problems

  • No Code Available8
  • Pain6
  • Bone Fracture(s)4
  • Discomfort3
  • Test Result3
  • Unspecified Infection3
  • Ambulation or Postural Difficulties2
  • Injury2

Recent adverse event reports

The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 29, 2024Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT THE PACKAGING WAS DAMAGED UPON RECEIPT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18603063 ↗
  • InjuryMar 17, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO PERIPROSTHETIC FRACTURE. DOI-UNKNOWN, DOR- (B)(6) 2022, UNKNOWN HIP.

    Patient problems: Limb Fracture

    MAUDE report 13801390 ↗
  • InjuryMar 29, 2021Osseointegration Problem; Osseointegration Problem

    REVISION OF A DEPUY CORAIL STEM. RIGHT SIDE. REASON FOR REVISION WAS DUE TO A PERI-PROSTHETIC FRACTURE CAUSING THE STEM TO BECOME LOOSE. THE STEM AND HEAD WERE REMOVED, AND A RECLAIM STEM AND A 36 +1.5 CERAMIC HEAD WERE IMPLANTED. CUP AND LINER WERE NOT IMPACTED.

    Patient problems: Inadequate Osseointegration; Limb Fracture

    MAUDE report 11584680 ↗
  • InjuryMay 18, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PERIPROSTHETIC FRACTURE. REMOVED STEM. IMPLANTED REVISION STEM. DOI: (B)(6) 2014. DOR: (B)(6) 2020. RIGHT HIP.

    Patient problems: Bone Fracture(s); No Code Available

    MAUDE report 10067244 ↗
  • InjuryApr 6, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED TO ADDRESS PERIPROSTHETIC FRACTURE. DOI: UNK. DOR: (B)(6) 2020, RIGHT HIP.

    Patient problems: Bone Fracture(s); No Code Available

    MAUDE report 9929190 ↗
  • InjuryDec 6, 2019Fracture; Fracture

    THE PATIENT HAD IN 2014 A HIP REPLACEMENT SURGERY THAT TODAY NEEDS TO BE REMOVED BECAUSE THE CORAIL PROSTHESIS HAS REPORTED NECK IMPLANT RUPTURE.

    Patient problems: No Code Available

    MAUDE report 9432793 ↗
  • InjurySep 18, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ADDITIONAL INFORMATION RECEIVED INDICATING A POSTTRAUMATIC PERIPROSTHETIC FRACTURE OF THE LEFT FEMUR PER- AND SUBTROCHANTERIC.

    Patient problems: Bone Fracture(s); No Code Available

    MAUDE report 9083503 ↗
  • InjuryApr 24, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PPF AND STICKER SHEETS RECORD RECEIVED. PPF ALLEGES INFECTION. DOI: (B)(6) 2016; DOR: (B)(6) 2016; (LEFT HIP). SEE (B)(4) FOR THE FIRST REVISION.

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 8546260 ↗
  • InjuryMar 15, 2019Fracture

    SUBJECT ID: (B)(6). CLINICAL ADVERSE EVENT RECEIVED FOR PERIPROSTHETIC FEMUR FRACTURE. EVENT IS SERIOUS AND IS CONSIDERED SEVERE. EVENT IS DEFINITELY RELATED TO DEVICE AND DEFINITELY NOT RELATED TO PROCEDURE. DOI: (B)(6) 2019; DOE: (B)(6) 2019; (RIGHT HIP). TREATMENT INCLUDES ORIF.

    Patient problems: Bone Fracture(s); No Code Available

    MAUDE report 8425860 ↗
  • InjuryJan 24, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    LITIGATION ALLEGES THAT THE PATIENT EXPERIENCED SEVERE PAIN AND DISCOMFORT, HAD ABNORMALLY HIGH LEVELS OF COBALT AND CHROMIUM, SUFFERED FROM INJURY, EMOTIONAL DISTRESS AND DISFIGUREMENT.

    Patient problems: Deformity/ Disfigurement; Discomfort; Injury; No Code Available; Pain; Test Result

    MAUDE report 8273934 ↗
  • InjuryDec 28, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    LITIGATION ALLEGES PAIN AND DECREASED RANGE OF MOTION. PPF ALLEGES ABDUCTOR MUSCLE REPAIR AND ELEVATED METAL IONS. DOI: (B)(6) 2010 - DOR: NONE REPORTED (RIGHT HIP).

    Patient problems: Loss of Range of Motion; Pain; Test Result; Tissue Damage

    MAUDE report 8202344 ↗
  • InjurySep 26, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    ASR REVISION , ASR XL ACETABULAR SYSTEM , RIGHT HIP , REASON(S) FOR REVISION: PAIN, ELEVATED BLOOD ION LEVELS. DOI: (B)(6) 2009; DOR: (B)(6) 2018; RIGHT HIP.

