NA L94010Recalls & reported problems
Depuy (Ireland) · LARGE SHIELDED CALCAR PLANER · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 22
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 22 adverse event reports that list this exact device (DI 10603295325192), received between May 15, 2018 and Jul 28, 2025 — classified as 22 malfunctions.
The most frequently reported specific device problem is “Failure to Cut” (15 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Cut15
- Dull, Blunt6
- Naturally Worn6
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions16
- No Consequences Or Impact To Patient5
- Insufficient Information2
- No Code Available1
Recent adverse event reports
The 22 most recent of 22 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 28, 2025Break; Break
IT WAS REPORTED THAT THE METAL FILINGS COMING OFF INSTRUMENT AFTER STERILIZATION. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22631422 ↗MalfunctionJun 2, 2023Failure to Cut; Failure to Cut
CALCAR MILLS ARE BLUNT, SURGEONS REQUESTING NEW ONES TO REAM CALCAR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17049367 ↗MalfunctionJun 2, 2023Failure to Cut; Failure to Cut
CALCAR MILLS ARE BLUNT, SURGEONS REQUESTING NEW ONES TO REAM CALCAR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17049343 ↗MalfunctionJun 2, 2023Failure to Cut; Failure to Cut
CALCAR MILLS ARE BLUNT, SURGEONS REQUESTING NEW ONES TO REAM CALCAR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17049341 ↗MalfunctionJun 2, 2023Failure to Cut; Failure to Cut
CALCAR MILLS ARE BLUNT, SURGEONS REQUESTING NEW ONES TO REAM CALCAR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17049340 ↗MalfunctionJun 2, 2023Failure to Cut; Failure to Cut
CALCAR MILLS ARE BLUNT, SURGEONS REQUESTING NEW ONES TO REAM CALCAR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17049339 ↗MalfunctionNov 30, 2022Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE CORAIL REAMERS ARE BLUNT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15886705 ↗MalfunctionNov 18, 2022Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE REAMER IS DULL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15827217 ↗MalfunctionJul 7, 2022Failure to Cut; Naturally Worn; Failure to Cut; Naturally Worn
IT WAS REPORTED THAT THE INSTRUMENTS ARE WORN. THERE HAS NOT BEEN ANY IMPACT TO PATIENTS AND THERE WAS A SURGICAL DELAY OF 5 MINUTES WHILE A NEW TRAY WAS OPENED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 14945066 ↗MalfunctionJul 7, 2022Failure to Cut; Naturally Worn; Failure to Cut; Naturally Worn
IT WAS REPORTED THAT THE INSTRUMENTS ARE WORN. THERE HAS NOT BEEN ANY IMPACT TO PATIENTS AND THERE WAS A SURGICAL DELAY OF 5 MINUTES WHILE A NEW TRAY WAS OPENED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 14945038 ↗MalfunctionDec 21, 2021Failure to Cut; Naturally Worn; Failure to Cut; Naturally Worn
SCREW DRIVER DAMAGED, NOT GRIPPING, EDGES WITH VISIBLE SIGNS OF WEAR CALCAR REAMERS NOT CUTTING, VISIBLE SIGNS OF WEAR WAS SURGERY DELAYED DUE TO THE REPORTED EVENT? YES, IF YES, NUMBER OF MINUTES: 10. ACTION TAKEN WHEN EVENT OCCURRED? ANOTHER STEP OF INSTRUMENTS OPENED. WAS PROCEDURE SUCCESSFULLY COMPLETED? YES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13054361 ↗MalfunctionDec 21, 2021Failure to Cut; Naturally Worn; Failure to Cut; Naturally Worn
