Skip to content

CORAIL L98014Recalls & reported problems

Depuy (Ireland) · CORAIL HIP SYSTEM CEMENTLESS FEMORAL STEM HA COATED REVISION 12/14 AMT 135 DEGREES STANDARD COLLAR SIZE 14 · Class 2 · Implant

All models of CORAIL (121)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10603295262336), received between Jan 5, 2018 and Dec 8, 2025 — classified as 4 injuries.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Fracture1
  • Loss of Osseointegration1
  • Osseointegration Problem1

Patient problems

  • Inadequate Osseointegration2
  • Insufficient Information1
  • No Code Available1
  • Pain1
  • Unspecified Infection1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 8, 2025Fracture; Osseointegration Problem; Fracture; Osseointegration Problem

    IT WAS REPORTED THAT THE CORAIL REVISION STEM BROKE AT MIDPOINT. DOI: (B)(6) 2023. DOR: (B)(6) 2025. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Inadequate Osseointegration; Insufficient Information

    MAUDE report 23743036 ↗
  • InjuryApr 26, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO INFECTION. RESECTION ARTHROPLASTY AND GIRDLE STONE. DOI: (B)(6) 2022. DOR: (B)(6) 2022. AFFECTED SIDE: LEFT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 14213366 ↗
  • InjuryJan 23, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED TO ADDRESS PAIN. DOI: (B)(6) 2019; DOR: (B)(6) 2020; LEFT HIP.

    Patient problems: No Code Available; Pain

    MAUDE report 9621796 ↗
  • InjuryJan 5, 2018Loss of Osseointegration; Loss of Osseointegration

    PATIENT WAS REVISED TO ADDRESS PATHOLOGICAL FEMUR FRACTURE AND LOOSENING OF THE STEM AT THE BONE TO IMPLANT INTERFACE. FEMORAL COMPONENT OF TOTAL HIP SUBSIDED SIGNIFICANTLY. SURGEON BELIEVED THIS WAS A RESULT OF THE COMPRISE OF THE BONE FROM CANCER AND THE RADIATION AND CHEMOTHERAPY TO TREAT IT. DOI: (B)(6) 2017; DOR: (B)(6) 2017; LEFT HIP.

    Patient problems: Inadequate Osseointegration

    MAUDE report 7167592 ↗

Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices

All Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page