This DePuy Attune patella clamp ring is a surgical instrument used during knee replacement to hold the kneecap in place. Its 155 reports mostly describe rings breaking or cracking, parts wearing out, and the clamp not holding. The reports include no deaths and 2 injuries, and 153 are malfunctions. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
153
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 153 adverse event reports that list this exact device (DI 10603295395454), received between Jul 21, 2017 and Aug 3, 2026 — classified as 2 injuries, 151 malfunctions.

The most frequently reported specific device problem is “Break” (75 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 3 reportsApr 2024: 0 reportsMay 2024: 3 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 8 reportsSep 2024: 4 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 6 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 3 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break75
  • Crack37
  • Naturally Worn22
  • Device-Device Incompatibility14
  • Material Twisted/Bent9
  • Detachment of Device or Device Component8
  • Scratched Material7
  • Appropriate Device Problem Term/Code Not Available6

Patient problems

  • No Clinical Signs, Symptoms or Conditions89
  • No Consequences Or Impact To Patient44
  • Insufficient Information12
  • No Information7
  • No Code Available4
  • Device Embedded In Tissue or Plaque1
  • Foreign Body In Patient1

Recent adverse event reports

The 25 most recent of 153 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 3, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50114359 ↗
  • MalfunctionAug 3, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50114357 ↗
  • MalfunctionJun 5, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT THE RUBBER RING OF PATELLA CLAMP FELL APART AFTER WASHING. THERE WERE NO ISSUES OR COMPLICATIONS. IT WAS REMOVED FROM SURGERY CENTER AFTER FALLING APART.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25411292 ↗
  • MalfunctionMay 8, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT CLAMP WILL NOT HOLD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25129956 ↗
  • MalfunctionApr 22, 2026Break

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE FOR THE HIP IT WAS OBSERVED THAT THE PINNACLE CUP IMPACTOR HANDLE BROKE WHILE IMPLANTING THE CUP. IT WAS REPORTED THAT THERE WAS NO DELAY IN THE PROCEDURE DUE TO THE EVENT, AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH THE EXISTING INSTRUMENT. THERE WERE NO FRAGMENTS GENERATED. THERE WERE NO PATIENT CONSEQUENCES. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24955047 ↗
  • MalfunctionMar 2, 2026Crack

    IT WAS REPORTED THAT THE PATELLLA COMPRESSION RING HAS CRACKED IN THE POLY ORANGE RING PRONE TO INFECTION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24489492 ↗
  • MalfunctionJan 13, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT PATELLA DRILL TRIALS WILL NOT STAY CONNECTED TO PATELLA DRILL CLAMP WHILE SIZING AND PREPPING THE PATELLA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24053551 ↗
  • MalfunctionNov 18, 2025Break; Break

    IT WAS REPORTED THAT THE PATELLA RING BECAME SEVERELY CHIPPED DURING NORMAL STERILIZATION PROCESS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23587136 ↗
  • MalfunctionNov 10, 2025Break; Break

    IT WAS REPORTED THAT THE ORANGE RING WAS BROKEN IN HALF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23513772 ↗
  • MalfunctionOct 23, 2025Crack; Crack

    IT WAS REPORTED THAT THE ORANGE RUBBER RING HAS CRACKS IN IT. REPLACEMENT IS NEEDED. THE EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23368407 ↗
  • MalfunctionOct 21, 2025Break; Break

    IT WAS REPORTED THAT THE RING WAS BROKEN. THE EVENT DID NOT HAPPENED DURING A SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23351046 ↗
  • MalfunctionJul 10, 2025Break; Break

    IT WAS REPORTED THE RUBBER PIECE OF THE PATELLA CLAMP RING BROKE OFF. THERE WAS NO REPORTED PATIENT OR PROCEDURE INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22469911 ↗
  • MalfunctionJun 20, 2025Break; Break

    IT WAS REPORTED THAT THE DEVICE IS BROKEN.