SCREW DRIVER DAMAGED, NOT GRIPPING, EDGES WITH VISIBLE SIGNS OF WEAR. CALCAR REAMERS NOT CUTTING, VISIBLE SIGNS OF WEAR. WAS SURGERY DELAYED DUE TO THE REPORTED EVENT? YES. IF YES, NUMBER OF MINUTES: 10. ACTION TAKEN WHEN EVENT OCCURRED? ANOTHER STEP OF INSTRUMENTS OPENED. WAS PROCEDURE SUCCESSFULLY COMPLETED? YES. WERE FRAGMENTS GENERATED? NO.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13053901 ↗MalfunctionDec 21, 2021Failure to Cut; Naturally Worn; Failure to Cut; Naturally Worn
ADDITIONAL INFORMATION RECEIVED INDICATED THAT THE INSTRUMENTS DID NOT BREAK INTO TWO OR MORE PIECES AND THERE WAS NO SURGICAL DELAY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13053113 ↗MalfunctionOct 4, 2021Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE CALCAR REAMERS ARE BLUNT. THERE WAS NOT A DELAY IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12569089 ↗MalfunctionSep 24, 2021Failure to Cut; Naturally Worn; Failure to Cut; Naturally Worn
WE HAVE 2 CALCAR REAMERS THAT ARE BLUNT THROUGH WEAR AND TEAR AT BEDFORD HOSPITAL. THEY WERE USED DURING A PROCEDURE. THERE WAS NO DELAY TO SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12525832 ↗MalfunctionJul 13, 2021Failure to Cut; Failure to Cut
THR CASE IT WAS REPORTED THAT THE CALCAR REAMER WAS BLUNT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12158041 ↗MalfunctionJul 8, 2020Dull, Blunt; Dull, Blunt
TOTAL HIP REPLACEMENT; CALCAR REAMER BLUNT AND UNABLE TO BE SHARPENED. NO AE TO PATIENT.
Patient problems: No Code Available
MAUDE report 10247470 ↗MalfunctionAug 13, 2018Dull, Blunt; Dull, Blunt
THR: (B)(6) HOSPITAL, (B)(6) 2018. BOTH OF THE CALCAR REAMERS WERE REPORTED BLUNT BY THE SURGEON DURING THE CASE. THERE WAS NO ADVERSE EVENT TO PATIENT OR DELAY IN SURGERY, AS WERE ABLE TO OPEN ANOTHER TRAY IMMEDIATELY AND USE A SHARPER REAMER FIT FOR PURPOSE. (B)(6).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7776776 ↗MalfunctionMay 18, 2018Dull, Blunt; Dull, Blunt
(B)(6) HOSPITAL; (B)(6) 2018. CALCAR ITEMS HAVE BEEN NOTED BY SURGEONS AND (B)(6) STAFF THAT THEY ARE BLUNT AND NOT MEETING STANDARDS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7525237 ↗MalfunctionMay 15, 2018Dull, Blunt; Dull, Blunt
(B)(6) 2018. CALCAR ITEMS HAVE BEEN NOTED BY SURGEONS AND CSSD STAFF THAT THEY ARE BLUNT AND NOT MEETING STANDARDS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7513811 ↗MalfunctionMay 15, 2018Dull, Blunt; Dull, Blunt
(B)(6) HOSPITAL, (B)(6) 2018. CALCAR ITEMS HAVE BEEN NOTED BY SURGEONS AND CSSD STAFF THAT THEY ARE BLUNT AND NOT MEETING STANDARDS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7513809 ↗MalfunctionMay 15, 2018Dull, Blunt; Dull, Blunt
(B)(6) 2018. CALCAR ITEMS HAVE BEEN NOTED BY SURGEONS AND CSSD STAFF THAT THEY ARE BLUNT AND NOT MEETING STANDARDS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7513795 ↗
Other Bone mill devices
- AML EXCEL 2001-48-000Depuy (Ireland) · Bone mill472 reports
- AML EXCEL 2001-47-000Depuy (Ireland) · Bone mill220 reports
- AML EXCEL 2001-49-000Depuy (Ireland) · Bone mill212 reports
- SUMMIT 2570-04-100Depuy (Ireland) · Bone mill193 reports
- SUMMIT 2570-04-200Depuy (Ireland) · Bone mill91 reports
- LPS 2987-70-001Depuy (Ireland) · Bone mill5 reports
- NA 4861-00-010Depuy (Ireland) · Bone mill4 reports
- NA 2001-80-503Depuy Orthopaedics, Inc. · Bone mill3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page