Skip to content

ATTUNE 2546-00-428Recalls & reported problems

Depuy (Ireland) · ATTUNE KNEE SYSTEM REVISION ASSEMBLY WRENCH ADAPTOR · Class 1

All models of ATTUNE (1475)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
14
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 10603295440307), received between Feb 7, 2018 and Jul 30, 2026 — classified as 14 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility6
  • Material Twisted/Bent4
  • Material Deformation2
  • Adverse Event Without Identified Device or Use Problem1
  • Material Integrity Problem1
  • Naturally Worn1
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • No Consequences Or Impact To Patient4
  • Insufficient Information2

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 30, 2026Material Deformation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50086865 ↗
  • MalfunctionApr 20, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT DURING IMPLANT ASSEMBLY THE SURGEON NOTICED THE WRENCH NO LONGER FIT TIGHTLY AGAINST THE STEM. WHEN HE ENGAGED THE WRENCH WITH ENOUGH FORCE TO MAKE TO REQUIRED LOAD THE WRENCH SLIGHTLY SLIPPED AROUND THE FITTINGS. UPON CLOSER EVALUATION AFTER THE CASE, IT WAS FOUND THAT ENOUGH WEAR ON THE INSIDE OF THE WRENCH ALLOWED SLIGHT SLIPPAGE. THE WRENCH WAS REMOVED FROM SERVICE. THERE WAS NO SURGICAL DELAY. NO PIECES WERE LEFT BEHIND IN THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24924800 ↗
  • MalfunctionJun 3, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE WRENCH WAS SLIPPING AND NEEDED TO BE REPLACED. THE IMPACTOR WAS WORN OUT AND NEEDED TO BE REPLACED AS WELL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22134885 ↗
  • MalfunctionJun 3, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE WRENCH IS SLIPPING AND NEEDS TO BE REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22132325 ↗
  • MalfunctionSep 10, 2024Device-Device Incompatibility; Material Integrity Problem; Device-Device Incompatibility; Material Integrity Problem

    THE STEM OF THE FEMORAL COMPONENT BECAME DAMAGED DURING TORQUEING AND THE MOUNTING KEY IS SEEN TO BE ROLLED IN ITS HEXAGON, WHICH COULD ALTER THE FIXING OF THE STEM.

    Patient problems: Insufficient Information

    MAUDE report 20179806 ↗
  • MalfunctionMar 10, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE WRENCH SPRUNG WHEN TIGHTENING TIBIAL STEM ONTO BASEPLATE. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16524525 ↗
  • MalfunctionAug 9, 2022Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE WRENCH ATTACHMENT IS SPRUNG.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15190988 ↗
  • MalfunctionOct 8, 2021Device-Device Incompatibility; Naturally Worn; Device-Device Incompatibility; Naturally Worn

    1. WAS THERE ANY ADVERSE CONSEQUENCES THAT AFFECTED THE PATIENT BECAUSE OF THE REPORTED EVENT? 2. THE DER IS MARKED YES FOR PIECES BEING RETRIEVED FROM PATIENT. PLEASE VERIFY IF THIS WAS MARKED BY MISTAKE OR IF THERE WERE PIECES BROKEN OFF FROM THE INSTRUMENT. IF YES, DID IT BREAK INTO 2 OR MORE PIECES? 3. PLEASE CONFIRM IF THE WRENCH WAS REALLY WORN. IF YES, WHAT FEATURE OF THE WRENCH WAS WORN AND NOT MAKING IT FUNCTION AS INTENDED? ANSWER: 1. NO ADVERSE CONSEQUENCES TO THE PATIENT. 2. NO PIECES BROKE OFF AT ALL. THE INSTRUMENT PARTS INVOLVED WERE BEING SENT BACK JUST FOR YOUR REVIEW. 3. I COULDN¿T BE CERTAIN IF THE WRENCH WAS IN FACT WORN DOWN SO I SENT IT BACK FOR YOUR REVIEW.

    Patient problems: Insufficient Information

    MAUDE report 12599465 ↗
  • MalfunctionJun 7, 2021Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE WRENCH IS WORN ON AND NEEDS TO BE REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11946971 ↗
  • MalfunctionNov 24, 2020Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT UPON TIGHTENING THE FINAL STEM, THE TORQUE WRENCH SLIPPED DURING THE PROCESS. APPEARS TO BE STRIPPED FROM USE. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10891548 ↗
  • MalfunctionJul 10, 2020Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE ATUN REV ASSEM ADAPTOR GOT STRIPPED. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10256087 ↗
  • MalfunctionMar 2, 2020Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE WRENCH IS STRIPPED. THERE WAS NOTHING LEFT IN PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9777719 ↗
  • MalfunctionMay 6, 2019Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED THAT THERE WAS AN INSTRUMENT THAT APPEARED TO NOT WORK CORRECTLY. THE 245600428 (STEM WRENCH) GOT STUCK ON THE STEM IMPLANT AND IT WAS DIFFICULT TO SEPARATE. IF YOU LOOK ON THE INSIDE OF THE WRENCH YOU CAN SEE WHERE THE CORNERS OF THE STEM BOUND WITH THE WRENCH.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8583755 ↗
  • MalfunctionFeb 7, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    TOLERENCE BETWEEN WRENCH AND IMPLANT CAUSED DAMAGE TO THE STEM PROSTHESIS.NO DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7247756 ↗

Other WRENCH devices

All WRENCH devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page