Skip to content

ACTIS 2010-02-100Recalls & reported problems

Depuy (Ireland) · ACTIS STANDARD NECK SIZE 0/1 · Class 1

All models of ACTIS (97)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
15
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 10603295393740), received between Aug 2, 2018 and Sep 9, 2026 — classified as 15 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (13 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility13
  • Break5
  • Loose or Intermittent Connection2
  • Material Twisted/Bent2
  • Appropriate Device Problem Term/Code Not Available1

Patient problems

  • No Consequences Or Impact To Patient10
  • Not Applicable3
  • No Clinical Signs, Symptoms or Conditions2
  • No Code Available2

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 9, 2026Break; Device-Device Incompatibility

    It was reported that the Broaches were broke during a case. Surgeon needed to order all new sets of broaches with the new Design in case this happens again.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50689328 ↗
  • MalfunctionMay 9, 2023Break; Break

    IT WAS REPORTED THAT THE INSTRUMENT WAS BROKEN DURING THE CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16898424 ↗
  • MalfunctionSep 30, 2020Loose or Intermittent Connection; Device-Device Incompatibility; Material Twisted/Bent; Loose or Intermittent Connection; Device-Device Incompatibility; Material Twisted/Bent

    IT WAS REPORTED THAT THE ACTIS TRIAL NECK SEGMENTS LOOSE TOLERANCE AND SLIGHTLY DISENGAGING WHEN TRIALING HIP. NO PROBLEMS WERE CAUSED DUE TO WORN INSTRUMENT. NEW NECK SEGMENTS WERE SWAPPED TO ADDRESS THIS ISSUES. NO SURGICAL DELAY.

    Patient problems: No Code Available; No Consequences Or Impact To Patient

    MAUDE report 10606196 ↗
  • MalfunctionSep 30, 2020Loose or Intermittent Connection; Device-Device Incompatibility; Material Twisted/Bent; Loose or Intermittent Connection; Device-Device Incompatibility; Material Twisted/Bent

    IT WAS REPORTED THAT THE ACTIS TRIAL NECK SEGMENTS LOOSE TOLERANCE, AND SLIGHTLY DISENGAGING WHEN TRIALING HIP. NO PROBLEMS WERE CAUSED DUE TO WORN INSTRUMENT. NEW NECK SEGMENTS WERE SWAPPED TO ADDRESS THIS ISSUES. NO SURGICAL DELAY.

    Patient problems: No Code Available; No Consequences Or Impact To Patient

    MAUDE report 10606059 ↗
  • MalfunctionOct 22, 2019Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE ACTIS TRIAL NECKS KEEP SLIDING OFF BROACHES . THEY CALLED R.O IN HIP MARKETING AND TOLD THEY HAVE FIXED THIS ISSUE IN THE NEWER SETS. NEEDING TO GET THREE SETS OF TRIAL NECKS FOR ACTIS STANDARD AND HIGH OFFSET . R.O TOLD HIM TO FILL OUT MDR FORM AND I CLD GET OLDER STYLE ACTIS NECKS REPLACED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9217815 ↗
  • MalfunctionOct 22, 2019Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE ACTIS TRIAL NECKS KEEP SLIDING OFF BROACHES. THEY CALLED R.O IN HIP MARKETING AND TOLD THEY HAVE FIXED THIS ISSUE IN THE NEWER SETS. NEEDING TO GET THREE SETS OF TRIAL NECKS FOR ACTIS STANDARD AND HIGH OFFSET . R.O TOLD HIM TO FILL OUT MDR FORM AND I CLD GET OLDER STYLE ACTIS NECKS REPLACED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9217813 ↗
  • MalfunctionOct 22, 2019Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE ACTIS TRIAL NECKS KEEP SLIDING OFF BROACHES . THEY CALLED R.O IN HIP MARKETING AND TOLD THEY HAVE FIXED THIS ISSUE IN THE NEWER SETS. NEEDING TO GET THREE SETS OF TRIAL NECKS FOR ACTIS STANDARD AND HIGH OFFSET . R.O TOLD HIM TO FILL OUT MDR FORM AND I CLD GET OLDER STYLE ACTIS NECKS REPLACED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9217812 ↗
  • MalfunctionMay 22, 2019Device-Device Incompatibility; Device-Device Incompatibility