    Patient problems: Not Applicable; Pain; Test Result

    MAUDE report 7909628 ↗
  • InjuryMay 10, 2018Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    NEW ETQ RECORD CREATED IN ORDER TO UPDATE ETQ (LEGACY SYSTEM) COMPLAINT NUMBER (B)(4). REASON FOR ORIGINAL COMPLAINT. -LITIGATION ALLEGES SEVERE PAIN, DISCOMFORT, INFLAMMATION, AND ELEVATED LEVELS OF COBALT CHROMIUM METAL IONS AND PARTICLES IN THE BLOOD AND TISSUE SURROUNDING THE IMPLANT. DOI: (B)(6) 2008 - DOR: NONE REPORTED; (RIGHT HIP). PATIENT IS A RESIDENT OF (B)(6). UPDATE: 11/2/2012- PFS WAS RECEIVED FROM LEGAL, MEDICAL RECORDS WERE RECEIVED FROM LEGAL, AND PART/LOT INFORMATION WAS IDENTIFIED. RECORDS ARE AVAILABLE FOR FURTHER REVIEW. UPDATE AD: 13 APR 2018-PFS, PPF, MEDICAL RECORD AND STICKER SHEET RECEIVED. PINNACLE PLAINTIFF FACT SHEET AND PLAINTIFF PROFILE FORM ALLEGES PAIN, INJURY AND DIFFICULTY WALKING. UPDATED COMPLAINT INFORMATION. DOI: (B)(6) 2008; DOR: NOT REVISED; RIGHT HIP.

    Patient problems: Ambulation or Postural Difficulties; Discomfort; Inflammation; Injury; No Code Available; Pain; Rash

    MAUDE report 7501991 ↗
  • InjuryJan 24, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    THE PATIENT WAS REVISED TO ADDRESS INFECTED HEMATOMA. (B)(6) 2017. (B)(6) 2017. RIGHT HIP.

    Patient problems: Hematoma; Unspecified Infection

    MAUDE report 7215076 ↗
  • InjuryJan 19, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    THE PATIENT WAS REVISED TO ADDRESS INFECTION. DOI: UNKNOWN; DOR: (B)(6) 2017; RIGHT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 7206423 ↗
  • InjurySep 11, 2017Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    PPF ALLEGES ELEVATED METAL IONS. ADDED LAWYER, LAW FIRM AND REVISION HOSPITAL. UPDATED PART AND LOT NUMBERS OF THE IMPACTED PRODUCTS. DOI: (B)(6) 2008. DOR: (B)(6) 2017. (RIGHT HIP).

    Patient problems: Discomfort; Irritation; Pain

    MAUDE report 6855834 ↗
  • InjurySep 1, 2017Corroded; Appropriate Device Problem Term/Code Not Available; Corroded; Appropriate Device Problem Term/Code Not Available

    NEW ETQ RECORD CREATED IN ORDER TO UPDATE ETQ (LEGACY SYSTEM) COMPLAINT NUMBER (B)(4). REASON FOR ORIGINAL COMPLAINT - LITIGATION PAPERS ALLEGE THE PATIENT SUFFERS FROM PAIN, METALLOSIS, THE INABILITY TO WALK AND ELEVATED METAL ION LEVELS. DOI: (B)(6) 2009 (LEFT HIP). DOR: NO DATE SET AS OF YET. RESIDENT OF (B)(6). UPDATE 9/19/12 - PLAINTIFF'S PRELIMINARY DISCLOSURE FORM WAS RECEIVED, WHICH IDENTIFIED PART/LOT INFORMATION AND CORRECT DATE OF IMPLANT. THE COMPLAINT AND ASSOCIATED MDRS WERE UPDATED. THERE WAS NO NEW INFORMATION THAT WOULD CHANGE THE OUTCOME OF THE INVESTIGATION. DOI: (B)(6) 2009. UPDATE 07 APRIL 2014 DER RECEIVED HOSPITAL (B)(4). DOR (B)(6) 2014. ADDITIONAL PRODUCT ADDED - TAPER SLEEVE. UPDATE AUG 4, 2017: MEDICAL RECORDS RECEIVED. AFTER REVIEW OF THE MEDICAL RECORDS FOR THE MDR REPORTABILITY, PATIENT WAS REVISED TO ADDRESS PAINFUL PROSTHESIS AND RISING ELEVATED METAL LEVELS. REVISION NOTES REPORTED GRAYISH BROWNISH TYPE FLUID SHOOTING OUT UNDER PRESSURE (WATER-FOUNTAIN EFFECT) ASSOCIATED WITH METAL ON METAL REACTION, AND TREMENDOUS AMOUNT OF BURNISHING AROUND THE TRUNNION. ADDED STEM TO THE COMPLAINT DUE TO ELEVATED METAL IONS. THERE WERE NO LABORATORY RESULTS PROVI

    Patient problems: Ambulation or Postural Difficulties; Foreign Body Reaction; Pain

    MAUDE report 6837844 ↗
  • InjuryJun 23, 2017Loss of Osseointegration; Loss of Osseointegration

    DER STATED THAT THE PATIENT HAD 2 DISLOCATIONS AND CAME INTO THE ER AT (B)(6) AND WAS SEEN BY A SURGEON. X-RAYS SHOWED THAT THE FEMORAL STEM HAD SPUN OUT OR CAME LOOSE IN THE FEMUR WHICH MADE THE JOINT UNSTABLE. THE PATIENT WAS REVISED ON (B)(6) 2017 WHERE THE FEMORAL STEM AND HIP BALL WAS REMOVED AND A NEW STEM WAS INSERTED WITH GOOD FIXATION AND THE JOINT WAS STABLE. SURGEON STATED THE STEM HAD CAME LOOSE AND THE HIP DISLOCATED AS A RESULT.

    Patient problems: No Information

    MAUDE report 6664935 ↗
  • InjuryJan 28, 2014Corroded

    PATIENT WAS REVISED TO ADDRESS TROCHANTERIC BURSA COLLECTION. IT WAS ALSO REPORTED THAT CORROSION WAS FOUND ON THE NECK OF THE FEMORAL STEM.

    Patient problems: Host-Tissue Reaction; No Information

    MAUDE report 3594913 ↗

Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices

All Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.