Skip to content

PINNACLE 2220-00-100Recalls & reported problems

Depuy (Ireland) · PINNACLE GRATER HEAD CASE · Class 1

All models of PINNACLE (198)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Pinnacle grater head case is a surgical tray that holds bone-grating instruments. The 89 reports mostly describe the tray's coating peeling or flaking off, along with cleaning and contamination problems; no deaths or injuries are reported. No recalls name this device. Reports have climbed sharply, from 10 in the previous 12 months to 43 in the last 12.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
74
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 74 adverse event reports that list this exact device (DI 10603295424178), received between Jun 16, 2020 and May 28, 2026 — classified as 74 malfunctions.

The most frequently reported specific device problem is “Peeled/Delaminated” (69 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 15 reportsApr 2024: 3 reportsMay 2024: 0 reportsJun 2024: 12 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 7 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 22 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 4 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202622
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Peeled/Delaminated69
  • Contamination /Decontamination Problem6
  • Material Integrity Problem2
  • Break1
  • Crack1
  • No Apparent Adverse Event1
  • Scratched Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions59
  • Insufficient Information17
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 25 most recent of 74 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 28, 2026Peeled/Delaminated

    IT WAS REPORTED THAT THE TRAYS ARE STARTING TO PEEL IN THE INSIDES DUE TO WEAR AND TEAR. NO DELAYS NOTHING LEFT INSIDE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25313284 ↗
  • MalfunctionMay 28, 2026Peeled/Delaminated

    IT WAS REPORTED THAT THE TRAYS ARE STARTING TO PEEL IN THE INSIDES DUE TO WEAR AND TEAR. NO DELAYS NOTHING LEFT INSIDE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25313279 ↗
  • MalfunctionMay 28, 2026Peeled/Delaminated

    IT WAS REPORTED THAT THE TRAYS ARE STARTING TO PEEL IN THE INSIDES DUE TO WEAR AND TEAR. NO DELAYS NOTHING LEFT INSIDE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25313273 ↗
  • MalfunctionMay 25, 2026Peeled/Delaminated

    IT WAS REPORTED THAT THE PLASTIC IN THE GRATER PAN WAS PEELING OFF. NO DELAY. NO HARM TO PATIENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25270214 ↗
  • MalfunctionApr 27, 2026Peeled/Delaminated

    IT WAS REPORTED THAT IT WAS NOTED BY STAFF THAT THE PLASTIC IS COMING OFF THE METAL OF THE PINNACLE GRATER PAN AND NEEDS TO BE REPLACED. NO HARM TO PATIENT. NO DELAY. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25008271 ↗
  • MalfunctionFeb 25, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THE INSTRUMENT HAD AN UNKNOWN ALLEGATION. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24445921 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN GREATER TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427477 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN GREATER TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427474 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN GREATER TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427473 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN GREATER TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427472 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN GREATER TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427467 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN GREATER TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427457 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN GREATER TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427455 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN GREATER TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427452 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN GREATER TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427444 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN GREATER TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427433 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN GREATER TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427201 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN GREATER TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427197 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN GREATER TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427196 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN GREATER TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427194 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN GREATER TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427187 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN GREATER TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427185 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN GREATER TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427184 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN GREATER TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427181 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN GREATER TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427177 ↗

FDA clearance and classification

The FDA regulates product code FSM (tray, surgical, instrument) under 21 CFR 878.4800 as a Class I device. 21 CFR 878.4800 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the PINNACLE 2220-00-100

Has the PINNACLE 2220-00-100 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10603295424178), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the PINNACLE 2220-00-100?

The FDA's MAUDE database lists 74 adverse event reports for this device, received between Jun 16, 2020 and May 28, 2026. In those reports, the most common specific device problems are “Peeled/Delaminated” (69), “Contamination /Decontamination Problem” (6) and “Material Integrity Problem” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the PINNACLE 2220-00-100?

None of the reports are classified as deaths, 0 as injuries and 74 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the PINNACLE 2220-00-100 FDA approved?

Devices of this type (product code FSM) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Depuy (Ireland).

Who makes the PINNACLE 2220-00-100, and is it still sold?

The FDA's device database lists Depuy (Ireland) as the company responsible for it (brand PINNACLE, model 2220-00-100). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other TRAY, SURGICAL, INSTRUMENT devices

All TRAY, SURGICAL, INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page