    Patient problems: Insufficient Information

    MAUDE report 22301757 ↗
  • MalfunctionApr 30, 2025Crack; Crack

    IT WAS REPORTED THAT CSSD DISCOVERED HOLES IN TWO RED SILICONE DISCS. IT WAS ALSO DISCOVERED THAT AND RP IMPACTOR HAD WEAR AND CRACKS. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21937456 ↗
  • MalfunctionApr 30, 2025Crack; Crack

    IT WAS REPORTED THAT CSSD DISCOVERED HOLES IN TWO RED SILICONE DISCS. IT WAS ALSO DISCOVERED THAT AND RP IMPACTOR HAD WEAR AND CRACKS. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21937447 ↗
  • MalfunctionApr 22, 2025Corroded; Corroded

    IT WAS REPORTED THAT THE INSTRUMENT WERE PITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21880965 ↗
  • MalfunctionFeb 20, 2025Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT THE RUBBER OF A PATELLA COMPRESSION RING HAS DEGRADED. THE EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21424084 ↗
  • MalfunctionJan 31, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE ITEMS ARE NO LONGER ATTACHING TOGETHER, FOLLOWING STERILIZATION.THERE WAS NO HARM TO PATIENT AND NO DELAY TO SURGERY OCCURRED DUE TO EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21288735 ↗
  • MalfunctionJan 27, 2025Break; Break

    IT WAS REPORTED THAT THE RUBBER RING ON THE PATELLA CLAMP RING HAS A LITTLE CHIP MISSING. THIS WAS NOTICED WHILE SETTING UP THE ROOM. THERE WERE NO DELAYS, AND THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21243364 ↗
  • MalfunctionJan 27, 2025Break; Break

    IT WAS REPORTED THAT THE INSTRUMENT HAD THE ISSUE OF WEAR AND TEAR. RUBBER HAS RIPPED APART FROM THE INSTRUMENT. IT WAS UNKNOWN IF THERE WAS A SURGICAL DELAY.

    Patient problems: Insufficient Information

    MAUDE report 21241229 ↗
  • MalfunctionJan 6, 2025Break; Break

    IT WAS REPORTED THAT THE HOSPITAL HAD FOUR TRAYS THAT WERE OUT OF ACTION WITH BROKEN INSTRUMENTS. THE ATTUNE PATELLA RING WAS TORN IN TWO PIECES. THE CEMENT RESTRICTOR INSERTER HAD FADED NUMBERS ON THE HANDLE. THE PINNACLE CUP IMPACTOR ROD THAT CONNECTS THE HANDLE WITH THE TIGHTENING KNOB WAS SPLIT IN TWO PIECES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21078518 ↗
  • MalfunctionJan 3, 2025Material Deformation; Material Deformation

    IT WAS REPORTED ON (B)(6) 2024, IT FOUND DURING STERILIZATION INSPECTION THAT THE PATELLA CLAMP RUBBER RING IS TORN AND NEEDS REPLACING. THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21071576 ↗
  • MalfunctionJan 2, 2025Break; Material Deformation; Break; Material Deformation

    IT WAS REPORTED THAT THE DEVICE WAS REPORTED FOR WEAR AND TEAR. NO OTHER INFORMATION WAS PROVIDED AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21057415 ↗
  • MalfunctionDec 13, 2024Crack; Detachment of Device or Device Component; Crack; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING LOAN KIT INSPECTION IT WAS IDENTIFIED THAT THE INSTRUMENT, COMPOSED OF TWO PARTS, THE RUBBER SEAL HAS CRACKED AND IS FALLING APART.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20925066 ↗
  • MalfunctionOct 23, 2024Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation

    SURGEON PLACED SCREW ONTO POWER TO INSERT INTO FEMORAL COMPONENT. WHEN ATTEMPTING TO PLACE SCREW INTO COMPONENT IT WAS CLEAR THERE WAS A BEND IN THE SCREW PREVENTING A STRAIGHT ENTRY POINT. NO HARM TO THE PATIENT. CLAMP WAS DAMAGED POST CASE. NO DELAY TO SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20513467 ↗