Skip to content

ACTIS 2010-01-070Recalls & reported problems

Depuy (Ireland) · ACTIS BROACH SIZE 7 · Class 1

Quick summary

This DePuy ACTIS broach is a surgical instrument used to shape bone during hip replacement surgery. All 161 reports are malfunctions, and many describe broaches that break, do not fit with other instruments, or fail to cut. The reports include no deaths or injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
162
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 162 adverse event reports that list this exact device (DI 10603295393696), received between May 4, 2018 and Sep 10, 2026 — classified as 162 malfunctions.

The most frequently reported specific device problem is “Break” (63 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 3 reportsJan 2024: 4 reportsFeb 2024: 5 reportsMar 2024: 2 reportsApr 2024: 2 reportsMay 2024: 1 reportsJun 2024: 3 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 3 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 2 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 4 reportsJun 2025: 1 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 3 reportsFeb 2026: 1 reportsMar 2026: 11 reportsApr 2026: 3 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 2 reportsOct 2026: 0 reports20242025202611
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break63
  • Device-Device Incompatibility58
  • Scratched Material20
  • Material Deformation17
  • Failure to Cut15
  • Material Twisted/Bent12
  • Difficult to Advance11
  • Naturally Worn4

Patient problems

  • No Clinical Signs, Symptoms or Conditions111
  • No Consequences Or Impact To Patient30
  • Insufficient Information21
  • No Code Available3
  • Not Applicable2

Recent adverse event reports

The 25 most recent of 162 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 10, 2026Material Deformation; Scratched Material

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50721105 ↗
  • MalfunctionSep 1, 2026Break; Device-Device Incompatibility

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50545694 ↗
  • MalfunctionAug 21, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50379538 ↗
  • MalfunctionAug 21, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50378926 ↗
  • MalfunctionApr 28, 2026Material Deformation; Scratched Material

    IT WAS REPORTED DURING MDRD INSPECTION SEVERAL ACTIS BROACHES WERE FOUND TO BE DAMAGED. THE BROACHES HAD LARGE DENTS AND CURLING OF THE METAL ON THE COMPACTION TEETH. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25014778 ↗
  • MalfunctionApr 13, 2026Break

    IT WAS REPORTED THAT THE DURING BROACHING WITH KINCISE THE MALE ATTACHMENT ON THE BROACH SNAPPED OFF THE BROACH. ALL PARTS WERE RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24858065 ↗
  • MalfunctionApr 7, 2026Break; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT ON UPON OFFICE IN HOUSE INSPECTION, IT WAS FOUND THAT THE INSTRUMENT WAS BROKEN. NO PREVIOUS CASES WAS AFFECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24811626 ↗
  • MalfunctionMar 31, 2026Break

    IT WAS REPORTED THE OLD STYLE ACTIS BROACHES WERE BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24740998 ↗
  • MalfunctionMar 31, 2026Break

    IT WAS REPORTED THE OLD STYLE ACTIS BROACHES WERE BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24740955 ↗
  • MalfunctionMar 31, 2026Break

    IT WAS REPORTED THE OLD STYLE ACTIS BROACHES WERE BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24740901 ↗
  • MalfunctionMar 13, 2026Break

    IT WAS REPORTED THE BROACH IS UNABLE TO EXTRACT DUE TO A BROKEN TRUNNION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24594744 ↗
  • MalfunctionMar 13, 2026Break

    IT WAS REPORTED THE BROACH IS UNABLE TO EXTRACT DUE TO A BROKEN TRUNNION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24591926 ↗
  • MalfunctionMar 13, 2026Break

    IT WAS REPORTED THE BROACH IS UNABLE TO EXTRACT DUE TO A BROKEN TRUNNION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24591885 ↗
  • MalfunctionMar 13, 2026Break

    IT WAS REPORTED THE BROACH IS UNABLE TO EXTRACT DUE TO A BROKEN TRUNNION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24591859 ↗
  • MalfunctionMar 13, 2026Break

    IT WAS REPORTED THE BROACH IS UNABLE TO EXTRACT DUE TO A BROKEN TRUNNION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24591821 ↗
  • MalfunctionMar 13, 2026Break

    IT WAS REPORTED THAT BROACHES ARE BROKEN INTO TWO PIECES. NO ADDITIONAL FRAGMENTS. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24591474 ↗
  • MalfunctionMar 12, 2026Break

    IT WAS REPORTED THAT THIS FACILITY HAD MULTIPLE VERSION 1 AND 2 BROACHES BREAK AND WERE REQUESTING ALL OF THEIR CONSIGNED BROACHES BE SWAPPED OUT FOR VERSION 3 TO AVOID FURTHER BREAKAGES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24579416 ↗
  • MalfunctionMar 12, 2026Break

    IT WAS REPORTED THAT THIS FACILITY HAD MULTIPLE VERSION 1 AND 2 BROACHES BREAK, AND WERE REQUESTING ALL OF THEIR CONSIGNED BROACHES BE SWAPPED OUT FOR VERSION 3 TO AVOID FURTHER BREAKAGES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24578931 ↗
  • MalfunctionFeb 10, 2026Break

    IT WAS REPORTED THAT THERE WAS BROKEN BROACHES/T HANDLE. THE EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24305550 ↗
  • MalfunctionJan 14, 2026Break

    IT WAS REPORTED THAT THE SIZE 6 BROACH TRUNION SNAPPED DURING EXTRACTION. ALL PARTS WERE RETRIEVED. ALL BROACHES WERE SWAPPED FOR SLOTTED BROACHES FOR EASE OF EXTRACTION AS PER SURGEON REQUEST. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24068753 ↗
  • MalfunctionJan 14, 2026Break

    IT WAS REPORTED THAT THE SIZE 6 BROACH TRUNION SNAPPED DURING EXTRACTION. ALL PARTS WERE RETRIEVED. ALL BROACHES WERE SWAPPED FOR SLOTTED BROACHES FOR EASE OF EXTRACTION AS PER SURGEON REQUEST. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24068556 ↗
  • MalfunctionJan 14, 2026Break

    IT WAS REPORTED THAT THE SIZE 6 BROACH TRUNION SNAPPED DURING EXTRACTION. ALL PARTS WERE RETRIEVED. ALL BROACHES WERE SWAPPED FOR SLOTTED BROACHES FOR EASE OF EXTRACTION AS PER SURGEON REQUEST. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24068475 ↗
  • MalfunctionNov 11, 2025Break; Break

    IT WAS REPORTED THAT THE TOP OF THE BROACH, WHICH CONNECTS TO THE NECK TRIAL AND BROACH HANDLE, BROKE INTO TWO PIECES. ALL FRAGMENTS WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23524550 ↗
  • MalfunctionOct 15, 2025Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    IT WAS REPORTED THAT ACTIS BROACHES NEEDED TO BE REPLACED. THERE WERE SEVERE STRESS RINGS ON THEM WHERE THEY ATTACH TO THE BROACH HANDLE. SURGEON FREQUENTLY BREAKS THEM. WHEN THEY BREAK, IT IS EXTREMELY HARD TO REMOVE FROM THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23279716 ↗
  • MalfunctionOct 10, 2025Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    IT WAS REPORTED THAT THERE ARE SEVERE STRESS RINGS ON ACTIS BROACHES WHERE THEY ATTACH TO BROACH HANDLE. SURGEON FREQUENTLY BREAKS THEM AND WHEN THEY BREAK IT IS EXTREMELY HARD TO REMOVE FROM THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23272737 ↗

Other BROACH devices

All BROACH devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.