Skip to content

ATTUNE 2545-01-965Recalls & reported problems

Depuy (Ireland) · ATTUNE FB CR ARTICULATION SURFACE SIZE 5 · Class 1

All models of ATTUNE (1475)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10603295423379), received between Dec 6, 2017 and Apr 29, 2025 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Break” (5 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 12, 2015Terminated

    Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal is separated from the post, it has the potential to enter the surgical site and be left in the patient if the surgeon is unaware of the disassociation.

    Z-2146-2015 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break5
  • Material Deformation2
  • Component Missing1
  • Detachment of Device or Device Component1
  • Scratched Material1

Patient problems

  • Insufficient Information2
  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient2
  • Not Applicable1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 29, 2025Break; Break

    IT WAS REPORTED THAT TWO SIZE 5 FIXED BEARING ARTICULAR SURFACE TRAILS WERE BROKEN ON AN UNKNOWN DATE. IT WAS UNKNOWN IF THERE WAS ANY DELAY IN SURGERY.

    Patient problems: Insufficient Information

    MAUDE report 21925367 ↗
  • MalfunctionApr 29, 2025Break; Break

    IT WAS REPORTED THAT TWO SIZE 5 FIXED BEARING ARTICULAR SURFACE TRAILS WERE BROKEN ON AN UNKNOWN DATE. IT WAS UNKNOWN IF THERE WAS ANY DELAY IN SURGERY.

    Patient problems: Insufficient Information

    MAUDE report 21925328 ↗
  • MalfunctionMar 5, 2024Detachment of Device or Device Component; Material Deformation; Detachment of Device or Device Component; Material Deformation

    EACH INSTRUMENT IS COMPOSED OF 1 PART. THE SPRINGS OF EACH INSTRUMENT ARE TANGLED OR NON-UNIFORM. THE PRODUCT DAMAGE DOCUMENTED IN THIS PC HAS BEEN IDENTIFIED DURING LOAN KIT INSPECTION, BY THE LOAN KIT TECHNICIAN. THE EVENT DATE AND ALERT DATE ARE THE DATE THAT THE INSPECTION TOOK PLACE. THERE IS NO SURGEON, PROCEDURE, OR PATIENT DETAILS AVAILABLE. NO FURTHER INFORMATION CAN BE OBTAINED AS THE CASE WAS NOT REPORTED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18841432 ↗
  • MalfunctionFeb 28, 2022Break; Break

    IT WAS REPORTED THAT THE INSTRUMENTS BROKE. ALL PIECES RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13619927 ↗
  • MalfunctionJan 2, 2020Break; Material Deformation; Scratched Material; Break; Material Deformation; Scratched Material

    IT WAS REPORTED THAT ON (B)(6) 2019, THE SPRING OF THE INSERT TRIAL (P/N: 254501965 AND 254500652) WAS BROKEN VIA THE TKA FOR RIGHT OA. AFTER FEMORAL AND TIBIAL IMPLANTS WERE INSERTED, THE SURGEON FIXED WITH THE INSERT TRIAL TO APPLY PRESSURE IN EXTENSION POSITION. WHEN THE SURGEON TRIED TO REMOVE THE INSERT TRIAL TO INSERT AN INSERT (NOT TRIAL), HE COULD NOT REMOVE THE INSERT TRIAL FROM A TIBIAL TRAY. THE SURGEON MANAGED TO REMOVE THE INSERT TRIAL WITH A PLIERS, AND HE FOUND THAT THE SPRING OF THE INSERT TRIAL HAD BEEN BROKEN. BECAUSE THE SURGEON CONFIRMED THAT THERE WAS NO FRAGMENT IN THE PATIENT BODY, THE SURGERY WAS CONTINUED, AND IT WAS COMPLETED SUCCESSFULLY. THE SURGERY WAS DELAYED BY LESS THAN 30 MINUTES. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Not Applicable

    MAUDE report 9540450 ↗
  • MalfunctionDec 17, 2019Break; Break

    IT WAS REPORTED THAT THE INSERT BROKE DURING THE CASE. ALL PARTS WERE RETRIEVED AND CASE WASN'T DELAYED. BROKEN ITEM BEING SHIPPED BACK TO OFFICE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9480373 ↗
  • MalfunctionDec 6, 2017Component Missing; Component Missing

    IT WAS REPORTED THAT WHILE SETTING UP THE CASE, THE ATTUNE SPACER BLOCK AND INSERT TRIAL WERE NOTED TO HAVE BALSEALS MISSING. THE INSTRUMENTS WERE REMOVED FROM THE CASE AND REPLACED WITH AN INTACT SPACER BLOCK AND INSERT TRIAL.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7089689 ↗

Other TEMPLATE devices

All TEMPLATE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page