    NECK SEGMENT IS LOOSE ON BROACH. PATIENT STATUS/ OUTCOME / CONSEQUENCES --> NO.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8633956 ↗
  • MalfunctionMay 22, 2019Device-Device Incompatibility; Device-Device Incompatibility

    NECK SEGMENT IS LOOSE ON BROACH. PATIENT STATUS/ OUTCOME / CONSEQUENCES --> NO.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8633858 ↗
  • MalfunctionMay 22, 2019Device-Device Incompatibility; Device-Device Incompatibility

    NECK SEGMENT IS LOOSE ON BROACH.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8631930 ↗
  • MalfunctionMay 21, 2019Device-Device Incompatibility; Device-Device Incompatibility

    NECK SEGMENT IS LOOSE ON BROACH.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8629110 ↗
  • MalfunctionMay 8, 2019Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    IT WAS REPORTED THAT THE ACTIS NECK TRIALS HAVE BEEN COMING APART DURING TRIAL REDUCTIONS. FORCING THE SURGEON TO DISLOCATE HIP, DROP THE LEG, RECONNECT THE NECK TRIAL, AND REDUCE THE HIP AGAIN. THIS HAS BEEN AN ONGOING ISSUE WITH OUR SETS, AND HAS ADDED TIME TO PROCEDURES. WE HAVE 3 SETS IN OUR HOSPITAL THAT NEED TO BE CHANGED OUT.

    Patient problems: Not Applicable

    MAUDE report 8591478 ↗
  • MalfunctionMay 8, 2019Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    IT WAS REPORTED THAT THE ACTIS NECK TRIALS HAVE BEEN COMING APART DURING TRIAL REDUCTIONS. FORCING THE SURGEON TO DISLOCATE HIP, DROP THE LEG, RECONNECT THE NECK TRIAL, AND REDUCE THE HIP AGAIN. THIS HAS BEEN AN ONGOING ISSUE WITH OUR SETS, AND HAS ADDED TIME TO PROCEDURES. WE HAVE 3 SETS IN OUR HOSPITAL THAT NEED TO BE CHANGED OUT.

    Patient problems: Not Applicable

    MAUDE report 8591388 ↗
  • MalfunctionMay 8, 2019Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    IT WAS REPORTED THAT THE ACTIS NECK TRIALS HAVE BEEN COMING APART DURING TRIAL REDUCTIONS. FORCING THE SURGEON TO DISLOCATE HIP, DROP THE LEG, RECONNECT THE NECK TRIAL, AND REDUCE THE HIP AGAIN. THIS HAS BEEN AN ONGOING ISSUE WITH OUR SETS, AND HAS ADDED TIME TO PROCEDURES. WE HAVE 3 SETS IN OUR HOSPITAL THAT NEED TO BE CHANGED OUT.

    Patient problems: Not Applicable

    MAUDE report 8591167 ↗
  • MalfunctionAug 2, 2018Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    USER REPORTED THAT FOR THE ACTIS TRIAL NECKS "IT WOULD BE AN IMPROVEMENT IF THE ¿FEMALE" TRIAL NECK HAD AN O RING IN THE HOLE SO THAT THE NECK WOULD NOT COME OFF DURING TRIAL REDUCTION AND DISLOCATION." COMPLAINT BEING RAISED AS USE INDICATED THAT THE NECK WOULD COME OFF DURING TRIALING AND DISLOCATE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7742507 ↗

Other TEMPLATE devices

All TEMPLATE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page